Parts-per-Billion Detection of Hydrogen Sulfide via Cavity Ring-Down Spectroscopy

被引:0
|
作者
Xu, Wei [1 ]
Wang, Xuejun [1 ]
Zhao, Lei [1 ]
Zou, Jun [1 ]
Chen, Bing [2 ]
机构
[1] Wuhan Second Ship Design and Research Institute, Wuhan,430205, China
[2] Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Sciences, Chinese Academy of Sciences, Hefei,230031, China
基金
中国国家自然科学基金;
关键词
Absorption spectroscopy - Atomic spectroscopy - Chemical sensors - Gas detectors - Laser mode locking - Laser safety - Spectrum analyzers;
D O I
10.3390/photonics12030284
中图分类号
学科分类号
摘要
Rapid and precise detection of hydrogen sulfide (H2S) at trace levels is critical for industrial safety and environmental air quality monitoring, yet existing methods often struggle with cost, speed, or sensitivity. A cost-effective cavity ring-down spectroscopy (CRDS) analyzer is presented, incorporating a novel digital locking circuit for sequential laser-cavity mode matching. This system demonstrates rapid and precise hydrogen sulfide (H2S) detection capability at parts-per-billion (ppb) concentration levels. Compared to traditional wavelength meters, our system delivers a 140-fold improvement in frequency interval precision (0.07 MHz, 0.027% relative uncertainty). Allan variance analysis under vacuum conditions demonstrates a sensitivity limit of 3 × 10−12 cm−1 at a 60-s averaging time. Validated through calibrated gas dilution tests, the analyzer detects a 4 ppb H2S absorption signal with a signal-to-noise ratio (SNR) > 6, establishing a 2 ppb detection limit (3σ criterion). This innovative approach meets stringent industrial and environmental requirements, offering a significant advancement in trace gas-sensing technology. © 2025 by the authors.
引用
收藏
相关论文
共 50 条
  • [1] Investigation of hydrogen sulfide adsorption using cavity ring-down spectroscopy
    Xiong, Feng
    Peng, Zhimin
    Ding, Yanjun
    Lyu, Junfu
    Du, Yanjun
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2025, 36 (04)
  • [2] Cavity ring-down spectroscopy sensor for detection of hydrogen chloride
    Hagen, C. L.
    Lee, B. C.
    Franka, I. S.
    Rath, J. L.
    VandenBoer, T. C.
    Roberts, J. M.
    Brown, S. S.
    Yalin, A. P.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2014, 7 (02) : 345 - 357
  • [3] DETECTION OF ATMOSPHERIC POLLUTANTS AT PARTS-PER-BILLION LEVELS BY INFRARED SPECTROSCOPY
    HANST, PL
    LEFOHN, AS
    GAY, BW
    APPLIED SPECTROSCOPY, 1973, 27 (03) : 188 - 198
  • [4] DETERMINATION OF SULFIDE AT THE PARTS-PER-BILLION LEVEL BY ION CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION
    HAN, K
    KOCH, WF
    ANALYTICAL CHEMISTRY, 1987, 59 (07) : 1016 - 1020
  • [5] Cavity ring-down spectroscopy
    Zare, RN
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1998, 215 : U83 - U83
  • [6] Cavity ring-down spectroscopy
    Wheeler, MD
    Newman, SM
    Orr-Ewing, AJ
    Ashfold, MNR
    JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1998, 94 (03): : 337 - 351
  • [7] MEMS based hydrogen sensing with parts-per-billion resolution
    Gurusamy, J. T.
    Putrino, Gino
    Jeffery, Roger D.
    Silva, K. K. M. B. Dilusha
    Martyniuk, Mariusz
    Keating, Adrian
    Faraone, Lorenzo
    SENSORS AND ACTUATORS B-CHEMICAL, 2019, 281 : 335 - 342
  • [8] Detection of Sulfur Dioxide by Cavity Ring-Down Spectroscopy
    Medina, David S.
    Liu, Yingdi
    Wang, Liming
    Zhang, Jingsong
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2011, 45 (05) : 1926 - 1931
  • [9] Detection of toxic compounds by cavity ring-down spectroscopy
    Research and Technology Park, Mississippi Stt. Univ. Diagn. I., 205 Research Boulevard, Starkville, MS 39759-9734, United States
    Environ. Sci. Technol., 11 (1936-1939):
  • [10] Detection of sputtered metals with cavity ring-down spectroscopy
    Yalin, AP
    Surla, V
    Butweiller, M
    Williams, JD
    APPLIED OPTICS, 2005, 44 (30) : 6496 - 6505