摘要
Background: A fluid-attenuated inversion recovery (FLAIR) method eliminates the cerebrospinal fluid (CSF) signal, enhancing white matter lesion detection by enhancing the contrast between the lesion and CSF. Three-dimensional (3D) volume acquisition has the advantage of multiplanar reformation of contiguous slices yielding improved signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs). To our knowledge, there are only three studies comparing 3D-and 2D-FLAIR sequences with respect to multiple sclerosis (MS) lesions at 3 tesla.Aims and objectives: This study aimed to determine the efficiency of 3D-FLAIR in the detection of lesions of multiple sclerosis in terms of spatial and contrast resolutions in comparison with 2D-FLAIR sequences.Methodology: A total of 75 patients with MS undergoing magnetic resonance imaging (MRI) brain at the Department of Radiology, Krishna Institute of Medical Sciences (KIMS), Secunderabad, Telangana, India. This is an observational comparative study. Independent-samples t-tests were performed in the present study to compare the number of lesions detected. The measured CNR and SNR values were subjected to Mann-Whitney U test.Results: As a result of the 3D-FLAIR, more lesions were found as compared to 2D-FLAIR (p = 0.001). There was a greater CNRs for 3D-FLAIR images than for 2D-FLAIR images (p = 0.001). Lesions, CSF, white matter, and gray matter showed significantly higher SNRs with 3D-FLAIR (p = 0.001).Conclusion: 3D-FLAIR has exhibited greater sensitivity in detecting lesions associated with MS when contrasted with the 2D-FLAIR sequence. Significantly more lesions and higher SNRs and CNRs were detected with 3D-FLAIR in contrast to 2D-FLAIR. 3D-FLAIR may be considered the sequence of choice for MS imaging in the future.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Utility of 3-T FLAIR and 3D short tau inversion recovery MR imaging in the preoperative diagnosis of hippocampal sclerosis: Direct comparison with 1.5-T FLAIR MR imaging
    Hashiguchi, Kimiaki
    Morioka, Takato
    Murakami, Nobuya
    Suzuki, Satoshi O.
    Hiwatashi, Akio
    Yoshiura, Takashi
    Sasaki, Tomio
    EPILEPSIA, 2010, 51 (09) : 1820 - 1828
  • [32] Fast fluid-attenuated inversion-recovery imaging: first experience with a 3D version in epilepsy
    U. C. Wieshmann
    G. J. Barker
    M. R. Symms
    P. A. Bartlett
    J. M. Stevens
    S. D. Shorvon
    Neuroradiology, 1998, 40 : 483 - 489
  • [33] Multiple sclerosis lesion detection with 3D double inversion recovery (DIR) as compared to 3D fluid low attenuation inversion recovery (T2-FLAIR): A systematic review and meta-analysis
    Mostardeiro, Thomaz R.
    Schmitt, Luiza Giuliani
    de Campos, Fillipe Thiago Xavier
    Xi, Yin
    Torri, Giovanni Brondani
    Carvalho, Bruno Murad
    Feltrin, Fabricio Stewan
    MULTIPLE SCLEROSIS AND RELATED DISORDERS, 2025, 93
  • [34] Fast fluid-attenuated inversion-recovery imaging: first experience with a 3D version in epilepsy
    Wieshmann, UC
    Barker, GJ
    Symms, MR
    Bartlett, PA
    Stevens, JM
    Shorvon, SD
    NEURORADIOLOGY, 1998, 40 (08) : 483 - 489
  • [35] A comparative MRI study for white matter hyperintensities detection: 2D-FLAIR, FSE PD 2D, 3D-FLAIR and FLAIR MIP
    Paniagua Bravo, A.
    Sanchez Hernandez, J. J.
    Ibanez Sanz, L.
    Alba de Caceres, I.
    Crespo San Jose, J. L.
    Garcia-Castano Gandariaga, B.
    BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1035):
  • [36] PARTIALLY SATURATED FLUID ATTENUATED INVERSION-RECOVERY (FLAIR) SEQUENCES IN MULTIPLE-SCLEROSIS - COMPARISON WITH FULLY RELAXED FLAIR AND CONVENTIONAL SPIN-ECHO
    BARATTI, C
    BARKHOF, F
    HOOGENRAAD, F
    VALK, J
    MAGNETIC RESONANCE IMAGING, 1995, 13 (04) : 513 - 521
  • [37] Isotropic 3D fast FLAIR imaging of the brain in multiple sclerosis patients:: initial experience
    Tan, IL
    Pouwels, PJW
    van Schijndel, RA
    Adèr, HJ
    Manoliu, RA
    Barkhof, F
    EUROPEAN RADIOLOGY, 2002, 12 (03) : 559 - 567
  • [38] Isotropic 3D fast FLAIR imaging of the brain in multiple sclerosis patients: initial experience
    I. Tan
    P. Pouwels
    R. van Schijndel
    H. Adèr
    R. Manoliu
    F. Barkhof
    European Radiology, 2002, 12 : 559 - 567
  • [39] The value of postcontrast delayed 3D fluid-attenuated inversion recovery MRI in the diagnosis of unilateral peripheral vestibular dysfunction
    Zhang, Menglu
    Zhao, Tongtong
    Feng, Yufei
    Li, Xiang
    Song, Ning
    Ma, Xinyan
    Yang, Xu
    Sui, Rubo
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (08) : 5072 - 5088
  • [40] MR Imaging in Multiple Sclerosis: The Accuracy of 3D Double Inversion Recovery at 3 Tesla and the Potential for Single Sequence Imaging
    Khangure, S. R.
    Khangure, M. S.
    NEURORADIOLOGY JOURNAL, 2011, 24 (01): : 92 - 99