Carboxymethylated maltodextrin as a chiral selector for the separation of some basic drug enantiomers using capillary electrophoresis

被引:6
|
作者
Alawadi, Mustafa [1 ]
Fakhari, Ali Reza [1 ]
Bayatloo, Mohammad Reza [1 ]
Nojavan, Saeed [1 ]
机构
[1] Shahid Beheshti Univ, Dept Analyt Chem & Pollutants, Tehran, Iran
关键词
Carboxymethylated maltodextrin; Basic drugs; Capillary electrophoresis; Enantiomer separation; PHARMACEUTICAL FORMULATIONS; AMLODIPINE ENANTIOMERS; IONIC LIQUID; ENANTIOSEPARATION; QUANTIFICATION; CYCLODEXTRIN;
D O I
10.1016/j.chroma.2023.464335
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this work, carboxymethylated maltodextrin (Cm-MD) was successfully synthesized as an efficient anionic chiral selector and applied for the enantiomer separation of some basic drugs including tramadol, venlafaxine, verapamil, hydroxyzine, citalopram, fluoxetine, and amlodipine by capillary electrophoresis (CE). The synthesized chiral selector was characterized by the nuclear magnetic resonance and Fourier transform infrared spectrophotometry. Under the optimized Cm-MD modified CE conditions (background electrolyte: phosphate buffer (pH 5.0, 50 mM) containing 5% (w/v) Cm-MD; applied voltage: 20 kV; and capillary column temperature: 25 C), successful enantiomer separation of all studied chiral drugs were observed. By comparison of Cm-MD and MD for enantiomer separation of the model drugs, it was revealed that Cm-MD exhibits a higher resolution in comparison to the MD modified CE. This enhanced resolution could be attributed to the electrostatic interactions between the cationic drugs and anionic Cm-MD and opposite direction mobility of the host-guest complex relative to the chiral analyte. The optimized Cm-MD modified CE method was successfully used for the assay of the enantiomers of citalopram and venlafaxine in commercial tablets. The proposed method showed the linear range of 5.0-150.0 mg/L and 10.0-150.0 mg/L for both enantiomers of citalopram and venlafaxine, respectively. The limits of quantification were 5.0 and 10.0 mg/L for the enantiomers of citalopram and venlafaxine, respectively. The limit of detection for all enantiomers was found to be < 3.0 mg/L. Intra-and inter-day RSDs (n = 4) were less than 9.7%. The relative errors were less than 9.4% for all enantiomers. The obtained results in this research show that Cm-MD as a new, efficient and inexpensive chiral selector can be used for enantiomer separation of basic drugs using the CE technique.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Separation of basic drug enantiomers by capillary electrophoresis using chicken α1-acid glycoprotein:: Insight into chiral recognition mechanism
    Matsunaga, H
    Sadakane, Y
    Haginaka, J
    ELECTROPHORESIS, 2003, 24 (15) : 2442 - 2447
  • [32] Separation of drug enantiomers by liquid chromatography and capillary electrophoresis, using immobilized proteins as chiral selectors
    Millot, MC
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 797 (1-2): : 131 - 159
  • [33] Study on the separation of ofloxacin enantiomers by hydroxylpropyl-β-cyclodextrin as a chiral selector in capillary electrophoresis: a computational approach
    Suliman, FakhrEldin O.
    Elbashir, Abdalla A.
    Schmitz, Oliver J.
    JOURNAL OF INCLUSION PHENOMENA AND MACROCYCLIC CHEMISTRY, 2015, 83 (1-2) : 119 - 129
  • [34] Separation of Enantiomers of Cathinones by Capillary Electrophoresis Using Cyclodextrins as Chiral Selectors
    Hegrova, Martina
    Rezankova, Klara
    Kuchar, Martin
    Rezanka, Pavel
    CHEMICKE LISTY, 2016, 110 (03): : 200 - 203
  • [35] Application of rifampicin as a chiral selector for enantioresolution of basic drugs using capillary electrophoresis
    Dixit, Shuchi
    Park, Jung Hag
    JOURNAL OF CHROMATOGRAPHY A, 2016, 1453 : 138 - 142
  • [36] Separation of profen enantiomers by capillary electrophoresis using cyclodextrins as chiral selectors
    Blanco, M
    Coello, J
    Iturriaga, H
    Maspoch, S
    Pérez-Maseda, C
    JOURNAL OF CHROMATOGRAPHY A, 1998, 793 (01) : 165 - 175
  • [37] CAPILLARY ZONE ELECTROPHORESIS FOR SEPARATION OF DRUG ENANTIOMERS USING CYCLODEXTRINS AS CHIRAL SELECTORS - INFLUENCE OF EXPERIMENTAL PARAMETERS ON SEPARATION
    PALMARSDOTTIR, S
    EDHOLM, LE
    JOURNAL OF CHROMATOGRAPHY A, 1994, 666 (1-2) : 337 - 350
  • [38] Boric acid modified β-CD polymer as chiral selector for chiral drug separation using gravity mediated capillary electrophoresis
    Nan, Yaqin
    Yuan, Xiaoqian
    Xu, Xinyu
    Xu, Yanmei
    Jin, Xinghua
    Li, Youxin
    MICROCHEMICAL JOURNAL, 2025, 210
  • [39] Chiral separation of bupivacaine enantiomers by capillary electrophoresis partial-filling technique with human serum albumin as chiral selector
    Martínez-Pla, JJ
    Martín-Biosca, Y
    Sagrado, S
    Villanueva-Camañas, RM
    Medina-Hernández, MJ
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1048 (01) : 111 - 118
  • [40] Chiral separation of simendan enantiomers by capillary zone electrophoresis
    Lu, XM
    Zhang, DD
    Peng, Y
    Li, FM
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (07) : 1003 - 1006