One-year follow-up comparison of two hybrid closed-loop systems in Italian children and adults with type 1 diabetes

被引:17
|
作者
Bassi, Marta [1 ,2 ]
Patti, Luca [3 ]
Silvestrini, Irene [3 ]
Strati, Marina Francesca [1 ,2 ]
Ponzano, Marta [4 ]
Minuto, Nicola [2 ]
Maggi, Davide [3 ,5 ]
机构
[1] Univ Genoa, Dept Neurosci Rehabil Ophthalmol Genet Maternal &, Genoa, Italy
[2] Univ Genoa, IRCCS Ist Giannina Gaslini, Dept Pediat, Genoa, Italy
[3] Univ Genoa, Dept Internal Med & Med Specialties DiMI, Genoa, Italy
[4] Univ Genoa, Dept Hlth Sci DiSSAL, Genoa, Italy
[5] IRCCS Osped Policlin San Martino Genoa, Diabet Clin, Genoa, Italy
来源
关键词
AHCL (Advanced Hybrid Closed Loop); type; 1; diabetes; CGM; continous glucose monitoring; TIR (time in range); CSII; continuous subcutaneous insulin infusion;
D O I
10.3389/fendo.2023.1099024
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and aimsTandem Control-IQ and MiniMed 780G are the main Advanced Hybrid Closed Loop (AHCL) systems currently available in pediatric and adult patients with Type 1 Diabetes (T1D). The aim of our study was to evaluate glycemic control after 1-year of follow-up extending our previous study of 1-month comparison between the two systems. MethodsWe retrospectively compared clinical and continuous glucose monitoring (CGM) data from the patients included in the previous study which have completed 1-year observation period. The study population consisted of 74 patients, 42 Minimed 780G users and 32 Tandem Control-IQ users. Linear mixed models with random intercept were performed to study the variations over time and the interaction between time and system; Mann-Whitney or T-test were used to compare systems at 1-year. ResultsBoth systems have been shown to be effective in maintaining the glycemic improvement achieved one month after starting AHCL. Significant changes over time were observed for TIR, TAR, TAR>250mg/dl, average glucose levels and SD (p<0.001). At 1-year follow-up Minimed 780G obtained better improvement in TIR (p<0.001), TAR (p=0.002), TAR>250mg/dl (p=0.001), average glucose levels (p<0.001). The comparison of the glycemic parameters at 1-year showed a significant superiority of Minimed 780G in terms of TIR (71% vs 68%; p=0.001), TAR (p=0.001), TAR>250 (p=0.009), average glucose levels(p=0.001) and SD (p=0.031). ConclusionsThe use of AHCL systems led to a significant improvement of glycemic control at 1-month, which is maintained at 1-year follow-up. MiniMed is more effective than Tandem in reaching the International recommended glycemic targets. Continuous training and education in the use of technology is essential to get the best out of the most advanced technological tools.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Closed-loop therapy in children with type 1 diabetes
    Pavlicek, Vojtech
    DIABETOLOGE, 2022, 18 (03): : 329 - 330
  • [22] PSYCHOLOGICAL WELL-BEING AND QUALITY OF LIFE AFTER ONE-YEAR EXPERIENCE OF ADVANCE HYBRID CLOSED-LOOP SYSTEM IN ADULTS WITH TYPE 1 DIABETES PREVIOUSLY NAIVE TO DIABETES TECHNOLOGY
    Cyranka, K.
    Matejko, B.
    Juza, A.
    Kiec-Wilk, B.
    Krzyzowska, S.
    Cohen, O.
    Maecki, M.
    Klupa, T.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2023, 25 : A110 - A110
  • [23] Closed-loop insulin delivery systems in children and adolescents with type 1 diabetes
    Dovc, Klemen
    Battelino, Tadej
    EXPERT OPINION ON DRUG DELIVERY, 2020, 17 (02) : 157 - 166
  • [24] Celiac Autoimmunity in Children with Type 1 Diabetes: A Two-Year Follow-Up
    Simmons, Jill H.
    Klingensmith, Georgeanna J.
    McFann, Kim
    Rewers, Marian
    Ide, Lisa M.
    Taki, Iman
    Liu, Edwin
    Hoffenberg, Edward J.
    JOURNAL OF PEDIATRICS, 2011, 158 (02): : 276 - U141
  • [25] Postprandial Glucose Control With Different Hybrid Closed-Loop Systems According to Type of Meal in Adults With Type 1 Diabetes
    Scida, Giuseppe
    Corrado, Alessandra
    Abuqwider, Jumana
    Lupoli, Roberta
    Rainone, Carmen
    Della Pepa, Giuseppe
    Masulli, Maria
    Annuzzi, Giovanni
    Bozzetto, Lutgarda
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024,
  • [26] A Real-Life Experience with Hybrid Closed-Loop Insulin Delivery Systems in People with Type 1 Diabetes: One Year After
    Convertino, Alessio
    Giuliani, Chiara
    Bitterman, Olimpia
    Napoli, Angela
    DIABETES, 2022, 71
  • [27] Incidences of psychiatric disorders before and after kidney transplantation: One-year follow-up comparison of adults and children
    Fukunishi, I
    Honda, M
    Hasegawa, H
    Obayashi, K
    Watanabe, T
    Hiraga, S
    NEUROLOGY PSYCHIATRY AND BRAIN RESEARCH, 1995, 3 (02) : 125 - 128
  • [28] Glycemic control metrics in a cohort of hospitalized patients with type 1 diabetes using hybrid closed-loop and advanced hybrid closed-loop systems
    Medina, Ana Maria Gomez
    Henao-Carrillo, Diana Cristina
    Yepes, Carlos
    Silva, Julio
    Gomzalez, Javier Alberto Gomez
    Cortes, David
    Robledo, Sofia
    Mejia, Gabriela
    Rondon, Martin
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2023, 204
  • [29] Hybrid Closed-Loop Control in Young Children with Type 1 Diabetes The authors reply
    Wadwa, R. Paul
    Beck, Roy W.
    Breton, Marc D.
    NEW ENGLAND JOURNAL OF MEDICINE, 2023, 388 (23): : 2206 - 2207
  • [30] Evaluation of hybrid closed-loop insulin delivery system in type 1 diabetes in real-world clinical practice: One-year observational study
    Eldib, Ahmed
    Dhaver, Shilton
    Kibaa, Karim
    Atakov-Castillo, Astrid
    Salah, Tareq
    Al-Badri, Marwa
    Khater, Abdelrahman
    Mccarragher, Ryan
    Elenani, Omnia
    Toschi, Elena
    Hamdy, Osama
    WORLD JOURNAL OF DIABETES, 2024, 15 (03)