Simulated design optimization of a prototype solid tank optical CT scanner for 3D radiation dosimetry

被引:0
|
作者
Ogilvy, A. [1 ]
Collins, S. [1 ]
Hare, W. [1 ]
Hilts, M. [1 ,2 ]
Tuokko, T. [1 ]
Deardon, R. [3 ,4 ]
Jirasek, A. [1 ]
机构
[1] Univ British Columbia, Dept Phys, Okanagan Campus, Kelowna, BC V1V 1V7, Canada
[2] BC Canc Kelowna, Med Phys, Kelowna, BC V1Y 5L3, Canada
[3] Univ Calgary, Dept Math & Stat, Calgary, AB T2N 1N4, Canada
[4] Univ Calgary, Fac Vet Med, Calgary, AB T2N 1N4, Canada
关键词
D O I
10.1088/1742-6596/2167/1/012009
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Optical computed tomography (CT) is one of the leading modalities for imaging gel dosimeters. There exist many prototype designs, as well as some commercial optical CT scanners that have showcased the value that gel dosimeters can provide to improve 3D dose verification for radiation treatments. However, due to factors including image accuracy, scan time, or demanding setup and maintenance there is currently no single scanner that has become a ubiquitous staple in a clinical setting. In this work, a prototype solid tank optical CT scanner is proposed that minimizes the need for a refractive index bath commonly found in optical CT systems. In addition to the design proposal, a ray-path simulator was created to optimize the design such that the solid tank geometry improves light collection across the detector array, maximizes the volume of the dosimeter scanned, and maximizes the dynamic range of the scanner.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Optimization of solid tank design for fan-beam optical CT based 3D radiation dosimetry
    Ogilvy, A.
    Collins, S.
    Tuokko, T.
    Hilts, M.
    Deardon, R.
    Hare, W.
    Jirasek, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (24):
  • [2] A prototype fan-beam optical CT scanner for 3D dosimetry
    Campbell, Warren G.
    Rudko, D. A.
    Braam, Nicolas A.
    Wells, Derek M.
    Jirasek, Andrew
    MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Commissioning of a solid tank design for fan-beam optical CT based 3D radiation dosimetry
    Ogilvy, A.
    Collins, S.
    Hilts, M.
    Hare, W.
    Jirasek, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (17):
  • [4] Cone beam optical CT scanner for 3D dosimetry
    Miller, J
    Adamovics, J
    Dietrich, J
    MEDICAL PHYSICS, 2005, 32 (06) : 2138 - 2138
  • [5] Optical CT scanner for in-air readout of gels for external radiation beam 3D dosimetry
    Ramm, Daniel
    Rutten, Thomas P.
    Shepherd, Justin
    Bezak, Eva
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12): : 3853 - 3868
  • [6] Performance of an improved first generation optical CT scanner for 3D dosimetry
    Qian, Xin
    Adamovics, John
    Wuu, Cheng-Shie
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (24): : N321 - N331
  • [7] Improvement of Optical CT Scanner for 3-D Dosimetry
    Qian, X.
    Admovics, J.
    Wuu, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Simulation of prototype optical CT scanner design
    Ogilvy, Andy
    Jirasek, Andrew
    MEDICAL PHYSICS, 2019, 46 (11) : 5393 - 5393
  • [9] Accuracy Comparison of the Optical 3D Scanner and CT Scanner
    Mendricky, Radomir
    Sobotka, Jiri
    MANUFACTURING TECHNOLOGY, 2020, 20 (06): : 791 - 801
  • [10] Optical Scanner for 3D Radiotherapy Polymer Gel Dosimetry
    Sobotka, P.
    Kozicki, M.
    Maras, P.
    Boniecki, L.
    Kacperski, K.
    Domanski, A. W.
    ACTA PHYSICA POLONICA A, 2012, 122 (05) : 969 - 974