A stuck leaflet after balloon post-dilatation in transcatheter aortic valve implantation with a SAPIEN-3 ultra RESILIA valve: a case report

被引:0
|
作者
Yamazoe, Shinji [1 ]
Ogawa, Yasuhiro [1 ]
Kano, Naoaki [1 ]
Mamiya, Keita [1 ]
Kawaguchi, Katsuhiro [1 ]
机构
[1] Komaki City Hosp, Dept Cardiol, 1-20 Joubushi, Komaki, Aichi 4858520, Japan
关键词
Transcatheter aortic valve implantation; SAPIEN 3 Ultra RESILIA; Stuck leaflet; Transoesophageal echocardiography; Case Report; PARAVALVULAR REGURGITATION; REPLACEMENT; IMPACT; MORTALITY;
D O I
10.1093/ehjcr/ytae697
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Transcatheter aortic valve implantation (TAVI) is a safe and effective therapy for patients with severe aortic stenosis. A Stuck leaflet and severe intraprosthetic regurgitation after valve implantation occur rarely but can lead to sudden haemodynamic deterioration. We encountered a case of a stuck leaflet following post-dilatation with the Edwards Sapien 3 Ultra RESILIA valve.Case summary A 72-year-old woman was referred to our hospital for severe aortic stenosis with shortness of breath. She underwent transfemoral TAVI. After deployment of a 23 mm Sapien 3 Ultra RESILIA valve, post-dilatation was performed due to the presence of paravalvular leak (PVL). Transoesophageal echocardiography revealed a stuck leaflet and severe intraprosthetic regurgitation. Aortography also demonstrated severe aortic regurgitation. We performed valve-in-valve procedure using the second 23 mm valve. Post-valve-in-valve transoesophageal echocardiography showed no PVL nor aortic regurgitation, and haemodynamics improved.Discussion A stuck leaflet is a rare complication following post-dilatation. Severe intraprosthetic regurgitation can lead to sudden haemodynamic changes and may, in some cases, necessitate the use of extracorporeal membrane oxygenation. If haemodynamic changes occur, it is essential to promptly investigate the cause through multiple diagnostic modalities, including transoesophageal echocardiography and angiography.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Cardiogenic shock following balloon post-dilatation in transcatheter aortic valve implantation: first case report of all three stuck leaflets (vol 2, yty073, 2018)
    Mehrani, Mehdi
    Nematollahi, Alireza
    Hatami, Motahare
    Hosseini, Kaveh
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2019, 3 (02)
  • [22] Hemodynamic Outcomes of Post-dilatation in Transcatheter Aortic Valve Replacement
    Forster, Molly
    Khalique, Omar K.
    Kodali, Susheel
    Williams, Mathew
    George, Isaac
    Nazif, Tamim
    Paradis, Jean-Michel
    Grubb, Kendra J.
    Leon, Martin
    Hahn, Rebecca
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : B216 - B217
  • [23] Stuck leaflet after transcatheter aortic valve implantation due to calcium chunk in a patient with bicuspid valve
    Kawamura, Hiroto
    Yoshizaki, Toru
    Sato, Akira
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2024, 8 (06)
  • [24] Conduction disorders after transcatheter aortic valve implantation: A comparison between SAPIEN 3 and SAPIEN 3 Ultra balloon-expandable valves
    Monizzi, Giovanni
    Olivares, Paolo
    Makmur, Giulio
    Fabbiocchi, Franco
    Grancini, Luca
    Mastrangelo, Angelo
    Ferrari, Cristina
    Galli, Stefano
    Montorsi, Piero
    Bartorelli, Antonio L. L.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [25] Aortic regurgitation after transcatheter aortic valve implantation of the edwards SAPIEN™ valve
    Nielsen, Hans Henrik Moller
    Egeblad, Henrik
    Andersen, Henning Rud
    Thuesen, Leif
    Poulsen, Steen Hvitfeldt
    Klaaborg, Kaj-Erik
    Jakobsen, Carl-Johan
    Hjortdal, Vibeke Elisabeth
    SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2013, 47 (01) : 36 - 41
  • [26] Transcatheter aortic valve-in-valve post-dilatation as an overlooked risk factor of delayed coronary obstruction: a case report
    Marchese, Alfredo
    Tarantini, Giuseppe
    Tito, Antonio
    Colombo, Antonio
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (05)
  • [27] Leaflet thrombosis after valve-in-valve transcatheter aortic valve implantation: a case series
    Aktuerk, Dincer
    Mirsadraee, Saeed
    Quarto, Cesare
    Davies, Simon
    Duncan, Alison
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (04) : 1 - 6
  • [28] Balloon Post-Dilatation Improves Long-Term Valve Performance After Balloon-Expandable Valve Implantation
    Nara, Yugo
    Watanabe, Yusuke
    Kataoka, Akihisa
    Nakashima, Makoto
    Hioki, Hirofumi
    Kawashima, Hideyuki
    Nagura, Fukuko
    Kozuma, Ken
    Yashima, Fumiaki
    Shirai, Shinichi
    Tada, Norio
    Yamawaki, Masahiro
    Naganuma, Toru
    Yamanaka, Futoshi
    Ueno, Hiroshi
    Tabata, Minoru
    Mizutani, Kazuki
    Takagi, Kensuke
    Yamamoto, Masanori
    Hayashida, Kentaro
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2022, 37 : 15 - 22
  • [29] Impact of New Pacing Requirement on Echocardiographic Outcomes After Transcatheter Aortic Valve Replacement With Sapien-3 Valve
    Sammour, Yasser
    Gajulapalli, Rama D.
    Lak, Hassan Mehmood
    Chawla, Sanchit
    Kumar, Arnav
    Sato, Kimi
    Banerjee, Kinjal
    Svensson, Lars G.
    Yun, James J.
    Popovic, Zoran B.
    Tarakji, Khaldoun
    Wazni, Oussama M.
    Reed, Grant W.
    Puri, Rishi
    Krishnaswamy, Amar
    Kapadia, Samir R.
    CIRCULATION, 2020, 142
  • [30] Is Post-dilatation Useful After Implantation of the Edwards Valve?
    Watanabe, Yusuke
    Hayashida, Kentaro
    Lefevre, Thierry
    Romano, Mauro
    Morice, Marie-Claude
    AMERICAN JOURNAL OF CARDIOLOGY, 2013, 111 (07): : 56B - 57B