Transvenous extraction of conduction system and lumenless pacing leads

被引:0
|
作者
Wagner, Ethan S. [1 ]
Lewis, Robert K. [2 ]
Pokorney, Sean D. [2 ]
Hegland, Donald D. [2 ]
Friedman, Daniel J. [2 ]
Piccini, Jonathan P. [2 ]
机构
[1] Duke Univ, Sch Med, Durham, NC USA
[2] Duke Univ Hosp, Dept Med, Div Cardiol, Clin Cardiac Electrophysiol Sect, Durham, NC USA
关键词
conduction system pacing; His bundle pacing; laser lead extraction; left bundle branch area pacing; transvenous lead extraction; EXPERT CONSENSUS STATEMENT; PACEMAKER;
D O I
10.1111/jce.16467
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionConduction system pacing (CSP), often accomplished with lumenless pacing leads, is increasingly employed to achieve physiologic ventricular activation. There are limited data on the extraction of these leads. The objective of this study was to describe the safety and efficacy of extraction of CSP pacing leads and compare outcomes with extraction of non-CSP lumenless leads.MethodsPatients undergoing CSP/non-CSP lumenless lead removal were included. Outcomes of interest included rates of complete procedural success, complications, and successful reimplantation.ResultsOverall, 23 patients were included (n = 14 with CSP and n = 9 with non-CSP lumenless leads implanted in the right atrium, right ventricle, or septum). The mean age was 52.7 +/- 24.0 years, 30% were female, and the mean lead age was 4.5 +/- 4.4 years. The complete procedural success rate was 100%. One serious complication occurred in the non-CSP group but was unrelated to the lead of interest. Manual traction alone was successful in 57% of CSP cases (mean lead age 2.4 +/- 1.7 years) and in 11% of non-CSP cases (mean lead age 7.9 +/- 5.3 years). Laser sheaths were used in 43% of CSP cases and 89% of non-CSP cases; rotational cutting tools were used in no CSP cases and in 33% of non-CSP cases. Reimplantation in the conduction system was attempted with a left bundle branch pacing lead and successful in 80% (n = 4/5).ConclusionExtraction of CSP and non-CSP lumenless leads is feasible with a high success rate and a good safety profile. CSP reimplantation after extraction is also feasible with good electrical performance. Extraction of CSP and non-CSP lumenless leads is feasible with a high success rate and a good safety profile. CSP reimplantation after extraction is also feasible with good electrical performance.image
引用
收藏
页码:2432 / 2443
页数:12
相关论文
共 50 条
  • [31] EVALUATION OF TINED LEADS VERSUS STANDARD LEADS IN TRANSVENOUS CARDIAC PACING
    GODIN, JF
    LHENAFF, HW
    MOUTEL, PC
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (03): : A87 - A87
  • [32] TRANSATRIAL IMPLANTATION OF TRANSVENOUS PACING LEADS AS AN ALTERNATIVE TO IMPLANTATION OF EPICARDIAL LEADS
    BYRD, CL
    SCHWARTZ, SJ
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1856 - 1859
  • [33] Caudal fluoroscopic guidance for the insertion of transvenous pacing leads
    Morton, Matthew B.
    William, Jeremy
    Kistler, Peter M.
    Prabhu, Sandeep
    Sugumar, Hariharan
    Brink, Olivier Van Den
    Patel, Hitesh
    Mariani, Justin
    Voskoboinik, Aleksandr
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2024, 35 (03) : 433 - 437
  • [34] ACTUARIAL SURVIVAL OF TRANSVENOUS PACING LEADS IN A PEDIATRIC POPULATION
    LAU, YR
    GILLETTE, PC
    BUCKLES, DS
    ZEIGLER, VL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (07): : 1363 - 1367
  • [35] TRANSVENOUS REMOVAL OF PACING AND DEFIBRILLATING LEADS: THE PISA EXPERIENCE
    Bongiorni, M. G.
    Soldati, E.
    Segreti, L.
    Zucchelli, G.
    Di Cori, A.
    Viani, S.
    Paperini, L.
    De Lucia, R.
    Levorato, D.
    Boem, A.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S138 - S138
  • [36] Implantable transvenous pacing leads: The shape of things to come
    Mond, HG
    Grenz, D
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 887 - 893
  • [37] Performance of transvenous pacing leads inserted into vascular stents
    Karpawich, P
    Vincent, J
    Forbes, T
    Keller, G
    Cavitt, D
    Webster, P
    NEW TRENDS IN ELECTROCARDIOLOGY, 2000, : 137 - 141
  • [38] Extraction of transvenous leads while anticoagulated
    Montgomery, Jay A.
    Crossley, George H.
    HEART RHYTHM, 2018, 15 (12) : 1782 - 1783
  • [39] Extraction of lumenless pacing leads from the His bundle and left bundle branch area: outcomes of the high-volume centre
    Gardas, Rafal
    Loboda, Danuta
    Biernat, Jolanta
    Soral, Tomasz
    Kulesza, Piotr
    Gladysz-Wanha, Sylwia
    Joniec, Michal
    Sajdok, Mateusz
    Zub, Kamil
    Golba, Krzysztof S.
    EUROPACE, 2024, 26 (09):
  • [40] Use of extendable helix leads for conduction system pacing: Differences in lead handling and performance lead design impacts conduction system pacing
    Tan, Eugene S. J.
    Lee, Jie-Ying
    Boey, Elaine
    Soh, Rodney
    Sim, Ming G.
    Yeo, Wee-Tiong
    Seow, Swee-Chong
    Kojodjojo, Pipin
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (07) : 1550 - 1557