Commissioning a clinical proton pencil beam scanning beamline for pre-clinical ultra-high dose rate irradiations on a cyclotron-based system

被引:0
|
作者
Saini, Jatinder [1 ,2 ]
Erickson, Danielle P. Johnson [1 ]
Stappen, Francois Vander [3 ]
Ruth, Matt [1 ,2 ]
Cui, Sunan [1 ,2 ]
Gorman, Vanessa [3 ]
Rossomme, Severine [3 ]
Cao, Ning [1 ,2 ]
Ford, Eric C. [1 ,2 ]
Meyer, Juergen [1 ,2 ]
Bloch, Charles [1 ,2 ]
Wong, Tony [1 ,2 ]
Grassberger, Clemens [1 ]
Rengan, Ramesh [1 ,2 ]
Zeng, Jing [1 ,2 ]
Schwarz, Marco [1 ,2 ]
机构
[1] Univ Washington, Sch Med, Dept Radiat Oncol, Seattle, WA 98195 USA
[2] Fred Hutchinson Canc Ctr, Radiat Oncol, Seattle, WA 98109 USA
[3] Proton Therapy Res & Dev, Ion Beam Applicat, Louvain La Neuve, Belgium
来源
FRONTIERS IN ONCOLOGY | 2024年 / 14卷
关键词
ultra-high dose rate; FLASH; protons; pencil beam scanning; pre-clinical;
D O I
10.3389/fonc.2024.1460288
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background This manuscript describes modifications to a pencil beam scanning (PBS) proton gantry that enables ultra-high dose rates (UHDR) irradiation, including treatment planning and validation. Methods Beamline modifications consisted of opening the energy slits and setting the degrader to pass-through mode to maximize the dose rate. A range shifter was inserted upstream from the isocenter to enlarge the spot size and make it rotationally symmetric. We measured the beamline transport efficiency and investigated the variation in output due to the recombination of charge in the dose monitoring chamber. The output calibration was performed through a parallel plate chamber (PPC05), and an intercomparison was performed for various detectors. The pre-clinical field for mice irradiation consisted of different dose levels to deliver uniform doses in transmission mode. The field dose rates were determined through log files while scripting in TPS was used to estimate PBS dose rates. The survival experiments consisted of irradiating the full pelvis of the mice at UHDR and conventional dose rates. Results The spot size was constant with beam current and had a sigma of 8.5 mm at the isocenter. The beam output increased by 35% at 720 nA compared to 5.6 nA, primarily due to recombination in the dose-monitoring ion chambers. The Faraday Cup and PPC05 agreed within 2%, while other detectors were within 3% of FC for dose rates <60 Gy/s. The pre-clinical fields' PBS dose rate is above 45 Gy/sec for all voxels within the target volume. The average and PBS dose rates decrease as field size increases and approaches 40 Gy/s for a field size of 7x7 cm2. All UHDR arms showed better survival than the corresponding conventional dose rate arms. Conclusions We successfully modified a clinical system to perform UHDR pre-clinical experiments. As part of our pre-clinical experiments, we observed the FLASH effect concerning mice survival.
引用
收藏
页数:13
相关论文
共 25 条
  • [21] Long-Term Clinical Safety of High-Dose Proton Radiation Therapy Delivered With Pencil Beam Scanning Technique for Extracranial Chordomas and Chondrosarcomas in Adult Patients: Clinical Evidence of Spinal Cord Tolerance
    Stieb, Sonja
    Snider, James W., III
    Placidi, Lorenzo
    Kliebsch, Ulrike
    Lomax, Anthony J.
    Schneider, Ralf A.
    Weber, Damien C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 100 (01): : 218 - 225
  • [22] Pencil-Beam Versus Monte Carlo Based Dose Calculation for Proton Therapy Patients with Complex Geometries. Clinical Use of the TOPAS Monte Carlo System
    Schuemann, J.
    Shin, J.
    Perl, J.
    Grassberger, C.
    Verburg, J.
    Faddegon, B.
    Paganetti, H.
    MEDICAL PHYSICS, 2012, 39 (06) : 3820 - 3820
  • [23] Log file-based dose reconstruction and accumulation for 4D adaptive pencil beam scanned proton therapy in a clinical treatment planning system: Implementation and proof-of-concept
    Meijers, A.
    Jakobi, A.
    Stuetzer, K.
    Guterres Marmitt, G.
    Both, S.
    Langendijk, J. A.
    Richter, C.
    Knopf, A.
    MEDICAL PHYSICS, 2019, 46 (03) : 1140 - 1149
  • [24] Ultra-High Dose-Rate Particle Therapy: On the Delivery of Active Beam Scanning at a Synchrotron-Based Facility With p, 4He, 12C and 16O
    Tessonnier, T.
    Mein, S.
    Brons, S.
    Haberer, T.
    Mairani, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E517 - E517
  • [25] Pre-Clinical Evaluation of a Novel Thin Strut (85 mm) Ultra-High Molecular Weight PLLA Sirolimus-Eluting Bioresorbable Scaffold: A Comparative Multi-Modality Imaging-Based Study
    Ferrone, Marco
    Melnick, Gilberto
    Isaza, Nicolas
    Cheng, Yanping
    McGregor, Jenn
    Ramzipoor, Kamal
    Lee, Chang
    Conditt, Gerard
    Kaluza, Grzegorz
    Granada, Juan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 72 (13) : B157 - B157