Impact of the automatic tube current modulation (ATCM) system on virtual monoenergetic image quality for dual-source CT: A phantom study

被引:4
|
作者
Dabli, D. [1 ,2 ,3 ]
Durand, Quentin [1 ,2 ]
Frandon, J. [1 ]
de Oliveira, Fabien [1 ]
Pastor, Maxime [1 ]
Beregi, J. P. [1 ]
Greffier, J. [1 ,2 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, IMAGINE UR UM 103, F-30029 Nimes, France
[2] Nimes Univ Hosp, Dept Med Phys, F-30029 Nimes 9, France
[3] Montpellier Univ, Nimes Univ Hosp, Dept Med Imaging, CHU Nimes,IMAGINE UR UM 103, Bd Prof Robert Debre, F-30029 Nimes 9, France
关键词
Dual-source CT scan; Dual-energy CT; Virtual monochromatic images; Automatic tube current modulation; Detectability; ITERATIVE RECONSTRUCTION ALGORITHM; ENERGY COMPUTED-TOMOGRAPHY; LIVER-LESIONS; IODINE QUANTIFICATION; MAGNETIC-RESONANCE; EXPOSURE CONTROL; DOSE REDUCTION; OPTIMIZATION; PERFORMANCE; ACCURACY;
D O I
10.1016/j.ejmp.2023.102574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of the automatic tube current modulation (ATCM) on virtual monoenergetic images (VMIs) quality in dual-source CT(DSCT).Materials and methods: Acquisitions were performed on DSCT using the Mercury phantom. The acquisition pa-rameters for an abdomen-pelvic examination with single-energy CT(SECT) and dual-energy CT(DECT) imaging were used. Acquisitions were performed for each imaging mode using fixed mAs and ATCM. The mAs value was set to obtain a volume CT dose index of 11 mGy in fixed mAs acquisitions. This value was used as the reference mAs in ATCM acquisitions. The noise power spectrum and task-based transfer function at 40,50,60 and 70 keV levels were computed on VMIs and SECT images. The detectability index (d ') was calculated for a lesion with an iodine concentration of 10 mg/mL.Results: The noise magnitude on VMIs was higher with the ATCM system than with fixed mAs for all energy levels and section diameters of 21,26 and 31 cm. The noise texture and spatial resolution were similar between the fixed mAs and ATCM acquisitions for both imaging modes. The d' values were lower for all energy levels with ATCM than with fixed mAs acquisitions for 21 and 26 cm diameters by-39.82 +/- 9.32%, similar at 31 cm diameter-4.13 +/- 0.24% and higher at 36 cm diameter 10.40 +/- 6.69%. It was higher on VMIs at all energy levels compared to SECT images.Conclusions: The ATCM system could be used with DECT imaging to optimize patient exposure without changing the noise texture and spatial resolution of VMIs compared to fixed mAs and SECT.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Evaluating Effective Doses and Image Quality for Multislice CT Using Automatic Tube Current Modulation
    Tsai, S. F.
    Tsai, H. Y.
    Tyan, Y. S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [32] Determining the Optimal Energy Level of Virtual Monoenergetic Images in Dual-Source CT for Diagnosis of Bowel Obstruction and Colitis
    Lahuna, Loris
    Greffier, Joel
    Goupil, Jean
    Frandon, Julien
    Pastor, Maxime
    De Oliveira, Fabien
    Beregi, Jean Paul
    Dabli, Djamel
    DIAGNOSTICS, 2023, 13 (23)
  • [33] Evaluation of automatic tube current modulation in a CT scanner using a customised homogeneous phantom
    Hishaam, Urshella
    Jeyasugiththan, Jeyasingam
    Viswakula, Sameera
    Satharasinghe, D. M.
    Amalaraj, T.
    Costa, M.
    Kumarihami, A. M. C.
    Pallewatte, Aruna
    Peterson, Steve
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (06):
  • [34] Dual-source CT coronary angiography: image quality and optimal reconstruction interval
    Bastarrika, G.
    Arraiza, M.
    Arias, J.
    Broncano, J.
    Zudaire, B.
    Pueyo, J. C.
    del Barrio, Y. L. Garcia
    RADIOLOGIA, 2009, 51 (04): : 376 - 384
  • [35] Detecting Intracranial Hemorrhage Using Automatic Tube Current Modulation With Advanced Modeled Iterative Reconstruction in Unenhanced Head Single-and Dual-Energy Dual-Source CT
    Scholtz, Jan-Erik
    Wichmann, Julian L.
    Bennett, Dennis W.
    Leithner, Doris
    Bauer, Ralf W.
    Vogl, Thomas J.
    Bodelle, Boris
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 208 (05) : 1089 - 1096
  • [36] Dual-source CT Angiography in Aortic Stent Grafting: An in vitro Aorta Phantom Study of Image Noise and Radiation Dose
    Sun, Zhonghua
    Ng, Curtise
    ACADEMIC RADIOLOGY, 2010, 17 (07) : 884 - 893
  • [37] Comparison of Image Quality and Radiation Dose between Fixed Tube Current and Combined Automatic Tube Current Modulation in Craniocervical CT Angiography
    Lee, E. J.
    Lee, S. K.
    Agid, R.
    Howard, P.
    Bae, J. M.
    terBrugge, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (09) : 1754 - 1759
  • [38] Optimization of image quality and accuracy of low iodine concentration quantification as function of kVp pairs for abdominal imaging using dual-source CT: A phantom study
    Dabli, Djamel
    Frandon, Julien
    Hamard, Aymeric
    Belaouni, Asmaa
    Addala, Takieddine
    Beregi, Jean-Paul
    Greffier, Joel
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 285 - 292
  • [39] Impact of four kVp combinations available in a dual-source CT on the spectral performance of abdominal imaging: A task-based image quality assessment on phantom data
    Greffier, Joel
    Frandon, Julien
    Sadate, Alexandre
    Akessoul, Philippe
    Belaouni, Asmaa
    Beregi, Jean-Paul
    Dabli, Djamel
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 243 - 254
  • [40] A Phantom Evaluation of the Use of Ct Automatic Tube Current Modulation with Low Tube Potentials for Iodinated Contrast Studies
    Sookpeng, Supawitoo
    Martin, Colin J.
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (03) : 188 - 195