Mandible bone mineral density estimation using spectral panoramic X-ray imaging

被引:0
|
作者
Somerkivi, Villeseveri [1 ,2 ,3 ]
Sellerer, Thorsten [1 ,2 ]
Berthe, Daniel [1 ,2 ]
Haemisch, York [4 ]
Pantsar, Tuomas [4 ]
Lohman, Henrik [4 ]
Kaasalainen, Touko [5 ,6 ]
Pfeiffer, Franz [1 ,2 ,7 ,8 ]
机构
[1] Tech Univ Munich, Sch Nat Sci, Dept Phys, James Franck Str 1, D-85748 Garching, Germany
[2] Tech Univ Munich, Munich Inst Biomed Engn, Garching, Germany
[3] Planmeca Oy, Helsinki, Finland
[4] Varex Imaging Corp, Salt Lake City, UT USA
[5] Univ Helsinki, HUS Diag Ctr, Helsinki, Finland
[6] Helsinki Univ Hosp, Helsinki, Finland
[7] Tech Univ Munich, Dept Diag & Intervent Radiol, Klinikum Rechts Isar, Munich, Germany
[8] Tech Univ Munich, Inst Adv Study, Garching, Germany
关键词
Radiography; Dual-Energy Scanned Projection; X-Rays; Bone Density; OSTEOPOROSIS; SKELETAL;
D O I
10.5624/isd.20240231
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study demonstrated the feasibility of obtaining mandible bone mineral density (BMD) scores using spectral panoramic imaging. Materials and Methods: Areal BMD scores were measured from the body and angle of the mandible in 3 anthropomorphic head phantoms using a spectral panoramic system (Planmeca Promax Mid, Planmeca Oy, Helsinki, Finland) equipped with a DC-Vela detector (Varex Imaging Corporation, Salt Lake City, USA). These results were compared to synthetic panoramic images generated from dual-energy CT acquisitions. Reproducibility was evaluated by repeatedly scanning 1 phantom with minor patient positioning errors, and the linearity of the BMD scores was assessed using calcium inserts in a Gammex 472 phantom (Sun Nuclear, Melbourne, USA). Results: The experimental and synthetic panoramic images appeared visually similar. The mean synthetic score was 0.640 g/cm2, and the anthropomorphic phantoms produced a root mean squared error of 0.0292 g/cm2 with a correlation coefficient of 0.969. Typical patient positioning errors did not substantially increase the error, which measured 0.0296 g/cm2 and 0.0474 g/cm2 for the left and right sides, respectively. Linearity tests using the Gammex phantom yielded a correlation coefficient of 0.998 for BMD scores ranging from 0.03 to 2.7 g/cm2. Conclusion: The BMD data obtained from spectral panoramic imaging are consistent with both dual-energy CT and Gammex phantom measurements. Consequently, spectral panoramic imaging shows potential as a method for osteoporosis screening, leveraging the widespread use of panoramic imaging. (Imaging Sci Dent 20240231)
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Quantitative evaluation of bone-mineral density loss using X-ray coherent scattering
    Barroso, Regina Cely
    Oliveira, Luis Fernando
    Castro, Carlos Roberto Ferreira
    Lima, Joao Carlos
    Braz, Delson
    Lopes, Ricardo Tadeu
    Droppa, Roosevel
    Tromba, Giuliana
    Mancini, Lucia
    Zanini, Franco
    Rigon, Luigi
    Dreossi, Diego
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01): : 318 - 321
  • [32] Bone mineral density measurements in otosclerosis using dual-energy X-ray absorptiometry
    Tonuk, S. B.
    Koybasi, S.
    Bicer, Y. O.
    Alkan Temel, S.
    CLINICAL OTOLARYNGOLOGY, 2016, 41 (02) : 149 - 153
  • [33] Comparison between quantitative X-ray imaging, dual energy X-ray absorptiometry and microCT in the assessment of bone mineral density in disuse-induced bone loss
    Mabilleau, G.
    Mieczkowska, A.
    Libouban, H.
    Simon, Y.
    Audran, M.
    Chappard, D.
    JOURNAL OF MUSCULOSKELETAL & NEURONAL INTERACTIONS, 2015, 15 (01) : 42 - 52
  • [34] Calcaneus-bone mineral density using single X-ray absorptiometry in Japanese women
    Sekioka, Y
    Kushida, K
    Yamazaki, K
    Inoue, T
    CALCIFIED TISSUE INTERNATIONAL, 1999, 65 (02) : 106 - 111
  • [35] Bone Mineral Density Screening System Using CMOS-Sensor X-ray Detector
    Maneerat, Areerat
    Visitsattapongse, Sarinporn
    Pintavirooj, Chuchart
    SENSORS, 2021, 21 (21)
  • [36] An image processing algorithm to estimate bone mineral density using digital X-ray images
    Khan, Shiyana Sherief
    Jayan, Archana S.
    Nageswaran, Sharmila
    PROCEEDINGS OF THE 2017 IEEE SECOND INTERNATIONAL CONFERENCE ON ELECTRICAL, COMPUTER AND COMMUNICATION TECHNOLOGIES (ICECCT), 2017,
  • [37] Bone Cartilage Imaging with X-ray Interferometry using a Practical X-ray Tube
    Kido, Kazuhiro
    Makifuchi, Chiho
    Kiyohara, Junko
    Itou, Tsukasa
    Honda, Chika
    Momose, Atsushi
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [38] Deep learning for automatic mandible segmentation on dental panoramic x-ray images
    Machado, Leonardo Ferreira
    Watanabe, Plauto Christopher Aranha
    Rodrigues, Giovani Aantonio
    Junior, Luiz Otavio Murta
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (03)
  • [39] Pitfalls in the measurement of bone mineral density by dual energy X-ray absorptiometry
    Antonacci, D
    Hanson, DS
    Heggeness, MH
    SPINE, 1996, 21 (01) : 87 - 90
  • [40] Bone mineral density of normal people by dual energy X-ray absorptiometry
    Zhang Yuan Xun
    Nuclear Science and Techniques, 1997, (04) : 54 - 56