A Method for Compensating Hemoglobin Interference in Total Serum Bilirubin Measurement Using a Simple Two-Wavelength Reflectance Photometer

被引:0
|
作者
Zucchini, Lorenzo [1 ]
Coda Zabetta, Carlos Daniel [2 ]
Ajcevic, Milos [1 ]
Accardo, Agostino [1 ]
机构
[1] Univ Trieste, Dept Engn & Architecture, Trieste, I-34127, Italy
[2] Bilimetrix srl, I-34149 Trieste, Italy
关键词
bilirubin; hemoglobin; neonatal hyperbilirubinemia; LMICs; SEVERE NEONATAL HYPERBILIRUBINEMIA; MANAGEMENT; HEMOLYSIS; JAUNDICE; CARE;
D O I
10.3390/s24206749
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Neonatal hyperbilirubinemia (NH) is a common condition in newborns, with elevated bilirubin levels potentially causing neurological damage or death. Accurate and timely measurements of total serum bilirubin are essential to prevent these outcomes. Direct spectrophotometry, a reliable method for measuring bilirubin, is particularly useful in constrained settings due to its potential for portable low-cost instrumentation. However, this method is susceptible to interference from hemoglobin, often present due to hemolysis. Typically, this interference is reduced using complex optical filters, reagents, multiple wavelengths, or combinations thereof, which increase costs and complexity while reducing usability. This study presents a hemoglobin compensation algorithm applied to a simple, portable, two-wavelength (465 and 590 nm) reflectance photometer designed to receive 30 mu L of plasma or whole blood samples and perform the measurement without any reagents. Testing across five bilirubin and hemoglobin levels (4.96 to 28 mg/dL and 0.06 to 0.99 g/dL, respectively) demonstrated that the algorithm effectively reduces hemoglobin interference and overestimation errors. The overall root mean square error was reduced from 4.86 to 1.45 mg/dL, while the measurement bias decreased from -4.46 to -0.10 mg/dL. This substantial reduction in overestimation errors supports future clinical trials with neonatal blood samples.
引用
收藏
页数:11
相关论文
共 10 条
  • [1] Optimization of an algorithm for hemoglobin interference compensation on a simple photometer for bilirubin measurement
    Zucchini, Lorenzo
    Zabetta, Carlos Daniel Coda
    Ajcevic, Milos
    Accardo, Agostino
    2024 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS, MEMEA 2024, 2024,
  • [3] Transcutaneous measurement (TcB) of total serum bilirubin (TSB) by multi-wavelength spectral reflectance (BillCheck™):: Accuracy and precision in newborn results
    Bhutani, V
    Johnson, LH
    Gourley, G
    Adler, S
    PEDIATRIC RESEARCH, 1999, 45 (04) : 186A - 186A
  • [4] Transcutaneous Measurement (TcB) of Total Serum Bilirubin (TSB) by Multi-Wavelength Spectral Reflectance (BiliCheck™): Accuracy and Precision in Newborn Babies
    V Bhutani
    L H Johnson
    G Gourley
    S Adler
    Pediatric Research, 1999, 45 (7) : 186 - 186
  • [5] Non-Invasive Measurement of Total Serum Bilirubin by Multi-Wavelength Spectral Reflectance by Bili Check (TM) in Newborn Patients † 968
    V Bhutani
    L Johnson
    G Gourley
    R Dworanczyk
    M Grous
    Pediatric Research, 1998, 43 (Suppl 4) : 167 - 167
  • [6] Strong interference of hemoglobin concentration on CSF total protein measurement using the trichloroacetic acid precipitation method
    Boer, Klas
    Isenmann, Stefan
    Deufel, Thomas
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2007, 45 (01) : 112 - 113
  • [7] Optical fiber-based full Mueller polarimeter for endoscopic imaging using a two-wavelength simultaneous measurement method
    Vizet, Jeremy
    Manhas, Sandeep
    Tran, Jacqueline
    Validire, Pierre
    Benali, Abdelali
    Garcia-Caurel, Enric
    Pierangelo, Angelo
    De Martino, Antonello
    Pagnoux, Dominique
    JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (07)
  • [8] Analysis Method Using Two-Wavelength Mach-Zehnder Interferometer for the Measurement of Soret Coefficients in Soret-Facet Mission on ISS
    Tomaru, Momoko
    Osada, Takuma
    Orikasa, Isamu
    Suzuki, Shinsuke
    Inatomi, Yuko
    MICROGRAVITY SCIENCE AND TECHNOLOGY, 2019, 31 (01) : 49 - 59
  • [9] Analysis Method Using Two-Wavelength Mach-Zehnder Interferometer for the Measurement of Soret Coefficients in Soret-Facet Mission on ISS
    Momoko Tomaru
    Takuma Osada
    Isamu Orikasa
    Shinsuke Suzuki
    Yuko Inatomi
    Microgravity Science and Technology, 2019, 31 : 49 - 59
  • [10] Simple and sensitive method for the quantification of total bilirubin in human serum using 3-methyl-2-benzothiazolinone hydrazone hydrochloride as a chromogenic probe
    Nagaraja, Padmarajaiah
    Avinash, Krishnegowda
    Shivakumar, Anantharaman
    Dinesh, Rangappa
    Shrestha, Ashwinee Kumar
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2010, 77 (04) : 782 - 786