Viability of focal dose escalation to prostate cancer intraprostatic lesions using HDR prostate brachytherapy

被引:1
|
作者
Poder, Joel [1 ,2 ,3 ,4 ]
Radvan, Samantha [3 ]
Howie, Andrew [1 ]
Kasraei, Farshad [1 ]
Parker, Annaleise [1 ]
Haworth, Annette [3 ]
Bucci, Joseph [1 ]
机构
[1] St George Hosp, Canc Care Ctr, Dept Radiat Oncol, Kogarah, NSW, Australia
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[3] Univ Sydney, Sch Phys, Camperdown, NSW, Australia
[4] St George Hosp, Canc Care Ctr, Gray St, Kogarah, NSW, Australia
关键词
Intraprostatic-lesion; Image registration; mpMRI; PSMA PET; IMAGE REGISTRATION; MRI; TUMOR; BOOST; RADIOTHERAPY; ALGORITHMS;
D O I
10.1016/j.brachy.2023.09.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: This study aimed to determine the viability of focal dose escalation to prostate cancer intraprostatic lesions (IPLs) from multiparametric magnetic resonance (mpMRI) and prostate -specific membrane antigen positron emission tomography (PSMA-PET) images using high-dose -rate (HDR) prostate brachytherapy (pBT). METHODS AND MATERIALS: Retrospective data from 20 patients treated with HDR pBT was utilized. The interobserver contouring variability of 5 observers was quantified using the dice similarity coefficient (DSC) and mean distance to agreement (MDA). Uncertainty in propagating IPL contours to trans-rectal ultrasound (TRUS) was quantified using a tissue equivalent prostate phantom. Feasibility of incorporating IPLs into HDR pBT planning was tested on retrospective patient data.RESULTS: The average observer DSC was 0.65 (PSMA-PET) and 0.52 (mpMRI). The uncer-tainty in propagating IPL contours was 0.6 mm (PSMA-PET), and 0.4 mm (mpMRI). Uncer-tainties could be accounted for by expanding IPL contours by 2 mm to create IPL PTVs. The mean D98% achieved using HDR pBT was 166% and 135% for the IPL and IPL PTV contours, respectively.CONCLUSIONS: Focal dose escalation to IPLs identified on either PSMA-PET or mpMRI is viable using TRUS-based HDR pBT. Utilizing HDR pBT allows dose escalation of up to 166% of the prescribed dose to the prostate. Crown Copyright (c) 2023 Published by Elsevier Inc. on behalf of American Brachytherapy Society. All rights reserved.
引用
收藏
页码:800 / 807
页数:8
相关论文
共 50 条
  • [21] HDR brachytherapy and online image-guided adaptive IMRT for dose escalation in prostate cancer: Comparison of brachytherapy and IMRT boosts
    Fatyga, M.
    Williamson, J.
    Dogan, N.
    Todor, D.
    Siebers, J.
    George, R.
    Hagan, M.
    Barani, I.
    MEDICAL PHYSICS, 2007, 34 (06) : 2632 - 2632
  • [22] Dose escalation by HDR brachytherapy using free-hand needle application without template in localized prostate cancer.
    Mueller, RP
    Bendel, M
    Kocher, M
    JOURNAL OF CLINICAL ONCOLOGY, 2005, 23 (16) : 431S - 431S
  • [23] PARAMETERS FAVORABLE TO INTRAPROSTATIC RADIATION DOSE ESCALATION IN MEN WITH LOCALIZED PROSTATE CANCER
    Housri, Nadine
    Ning, Holly
    Ondos, John
    Choyke, Peter
    Camphausen, Kevin
    Citrin, Deborah
    Arora, Barbara
    Shankavaram, Uma
    Kaushal, Aradhana
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (02): : 614 - 620
  • [24] PARAMETERS AFFECTING THE UNKNOWN RECTUM DOSE IN HDR BRACHYTHERAPY OF PROSTATE CANCER WHEN USING A PROSTATE DISPLACEMENT KIT
    Bo, B.
    Leite, C.
    Kofoed-Larsen, E.
    Hagen, K. Hammerfjell
    Folkedal
    Bergstrand, E. S.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S403 - S403
  • [25] A Computational model for predicting dose escalation to the dominant intraprostatic lesion during high dose rate prostate brachytherapy
    Wronski, Matt
    Ravi, Ananth
    Davidson, Melanie
    Loblaw, Andrew
    Morton, Gerard
    Chung, Hans
    Tseng, Chia-Lin
    Mendez, Lucas
    Paudel, Moti
    MEDICAL PHYSICS, 2017, 44 (08) : 4399 - 4400
  • [26] High dose rate (HDR) brachytherapy for prostate cancer in the early stages
    Bobkov, M.
    ANNALS OF ONCOLOGY, 2016, 27
  • [27] The study of dose distribution of HDR brachytherapy for prostate cancer with glass dosimeter
    Hsu, S.
    Yeh, T.
    Yeh, C.
    Hong, J.
    Kuan, W.
    Chen, W.
    Huang, D.
    MEDICAL PHYSICS, 2007, 34 (06) : 2497 - 2497
  • [28] Needle displacement effect on dose distribution in HDR brachytherapy of prostate cancer
    Huang, Y
    Blough, M
    Sadeghi, A
    Prestidge, B
    MEDICAL PHYSICS, 2004, 31 (06) : 1906 - 1906
  • [29] Results of high dose rate (HDR) brachytherapy boost for prostate cancer
    Kelly, D
    Flynn, J
    Brookover, W
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S18 - S18
  • [30] The best method for dose escalation: Prostate brachytherapy
    Morton, Gerard
    CUAJ-CANADIAN UROLOGICAL ASSOCIATION JOURNAL, 2012, 6 (03): : 196 - 198