Non-enzymatic glucose sensing using a nickel hydroxide/chitosan modified screen-printed electrode incorporated into a flow injection analysis system

被引:3
|
作者
McCormick, Wesley J. [1 ]
Mcloughlin, Eva [1 ]
Mccrudden, Denis [1 ]
机构
[1] Atlantic Technol Univ, Dept Life & Phys Sci, Letterkenny F92 FC93, Donegal, Ireland
关键词
SENSORS; NANOCOMPOSITE; FOAM; FILM;
D O I
10.1039/d3ay01145c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This works presents a novel screen-printed carbon electrode modified with nickel hydroxide nanoparticles and chitosan (Ni(OH)2/CS/SPCE) for the non-enzymatic flow injection amperometric detection of glucose. The electrode was modified by drop-casting a suspension of the synthesised nanocomposite onto the screen-printed electrode and dried for 1 hour at room temperature. EDX analysis was used to investigate the chemical composition of the electrode before and after modifying. The electrochemical response of the unmodified SPCE and modified electrode was initially investigated by cyclic voltammetry (CV) using 0.1 M NaOH as the supporting electrolyte. CVs showed catalytic activity for glucose oxidation using the Ni(OH)2/CS/SPCE at 0.55 V. During flow injection analysis (FIA), 0.60 V and 1.5 mL min-1 were identified as the optimal potential and flow rate, respectively. A wide linear range of detection was observed (0.2 to 10.0 mM) with a sensitivity and limit of detection of 913 mu A mM-1 cm-2 and 0.0174 mM, respectively. The modified electrode also displayed excellent repeatability (RSD = 0.47%, n = 20) and good reproducibility (RSD = 2.52%, n = 6). The modified electrode was shown to be very selectivity for glucose over other interferences commonly found in human blood samples. The practicality of the developed flow injection-amperometric system (FIA-Amp) was validated by the quantification of glucose in real serum samples, where results were in close agreement with those obtained from the local hospital. This works presents a novel screen-printed carbon electrode modified with nickel hydroxide nanoparticles and chitosan (Ni(OH)2/CS/SPCE) for the non-enzymatic flow injection amperometric detection of glucose.
引用
收藏
页码:5071 / 5077
页数:7
相关论文
共 50 条
  • [1] Nickel Hydroxide Nanoparticles on Screen-printed Electrodes as an Impedimetric Non-enzymatic Glucose Sensor
    Laura Rinaldi, Ana
    Sobral, Santiago
    Carballo, Romina
    ELECTROANALYSIS, 2017, 29 (08) : 1961 - 1967
  • [2] Biosynthesis of Ag@CuO core–shell nanostructures for non-enzymatic glucose sensing using screen-printed electrode
    T. Dayakar
    K. Venkateswara Rao
    Jinsub Park
    Potharaju Krishna
    P. Swaroopa
    Yuexing Ji
    Journal of Materials Science: Materials in Electronics, 2019, 30 : 9725 - 9734
  • [3] Non-enzymatic sensing of glucose using screen-printed electrode modified with novel synthesized CeO2@CuO core shell nanostructure
    Dayakar, T.
    Rao, K. Venkateswara
    Bikshalu, K.
    Malapati, V.
    Sadasivuni, Kishor Kumar
    BIOSENSORS & BIOELECTRONICS, 2018, 111 : 166 - 173
  • [4] Biosynthesis of Ag@CuO core-shell nanostructures for non-enzymatic glucose sensing using screen-printed electrode
    Dayakar, T.
    Rao, K. Venkateswara
    Park, Jinsub
    Krishna, Potharaju
    Swaroopa, P.
    Ji, Yuexing
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2019, 30 (10) : 9725 - 9734
  • [5] Glucose sensing on graphite screen-printed electrode modified by sparking of copper nickel alloys
    Riman, Daniel
    Spyrou, Konstantinos
    Karantzalis, Alexandros E.
    Hrbac, Jan
    Prodromidis, Mamas I.
    TALANTA, 2017, 165 : 466 - 473
  • [6] Microliter Sample Insulin Detection Using a Screen-Printed Electrode Modified by Nickel Hydroxide
    Zhan, Zhikun
    Zhang, Hongyu
    Niu, Xuanyu
    Yu, Xiaodong
    Sun, Hui
    Sha, Xiaopeng
    Zhao, Yuliang
    Wang, Ying
    Li, Wen Jung
    ACS OMEGA, 2020, 5 (11): : 6169 - 6176
  • [7] Non-enzymatic Sensing of Glucose Using Glassy Carbon Electrode Modified with Organometallic Complex of Nickel
    Ciszewski, Aleksander
    Stepniak, Izabela
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2015, 10 (10): : 8298 - 8307
  • [8] Application of a nickel hydroxide gold nanoparticles screen-printed electrode for impedimetric sensing of glucose in artificial saliva
    Rinaldi, Ana L.
    Rodriguez-Castellon, Enrique
    Sobral, Santiago
    Carballo, Romina
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2019, 832 : 209 - 216
  • [9] Flow injection analysis of sulfide using a cinder/tetracyano nikelate modified screen-printed electrode
    Zen, JM
    Chang, JL
    Chen, PY
    Ohara, R
    Pan, KC
    ELECTROANALYSIS, 2005, 17 (09) : 739 - 743
  • [10] The effect of laser surface texturing on ZnO/MWCNT nanocomposite modified screen-printed carbon electrode for non-enzymatic glucose biosensor
    Jazi, Hamidreza Kaviani
    Sarafbidabad, Mohsen
    Ben Henda, Mouna
    Ahmadipour, Mohsen
    DIAMOND AND RELATED MATERIALS, 2025, 151