Vespa: Logic-Level Constraint-Based Validation for Continuous-Flow Microfluidic Devices

被引:0
|
作者
Zhou, Yangruirui [1 ]
Oliveira, Samuel M. D.
Sanka, Radhakrishna
Mcintyre, David [2 ]
Densmore, Douglas [3 ,4 ]
机构
[1] Boston Univ, Elect & Comp Engn Dept, Boston, MA 02215 USA
[2] Boston Univ, Biomed Engn Dept, Boston, MA 02215 USA
[3] Boston Univ, Elect & Comp Engn Dept, Boston, MA 02215 USA
[4] Boston Univ, Biomed Engn Dept, Boston, MA 02215 USA
基金
美国国家科学基金会;
关键词
Fluids; Fabrication; Benchmark testing; Microfluidics; Biochips; Logic; Transportation; Biochip; computer-aided design; control logic; microfluidics; pathfinding algorithms; validation; BIOCHIPS; SYSTEM; VALVES; CELLS; PUMPS;
D O I
10.1109/TCAD.2024.3435706
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
Continuous-flow microfluidic devices (CFMDs), also known as biochips provide automated and cost-effective solutions for the biomolecular analysis in life science applications. Multioperation biochemical reactions require incorporating microfluidic components into CFMDs, which may raise their design complexity and lead to design-objective-constraint compatibility (DOCC) conflicts. Existing validation methods excel in optimizing control-layer pressurization protocol (CLPP) for small networks. However, they face challenges in describing operation constraints for the microfluidic large-scale integration (mLSI) experiments and addressing the undesired flow risk. To overcome these challenges, we develop Vespa, an open-source validation framework that uses logic expressions to describe operation constraints with lower human labor costs. For each operation, taking an objective list, a constraint list, a design netlist, and a super-parameter as inputs, Vespa validates the DOCC by constructing a target fluid transportation path. In addition, it introduces an undesired flow risk mitigation mechanism, issuing warnings for incorrect constraints and potential fluid undesired flow risks. The work includes over 900 real-world and synthetic benchmarks in three complexity ranges to support researchers in the field. As a result, more than 85% of benchmark experiments yield correct results within 0.3 s, enabling Vespa for real-time validation and integration into interactive CFMD design tools. Finally, to show Vespa's practical efficacy and real-world impact, we create case studies using a real-world CFMD. We demonstrate that Vespa eliminates over 90% of DOCC-related wet lab tests by detecting issues and making updates before fabrication.
引用
收藏
页码:483 / 496
页数:14
相关论文
共 47 条
  • [21] Continuous-flow in-droplet magnetic particle separation in a droplet-based microfluidic platform
    Lee, Hun
    Xu, Linfeng
    Ahn, Byungwook
    Lee, Kangsun
    Oh, Kwang W.
    MICROFLUIDICS AND NANOFLUIDICS, 2012, 13 (04) : 613 - 623
  • [22] Microsphere based continuous-flow immunoassay in a microfluidic device for determination of clinically relevant insulin levels
    Cohen, Noa
    Sabhachandani, Pooja
    Sarkar, Saheli
    Kahanovitz, Lindy
    Lautsch, Norman
    Russell, Steven J.
    Konry, Tania
    MICROCHIMICA ACTA, 2017, 184 (03) : 835 - 841
  • [23] Continuous-flow in-droplet magnetic particle separation in a droplet-based microfluidic platform
    Hun Lee
    Linfeng Xu
    Byungwook Ahn
    Kangsun Lee
    Kwang W. Oh
    Microfluidics and Nanofluidics, 2012, 13 : 613 - 623
  • [24] Microsphere based continuous-flow immunoassay in a microfluidic device for determination of clinically relevant insulin levels
    Noa Cohen
    Pooja Sabhachandani
    Saheli Sarkar
    Lindy Kahanovitz
    Norman Lautsch
    Steven J. Russell
    Tania Konry
    Microchimica Acta, 2017, 184 : 835 - 841
  • [25] Sequence-Pair-Based Flow-Layer Physical Design Algorithm for Continuous-Flow Microfluidic Biochips
    Zhu Y.
    Huang H.
    Lin H.
    Chen W.
    Liu G.
    Xu N.
    Huang X.
    Jisuanji Fuzhu Sheji Yu Tuxingxue Xuebao/Journal of Computer-Aided Design and Computer Graphics, 2022, 34 (04): : 535 - 544
  • [26] Anomaly Detection Method based on Discrete Particle Swarm Optimization for Continuous-Flow Microfluidic Biochips
    Wu, Yangjie
    Zhu, Yuhan
    Liu, Genggeng
    Huang, Xing
    PROCEEDING OF THE GREAT LAKES SYMPOSIUM ON VLSI 2024, GLSVLSI 2024, 2024, : 507 - 510
  • [27] Continuous-flow pl-based sorting of proteins and peptides in a microfluidic chip using diffusion potential
    Song, Yong-Ak
    Hsu, Stephanie
    Stevens, Anna L.
    Han, Jongyoon
    ANALYTICAL CHEMISTRY, 2006, 78 (11) : 3528 - 3536
  • [28] Magnetic Bead Chain-Based Continuous-Flow DNA Extraction for Microfluidic PCR Detection of Salmonella
    Wang, Yuhe
    Qi, Wuzhen
    Wang, Lei
    Lin, Jianhan
    Liu, Yuanjie
    MICROMACHINES, 2021, 12 (04)
  • [29] COMSOL-Based Design and Validation of Dilution Algorithm with Continuous-Flow Lab-on-Chip
    Tapalina Banerjee
    Sarmishtha Ghoshal
    Bhargab B. Bhattacharya
    INAE Letters, 2017, 2 (2): : 55 - 63
  • [30] A PROSPECTIVE VALIDATION OF THE FIRST ENDOSCOPIC MANAGEMENT ALGORITHM FOR GASTROINTESTINAL BLEEDING IN PATIENTS WITH CONTINUOUS-FLOW LEFT VENTRICULAR ASSIST DEVICES
    Axelrad, Jordan
    Pinsino, Alberto
    Trinh, Pauline
    Thanataveerat, Anusorn
    Ramirez, Ivonne
    Garcia-Carrasquillo, Reuben J.
    Colombo, Paolo
    Yuzefpolskaya, Melana
    Gonda, Tamas A.
    GASTROINTESTINAL ENDOSCOPY, 2018, 87 (06) : AB299 - AB300