Microwave Measurement System for Characterizing Liquid Samples Based on a Modified HMSIW

被引:3
|
作者
Wu, Wen-Jing [1 ]
Zhao, Wen-Sheng [2 ]
机构
[1] Hangzhou City Univ, Sch Informat & Elect Engn, Hangzhou 310015, Peoples R China
[2] Hangzhou Dianzi Univ, Sch Elect & Informat, Zhejiang Prov Key Lab Large Scale Integrated Circ, Hangzhou 310018, Peoples R China
基金
中国国家自然科学基金;
关键词
Sensors; Microwave sensors; Permittivity; Resonant frequency; Permittivity measurement; Microwave measurement; Magnetic field measurement; Complex permittivity; dc voltage; half-mode substrate integrated waveguide (HMSIW); measurement system; RING-RESONATOR; SIMULTANEOUS PERMITTIVITY; INTERDIGITAL CAPACITOR; THICKNESS MEASUREMENT; MICROFLUIDIC SENSOR; DISPLACEMENT; PHASE; PAIRS;
D O I
10.1109/JSEN.2023.3307193
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A microwave measurement system for retrieving complex permittivity of liquid sample based on the modified half-mode substrate integrated waveguide (HMSIW) is proposed in this article. The proposed sensing system consists of the passive resonant sensor and the RF circuits. Meanwhile, the passive sensor is evolved from the traditional full-mode SIW, the HMSIW is obtained by cutting along the magnetic wall, and the resonant frequency for TE101 is almost unchanged. By adding a serrated interdigital structure on the top surface of HMSIW, the confinement of the electric field is enhanced and it is beneficial to improve the sensitivity. In addition, the resonant frequency of TE101 shifts toward the low-frequency band, which is beneficial to reduce the electrical size of the passive sensor. The RF circuits are composed of a power divider and a gain/phase detector; it should be noted that the detector can output two dc voltages, i.e., V-mag and V-phs. As liquid samples with different complex permittivity are loaded at the sensing area, V-mag and V-phs will have different output dc voltages. According to the relationship among complex permittivity of liquid samples, Vmag and Vphs, the mathematical models can be established to predict the unknown liquid samples. In measurement, the maximum errors for extracting real and imaginary parts of complex permittivity are -3.84% and 11.72%, respectively. The proposed sensing system has the advantages of discarding the dependence on an expensive vector network analyzer (VNA) and lower the cost.
引用
收藏
页码:22466 / 22475
页数:10
相关论文
共 50 条
  • [41] The Design of Frequency measure System Based on Microwave Resonator of Wetness measurement
    Wang Hongyun
    Yao Zhimin
    Shi Lianyan
    ITCS: 2009 INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY AND COMPUTER SCIENCE, PROCEEDINGS, VOL 2, PROCEEDINGS, 2009, : 346 - 349
  • [42] Plant Moisture Content Measurement System Based on Microwave Pseudo Waveguide
    Zhou, Tao
    Guo, Zixuan
    Sun, Qiaoling
    Li, Yi
    Deng, Liangbin
    Jiang, Geng
    Yin, Bangyong
    Yang, Yuqiu
    Wu, Junyao
    Cai, Huan
    Jiang, Xi
    Yin, Jungang
    Deng, Linfeng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2024, 73
  • [43] Automatic characterization of capillary flow profile of liquid samples on μTADs based on capacitance measurement
    Zhou, Jie
    Li, Haonan
    Li, Xionghui
    Liang, Xuanying
    Feng, Zitao
    He, Qinghao
    Zhang, Muyang
    Chen, Xinyi
    Chen, Huilin
    Zhang, Huiru
    Guo, Weijin
    JOURNAL OF CHROMATOGRAPHY A, 2024, 1735
  • [44] Liquid Level Measurement System based on a Coherent Optical Sensor
    Melchionni, Dario
    Pesatori, Alessandro
    Norgia, Michele
    2014 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC) PROCEEDINGS, 2014, : 968 - 971
  • [45] Liquid Refractive Index Measurement System Based on Electrowetting Lens
    Li, Shi-Long
    Nie, Zhong-Quan
    Tian, Yan-Ting
    Liu, Chao
    MICROMACHINES, 2019, 10 (08)
  • [46] Research of the Measurement System of Liquid Displacement based on Angular Tracking
    Tao, Huanqi
    Ma, Shuangbao
    INFORMATION TECHNOLOGY FOR MANUFACTURING SYSTEMS, PTS 1 AND 2, 2010, : 493 - +
  • [47] A Dielectric Constant Measurement System for Liquid Based on SIW Resonator
    Chen, Shichang
    Guo, Mengchu
    Xu, Kuiwen
    Zhao, Peng
    Hu, Yue
    Dong, Linxi
    Wang, Gaofeng
    IEEE ACCESS, 2018, 6 : 41163 - 41172
  • [48] MEASUREMENT OF THE MICROWAVE-ABSORPTION FOR SMALL SAMPLES IN A COAXIAL LINE
    GARNER, HR
    LEWIS, AC
    OHKAWA, T
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1991, 39 (05) : 890 - 892
  • [49] MEASUREMENT OF THE ELECTRICAL PARAMETERS OF THIN SAMPLES OF POLYDISPERSE, LIQUID AND SOLID DIELECTRICS THAT EXHIBIT HIGH ABSORPTION AT MICROWAVE FREQUENCIES.
    Shchegoleva, T.Yu.
    Radio Engineering and Electronic Physics (English translation of Radiotekhnika i Elektronika), 1981, 26 (11): : 58 - 62
  • [50] SURVEY OF MICROWAVE DIELECTRIC TECHNIQUES FOR SMALL LIQUID AND SOLID SAMPLES
    BIRNBAUM, G