Central venous pressure as a method of optimising atrio-ventricular delay after cardiac surgery

被引:0
|
作者
Tindale, Alexander [1 ,2 ]
Cretu, Ioana [3 ]
Gomez, Naomi [1 ]
Haynes, Ross [1 ]
Meng, Hongying [3 ]
Mason, Mark J. [1 ,3 ]
Francis, Darrel P. [2 ]
机构
[1] Guys & St Thomas Fdn Trust, Harefield Hosp, Dept Cardiol, London, England
[2] Imperial Coll London, Natl Heart & Lung Inst, London, England
[3] Brunel Univ London, Uxbridge, England
来源
PLOS ONE | 2025年 / 20卷 / 01期
关键词
RESYNCHRONIZATION THERAPY; AV DELAY; OPTIMIZATION;
D O I
10.1371/journal.pone.0310905
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Introduction Haemodynamic atrioventricular delay (AVD) optimisation has primarily focussed on signals that are not easy to acquire from a pacing system itself, such as invasive left ventricular catheterisation or arterial blood pressure (ABP). In this study, standard clinical central venous pressure (CVP) signals are tested as a potential alternative. Methods Sixteen patients with a temporary pacemaker after cardiac surgery were studied. AV delay optimisation was performed by alternating between a reference AVD of 120ms and tested settings ranging from 40 to 280ms, with 8 replicates for each setting. Alongside (a) the raw data, three methods of correcting for respiration were tested: (b) limiting analysis to a respiratory cycle, (c) asymmetric least squares (ALS) and (d) discrete wavelet transform (DWT). The utility of a quality control step was tested. Results CVP signals were a mirror image of the systolic ABP signals: The four R values were -0.674, -0.692, -0.631, -0.671 respectively (all p<0.001). With quality control, the mirror image was best for DWT (R = -0.76, p<0.001), with the CVP and ABP optima agreeing well (R = 0.78, p<0.001). The automated quality control signal correctly predicted the gap between the AVD optima calculated from ABP and CVP (R = 0.8, p<0.001). Conclusions Central venous pressure signals could be used to optimise AVD, because they have a reliable inverse relationship with ABP when pacemaker settings undergo protocolised testing. However, protocols need careful design to circumvent spontaneous biological variability.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Optimizing atrio-ventricular delay in pacemakers using potentially implantable physiological biomarkers
    Keene, Daniel
    Miyazawa, Alejandra A.
    Johal, Monika
    Arnold, Ahran D.
    Ali, Nadine
    Saqi, Khulat A.
    March, Katherine
    Burden, Leah
    Francis, Darrel P.
    Whinnett, Zachary, I
    Shun-Shin, Matthew J.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2022, 45 (04): : 461 - 470
  • [22] Left ventricular flow redirection towards the outflow tract: a marker for atrio-ventricular delay optimization
    Rodriguez Munoz, D.
    Lozano Granero, C.
    Carbonell San Roman, A.
    Moya Mur, J. L.
    Fernandez-Golfin, C.
    Casas Rojo, E.
    Garcia Martin, A.
    Fernandez Santos, S.
    Moreno Planas, J.
    Zamorano Gomez, J. L.
    EUROPEAN HEART JOURNAL, 2015, 36 : 945 - 946
  • [23] Does Atrio-Ventricular synchrony during ventricular tachycardia improve blood pressure response?
    Taneja, T
    Kadish, A
    Goldberger, J
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 119A - 120A
  • [24] USE OF CENTRAL VENOUS PRESSURE IN CARDIAC SURGERY
    SHAPIRO, G
    BAILEY, CP
    MILOWSKY, J
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1966, S : 183 - +
  • [25] Distribution of the device based intrinsic atrio-ventricular conduction delay in a broad pacemaker population with sinus node dysfunction or atrio -ventricular block
    Stockburger, M.
    Jauvert, G.
    Hatala, R.
    Butter, C.
    Bode, F.
    Ranalli, G.
    Rolando, M.
    Moreno, J.
    EUROPEAN HEART JOURNAL, 2012, 33 : 996 - 996
  • [26] Atrio-ventricular valve regurgitation in univentricular hearts: outcomes after repair
    Laux, Daniela
    Vergnat, Mathieu
    Lambert, Virginie
    Gouton, Marielle
    Ly, Mohamed
    Peyre, Marianne
    Roussin, Regine
    Belli, Emre
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2015, 20 (05) : 622 - 630
  • [27] Optimal atrio-ventricular delay in CRT patients can be approximated using surface ECG
    Jones, Chris
    Levin, Vadim
    Rubin, Andrew
    Phang, Robert
    Svinarich, Tom
    Murillo, Jaime
    Sambelashvili, Alex
    JOURNAL OF CARDIAC FAILURE, 2008, 14 (06) : S63 - S64
  • [28] Histopathology and prognosis of new-onset atrio-ventricular block after coronary artery bypass graft surgery
    Toiviainen, A.
    Relander, A.
    Lahtinen, V
    Ruohonen, I
    Lehto, J.
    Vasankari, T.
    Gunn, J.
    Hollmen, M.
    Kiviniemi, T.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [29] Outcomes and reoperations after total correction of complete atrio-ventricular septal defect
    Dodge-Khatami, Ali
    Herger, Stefan
    Rousson, Valentin
    Comber, Maurice
    Knirsch, Walter
    Bauersfeld, Urs
    Pretre, Rene
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 34 (04) : 745 - 750
  • [30] Does correlation exist between invasive and non-invasive measurements for optimal atrio-ventricular delay in cardiac resynchronization therapy?
    Wjasow, C
    Abramson, S
    Heimann, M
    Britton, N
    Rothman, S
    Kocovic, D
    JOURNAL OF CARDIAC FAILURE, 2005, 11 (06) : S142 - S142