Calibration of a neutron dose rate meter in various neutron standard fields

被引:0
|
作者
Tuan Khai Nguyen [1 ]
Hoai-Nam Tran [2 ,3 ]
Quynh Ngoc Nguyen [4 ]
Thi My Linh Dang [4 ]
Van Loat Bui [5 ]
Thiansin Liamsuwan [6 ,7 ]
Ngoc-Thiem Le [4 ]
机构
[1] Vietnam Agency for Radiation and Nuclear Safety
[2] Institute of Fundamental and Applied Sciences, Duy Tan University
[3] Faculty of Natural Sciences, Duy Tan University
[4] Institute for Nuclear Science and Technology,VINATOM5. Faculty of Physics, VNU University of Science6. Nuclear Research and Development Division, Thailand Institute of Nuclear Technology7. Faculty of Medicine and Public Health, HRH Princess Chulabhorn Coll
关键词
Neutron standard fields; Neutron dose rate meter; Calibration factor; 252Cf source; 241Am-Be source;
D O I
暂无
中图分类号
TL818.1 [];
学科分类号
082704 ;
摘要
This paper presents the calibration of a neutron dose rate meter and the evaluation of its calibration factors(CFs)in several neutron standard fields(i.e.,two standard fields with bare sources of;Cf and;Am-Be,and five simulated workplace fields with;Am-Be moderated sources).The calibration in standard fields with bare sources was conducted by following the recommendations of the ISO 8529 standard.The measured total neutron ambient dose equivalent rates,denoted as■,were analyzed to obtain direct components,denoted as■,using a reduced fitting method.The CF was then calculated as the ratio between the conventional true value of the neutron ambient dose equivalent rate in a free field,denoted as■,and the value of■.In contrast,in the simulated workplace neutron fields,the calibration of the neutron dose rate meter was conducted by following the ISO 12789 standard.The CF was calculated as the ratio between the values of■measured by a standard instrument(i.e.,Bonner sphere spectrometer)and the neutron dose rate meter.The CF values were obtained in the range of 0.88–1.0.The standard uncertainties■of the CFs were determined to be in the range of approximately 6.6–13.1%.
引用
收藏
页码:11 / 18
页数:8
相关论文
共 50 条
  • [31] CALIBRATION OF VARIOUS TYPES OF NEUTRON DOSIMETERS IN A HEAVY-WATER MODERATED AND MONOENERGETIC NEUTRON FIELDS .2. EXPERIMENTAL
    MATIULLAH
    KUDO, K
    YANG, X
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1993, 22 (1-4): : 687 - 690
  • [32] CALIBRATION OF VARIOUS TYPES OF NEUTRON DOSIMETERS IN A HEAVY-WATER MODERATED AND MONOENERGETIC NEUTRON FIELDS .1. THEORETICAL
    MATIULLAH
    KUDO, K
    YANG, X
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1993, 22 (1-4): : 683 - 686
  • [33] NEUTRON MODERATION SNOWFALL RATE-METER
    POWERS, CM
    WOODRUFF, GL
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1974, 18 (JUN23): : 68 - 68
  • [34] Analysis of the standard calibration methods for neutron monitors
    Campo, Xandra
    Mendez, Roberto
    Rivera, Sergio
    Gonzalez, Francisco
    RADIATION PHYSICS AND CHEMISTRY, 2024, 222
  • [35] Investigation of Neutron Survey Meter Response in Pulsed Fields
    Niven, D.
    Picard, Y.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [36] NBS NEUTRON FIELDS FOR PERSONNEL DOSIMETRY CALIBRATION
    SCHWARTZ, RB
    EISENHAUER, CM
    GILLIAM, DM
    SPIEGEL, V
    WASSON, OA
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1980, 34 (JUN): : 621 - 623
  • [38] The Design of an Ambient Neutron Dose-Equivalent Meter
    Xia Wenming
    Jia Mingchun
    Guo Zhirong
    JOURNAL OF ENGINEERING FOR GAS TURBINES AND POWER-TRANSACTIONS OF THE ASME, 2011, 133 (08): : 84501 - 1
  • [40] FIELD CALIBRATION OF A NEUTRON MOISTURE METER IN A CRACKING GRAY CLAY
    HODGSON, AS
    CHAN, KY
    IRRIGATION SCIENCE, 1987, 8 (04) : 233 - 244