Immunosuppressant Monitoring-Performance of the First Mass Spectrometry-Based Automated Clinical Analyzer Cascadion

被引:1
|
作者
Grote-Koska, Denis [1 ]
Czajkowski, Stefan [1 ]
Staaden, Antje [1 ]
Brand, Korbinian [1 ]
机构
[1] Hannover Med Sch, Inst Clin Chem, Carl Neuberg Str 1, D-30625 Hannover, Germany
关键词
immunosuppressants; LC-MS; MS; automated analyzer; CYCLOSPORINE-A; TACROLIMUS; ASSAY;
D O I
10.1097/FTD.0000000000001051
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background:Automatic analyzers simplify processes and may help improve standardization. The first automated analyzer based on mass spectrometry is available and offers a panel for monitoring cyclosporin A, tacrolimus, sirolimus, and everolimus. Method comparisons and evaluation tests are presented to verify the capability of the Cascadion system for use in a clinical laboratory.Methods:Sample preparation and measurements were performed using the Cascadion clinical analyzer. More than 1000 measurement values of patient samples were compared with an in vitro diagnostic-certified assay run on a liquid chromatography tandem mass spectrometry instrument. Precision and accuracy were determined using commercial quality control and external quality assessment (EQA) samples.Results:A good correlation between the 2 instruments was observed (Pearson correlation r = 0.956-0.996). Deming regression revealed 95% confidence intervals of slopes and intercepts covering the values 1 and 0, for sirolimus and everolimus, respectively, indicating equivalence of both measuring systems. However, for cyclosporin A, a bias was observed and confirmed using a Bland-Altman plot (-9.1%). Measurement repeatability and intermediate measurement precision were appropriate showing coefficients of variation of 0.9%-6.1% and 2.0%-5.3%, respectively. Accuracy according to internal quality controls was 85%-111% and 81%-100% in the EQA samples of Reference Institute of Bioanalytics and Laboratory of the Government Chemist, respectively. High robustness was found with regard to the linearity of the calibration lines (linear regression coefficient r(2) > 0.99). Carryover was negligible (0.1%).Conclusions:The Cascadion automatic analyzer produced convincing results in the measurement of patient, control, and EQA samples. The throughput was sufficient for routine use. Overall, it can be used as an alternative to open liquid chromatography tandem mass spectrometry instruments for immunosuppressant monitoring, simplifying processes without the need for specially trained personnel.
引用
收藏
页码:14 / 19
页数:6
相关论文
共 50 条
  • [31] Frontiers in mass spectrometry-based clinical proteomics for cancer diagnosis and treatment
    Haga, Yoshimi
    Minegishi, Yuriko
    Ueda, Koji
    CANCER SCIENCE, 2023, 114 (05) : 1783 - 1791
  • [32] A Comprehensive Mass Spectrometry-Based Workflow for Clinical Metabolomics Cohort Studies
    Shi, Zhan
    Li, Haohui
    Zhang, Wei
    Chen, Youxiang
    Zeng, Chunyan
    Kang, Xiuhua
    Xu, Xinping
    Xia, Zhenkun
    Qing, Bei
    Yuan, Yunchang
    Song, Guodong
    Caldana, Camila
    Hu, Junyuan
    Willmitzer, Lothar
    Li, Yan
    METABOLITES, 2022, 12 (12)
  • [33] Mass spectrometry-based proteomics and analyses of serum: A primer for the clinical investigator
    Fusaro, VA
    Stone, JH
    CLINICAL AND EXPERIMENTAL RHEUMATOLOGY, 2003, 21 (06) : S3 - S14
  • [34] Mass spectrometry-based metabolomics for clinical study: Recent progresses and applications
    Ding, Jun
    Feng, Yu-Qi
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2023, 158
  • [35] Mass Spectrometry-Based Steroid Profiling: Meeting Unmet Clinical Needs
    Choi, Man Ho
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 2021, 253 (03): : 171 - 180
  • [36] Biomarkers, Proteoforms, and Mass Spectrometry-Based Assays for Diabetes Clinical Research
    Nierves, Lorenz A.
    Lin, Tai-Tu
    Moradian, Annie
    Shen, Qingqing
    Sechi, Salvatore
    Maccoss, Michael J.
    Qu, Jun
    van Eyk, Jennifer E.
    Hoofnagle, Andrew N.
    Qian, Wei-Jun
    JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2025,
  • [37] Integration of a Dilution Module in a Mass Spectrometry-Based Online Reaction Monitoring System
    Vinh Quang Do
    Fleischer, Heidi
    Hoffmann, Dany
    Thurow, Kerstin
    AMERICAN LABORATORY, 2017, 49 (01) : 40 - 44
  • [38] Sample Matrices for Mass Spectrometry-Based Adherence Monitoring: A Systematic Critical Review
    Jacobs, Cathy M.
    Wagmann, Lea
    Meyer, Markus R.
    THERAPEUTIC DRUG MONITORING, 2024, 46 (01) : 6 - 15
  • [39] Mass spectrometry-based glycoprofiling of biopharmaceuticals by using an automated data processing tool: SimGlycan®
    Puranik, Amita
    Goswami, Rupanjan
    Sutar, Purushottam
    Tupe, Devika
    Rasam, Pratap
    Dandekar, Prajakta
    Jain, Ratnesh
    JOURNAL OF SEPARATION SCIENCE, 2023, 46 (03)
  • [40] msPurity: Automated Evaluation of Precursor Ion Purity for Mass Spectrometry-Based Fragmentation in Metabolomics
    Lawson, Thomas N.
    Weber, Ralf J. M.
    Jones, Martin R.
    Chetwynd, Andrew J.
    Rodriguez-Blanco, Giovanny
    Di Guida, Riccardo
    Viant, Mark R.
    Dunn, Warwick B.
    ANALYTICAL CHEMISTRY, 2017, 89 (04) : 2432 - 2439