Risk Factors for the Drift Phenomenon in O-arm Navigation-Assisted Pedicle Screw Placement during Spinal Deformity Surgery

被引:3
|
作者
Zhao, Jianquan [1 ]
Jiang, Heng [1 ]
Zhuge, Yingjie [2 ]
Gao, Rui [1 ]
Wang, Ce [1 ]
Ma, Jun [1 ]
Zhou, Xuhui [1 ]
机构
[1] Second Mil Med Univ, Shanghai Changzheng Hosp, Dept Orthopaed Surg, Shanghai 200003, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 6, Dept Orthopaed Surg, Shanghai, Peoples R China
关键词
Deformity; Drift phenomenon; Navigation; O-arm; Pedicle screw; ANTERIOR VERTEBRAL HEIGHT; FREE-HAND; FRACTURED VERTEBRA; ACCURACY; SCOLIOSIS; RESTORATION; FLUOROSCOPY; INSERTION; FIXATION; SAFETY;
D O I
10.1111/os.13557
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective Intraoperative O-arm navigation systems improve the accuracy of spinal instrumentation placement. However, deviation of the pedicle screw from the guide line might occur. The aim of the present study was to explore the causes of and countermeasures for the drift phenomenon during pedicle screw implantation with the aid of an O-arm three-dimensional navigation system in spinal deformity surgery. Methods This was a retrospective analysis of 341 patients with spinal deformity who underwent O-arm navigation system-assisted pedicle screw placement from July 2015 to June 2019. The patient's general condition, Cobb angle, apical vertebra position, softness index, spinal release status, fixed reference frame position, and distance between the navigation vertebral body and the reference frame were collected and compared by independent-samples t test or Pearson's chi-square analysis. The potential risk factors for the drift phenomenon were identified using binary logistic regression analysis. Results The drift phenomenon occurred in 57 patients during the first navigation-assisted pedicle screw placement, for an incidence of 16.7% (57/341). There were significant differences in factors such as the apical vertebra position, softness index, spinal release status, and distance between the vertebral body and the reference frame when the drift phenomenon occurred (P < 0.05). Binary logistic regression analysis showed that the softness index, spinal release status, and distance between the vertebral body and the reference frame when drifting occurred were independent risk factors for the drift phenomenon during O-arm navigation-assisted pedicle screw placement. Conclusion During the use of an O-arm navigation system to assist with pedicle screw placement, pedicle screws should not be placed away from the reference frame, and spinal osteotomy and release should be performed after pedicle screw placement. In addition, the accuracy of O-arm navigation-assisted pedicle screw placement will be affected more in those with larger softness indices.
引用
收藏
页码:118 / 123
页数:6
相关论文
共 50 条
  • [1] Pedicle Screw Placement With O-arm and Stealth Navigation
    Patil, Suresh
    Lindley, Emily M.
    Burger, Evalina L.
    Yoshihara, Hiroyuki
    Patel, Vikas V.
    ORTHOPEDICS, 2012, 35 (01) : E61 - E65
  • [2] Benefits and Pitfalls of O-Arm Navigation in Cervical Pedicle Screw
    Shin, Hong Kyung
    Jeon, Sang Ryong
    Roh, Sung Woo
    Park, Jin Hoon
    WORLD NEUROSURGERY, 2022, 159 : E460 - E465
  • [3] O-arm navigation versus C-arm guidance for pedicle screw placement in spine surgery: a systematic review and meta-analysis
    Weili Feng
    Weihao Wang
    Shubiao Chen
    Kezhou Wu
    Hu Wang
    International Orthopaedics, 2020, 44 : 919 - 926
  • [4] O-arm navigation versus C-arm guidance for pedicle screw placement in spine surgery: a systematic review and meta-analysis
    Feng, Weili
    Wang, Weihao
    Chen, Shubiao
    Wu, Kezhou
    Wang, Hu
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (05) : 919 - 926
  • [5] Robot-assisted versus navigation-assisted screw placement in spinal vertebrae
    Yu, Tong
    Jiao, Jian-Hang
    Wang, Yang
    Wang, Qing-Yu
    Jiang, Wei-Bo
    Wang, Zhong-Han
    Wu, Min-Fei
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (02) : 527 - 532
  • [6] Robot-assisted versus navigation-assisted screw placement in spinal vertebrae
    Tong Yu
    Jian-Hang Jiao
    Yang Wang
    Qing-Yu Wang
    Wei-Bo Jiang
    Zhong-Han Wang
    Min-Fei Wu
    International Orthopaedics, 2023, 47 : 527 - 532
  • [7] O-Arm Assisted Cervicothoracic Spine Pedicle Screw Placement Accuracy Is Higher Than C-Arm Fluoroscopy
    Inoue, Tomohisa
    Wada, Keiji
    Tominaga, Ayako
    Tamaki, Ryo
    Hirota, Tomoya
    Okazaki, Ken
    WORLD NEUROSURGERY, 2022, 158 : E996 - E1001
  • [8] Safety and risk factors of TINAVI robot-assisted percutaneous pedicle screw placement in spinal surgery
    Ren-Jie Zhang
    Lu-Ping Zhou
    Lai Zhang
    Hua-Qing Zhang
    Jian-Xiang Zhang
    Cai-Liang Shen
    Journal of Orthopaedic Surgery and Research, 17
  • [9] Safety and risk factors of TINAVI robot-assisted percutaneous pedicle screw placement in spinal surgery
    Zhang, Ren-Jie
    Zhou, Lu-Ping
    Zhang, Lai
    Zhang, Hua-Qing
    Zhang, Jian-Xiang
    Shen, Cai-Liang
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [10] Case series: O-arm navigation assisted by the Wiltse approach improves the accuracy of pedicle screw placement in ankylosing spondylitis combined with thoracolumbar fractures
    Li, Chang-Ming
    Zhao, Shi-Jie
    Xu, Jian-Zhu
    Li, Qiang
    Quan, Ren-Fu
    Deng, Xiao-Mei
    MEDICINE, 2023, 102 (52) : E36807