Modelling and commissioning validation of eclipse conical cone collimator for stereotactic radiosurgery using Monte Carlo simulation

被引:0
|
作者
Shende, Ravindra [1 ,2 ]
Dhoble, S. J. [2 ]
Saroj, Dinesh [1 ]
Gupta, Gourav [1 ]
机构
[1] Balco Med Ctr, Dept Radiat Oncol, New Raipur, Chhattisgarh, India
[2] Rashtrasant Tukadoji Maharaj Nagpur Univ, Dept Phys, Nagpur, Maharashtra, India
关键词
commissioning validation; conical cone; Monte Carlo; PRIMO; SRS; VARIAN; DETECTORS; DEPTH; BEAMS;
D O I
10.1017/S1460396924000025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The miniaturized conical cones for stereotactic radiosurgery (SRS) make it challenging in measurement of dosimetric data needed for commissioning of treatment planning system. This study aims at validating dosimetric characteristics of conical cone collimator manufactured by Varian using Monte Carlo (MC) simulation technique.Methods & Material: Percentage depth dose (PDD), tissue maximum ratio (TMR), lateral dose profile (LDP) and output factor (OF) were measured for cones with diameters of 5mm, 7 center dot 5mm, 10mm, 12 center dot 5 mm, 15 mm and 17 center dot 5 mm using EDGE detector for 6MV flattening filter-free (FFF) beam from Truebeam linac. Similarly, MC modelling of linac for 6MVFFF beam and simulation of conical cones were performed in PRIMO. Subsequently, measured beam data were validated by comparing them with results obtained from MC simulation.Results: The measured and MC-simulated PDDs or TMRs showed close agreement within 3% except for cone of 5mm diameter. Deviations between measured and simulated PDDs or TMRs were substantially higher for 5mm cone. The maximum deviations at depth of 10cm, 20cm and at range of 50% dose were found 4 center dot 05%, 7 center dot 52%, 5 center dot 52% for PDD and 4 center dot 04%, 7 center dot 03%, 5 center dot 23% for TMR with 5mm cone, respectively. The measured LDPs acquired for all the cones showed close agreement with MC LDPs except in penumbra region around 80% and 20% dose profile. Measured and MC full-width half maxima of dose profiles agreed with nominal cone size within +/- 0 center dot 2 mm. Measured and MC OFs showed excellent agreement for cone sizes >= 10 mm. However, deviation consistently increases as the size of the cone gets smaller.Findings: MC model of conical cones for SRS has been presented and validated. Very good agreement was found between experimentally measured and MC-simulated data. The dosimetry dataset obtained in this study validated using MC model may be used to benchmark beam data measured for commissioning of SRS for cone planning.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Monte Carlo simulations of a scintillation camera using GATE: validation and application modelling
    Staelens, S
    Strul, D
    Santin, G
    Vandenberghe, S
    Koole, M
    D'Asseler, Y
    Lemahieu, I
    van de Walle, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (18): : 3021 - 3042
  • [42] Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPE
    Gonzalez, W.
    Lallena, A. M.
    Alfonso, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (11): : 3417 - 3431
  • [43] Effect of collimator characteristics on 123I-MIBG phantom imaging using Monte Carlo simulation
    Okuda, K.
    Nakajima, K.
    Kitamura, C.
    Kirihara, Y.
    Taki, J.
    Hashimoto, M.
    Kinuya, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S20 - S20
  • [44] Monte Carlo simulation of a multi-leaf collimator design for telecobalt machine using BEAMnrc code
    Ayyangar, Komanduri M.
    Kumar, M. Dinesh
    Narayan, Pradush
    Jesuraj, Fenedit
    Raju, M. R.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (01) : 23 - 32
  • [45] MONTE-CARLO SIMULATION OF TRANSMISSION STUDIES USING A PLANAR SOURCE WITH A PARALLEL COLLIMATOR AND A LINE SOURCE WITH A FAN-BEAM COLLIMATOR
    LJUNGBERG, M
    STRAND, SE
    RAJEEVAN, N
    KING, MA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1577 - 1584
  • [46] Investigation of Collimator Broad Correction for Dopamine Transporter SPECT Imaging using Monte-Carlo Simulation
    Funada, Keita
    Takahashi, Akihiko
    Himuro, Kazuhiko
    Baba, Shingo
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [47] Optimization of the design parameters for a thyroid care nuclide monitoring diverging collimator using Monte Carlo simulation
    Dong-Hee Han
    Seung-Jae Lee
    Jang-Oh Kim
    Da-Eun Kwon
    Kyung-Hwan Jung
    Cheol-Ha Baek
    Journal of the Korean Physical Society, 2022, 81 : 675 - 679
  • [48] Optimization of the design parameters for a thyroid care nuclide monitoring diverging collimator using Monte Carlo simulation
    Han, Dong-Hee
    Lee, Seung-Jae
    Kim, Jang-Oh
    Kwon, Da-Eun
    Jung, Kyung-Hwan
    Baek, Cheol-Ha
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 81 (07) : 675 - 679
  • [49] Monte Carlo simulation for Ga-67 imaging using a low-medium energy collimator
    Kojima, A.
    Watanabe, H.
    Takaki, A.
    Matsumoto, M.
    Uehara, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S338 - S338
  • [50] Study on the tongue and groove effect of the Elekta multileaf collimator using Monte Carlo simulation and film dosimetry
    Haryanto, F
    Fippel, M
    Bakai, A
    Nüsslin, F
    STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (01) : 57 - 61