Impact of leadless pacemaker implantation site on cardiac synchronization and tricuspid regurgitation

被引:0
|
作者
Gao, Xiao-fei [1 ]
Zhu, Hong [2 ]
Zhang, Jia-sheng [1 ]
Pan, Xiao-hong [2 ]
Xu, Yi-Zhou [1 ]
机构
[1] Hangzhou First Peoples Hosp, Dept Cardiol, 261 Huansha Rd, Hangzhou 310000, Zhejiang, Peoples R China
[2] Zhejiang Univ, Dept Cardiol, Affiliated Hosp 2, Sch Med, Hangzhou, Zhejiang, Peoples R China
来源
EGYPTIAN HEART JOURNAL | 2025年 / 77卷 / 01期
关键词
Leadless pacemaker; Synchronization; Pacemaker-induced cardiomyopathy; RESYNCHRONIZATION THERAPY; ATRIOVENTRICULAR-BLOCK; DYSSYNCHRONY; ACTIVATION; INDEXES; RHYTHM;
D O I
10.1186/s43044-024-00602-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background To investigate the optimization of leadless pacemaker placement and to assess its impact on heart synchronization and tricuspid regurgitation. Results A clinical trial was conducted involving 53 patients who underwent leadless pacemaker implantation at the Second Affiliated Hospital of Zhejiang University School of Medicine and Hangzhou First People's Hospital between March 2022 and February 2023. Implantation site localization was determined using the 18-segment method under RAO 30 degrees imaging. Intraoperative and 1-month post-operative echocardiography was performed to assess cardiac electromechanical synchronization and tricuspid regurgitation; parameters of interest included interventricular mechanical delay (IVMD), pre-ejection period of the aorta (L-PEI), and septal-to-posterior wall motion delay (SPWMD). Pacing thresholds, sensing, and impedance exhibited no significant differences between the 8/9 zone and other sites (P > 0.05). In contrast, the 8/9 zone group manifested a significant reduction in L-PEI (128.24 +/- 12.27 vs. 146.50 +/- 18.17 ms, P < 0.001), IVMD (17.92 +/- 8.47 vs. 28.56 +/- 15.16 ms, P < 0.001), and SPWMD (72.84 +/- 19.57 vs. 156.56 +/- 81.54 ms, P < 0.001), compared to the non-8/9 group. Post-pacing QRS duration showed no significant difference between the two groups (139.21 +/- 11.36 vs. 143.83 +/- 16.35 ms P = 0.310). Notably, for patients with atrial fibrillation, the 8/9 zone placement significantly reduced tricuspid regurgitation. During the 1-month follow-up, neither group reported major complications such as bleeding, cardiac tamponade, pacemaker detachment, or malignant arrhythmias. Conclusion Implantation of the leadless pacemaker in the right ventricular 8/9 zone provides superior electromechanical synchronization compared to other sites.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Impact of leadless pacemaker therapy on tricuspid regurgitation
    Mendoza, I. M.
    Ukani, Z.
    Rodriguez, E. R.
    Rangel, A. R.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [2] Leadless cardiac pacemaker implantation in patient with active tricuspid endocarditis
    Jacobs, Mathieu
    Bisson, Arnaud
    Bodin, Alexandre
    Babuty, Dominique
    Niro, Marjorie
    Clementy, Nicolas
    EUROPACE, 2021, 23 (11): : 1794 - 1794
  • [3] Severe Tricuspid Regurgitation With Chordae Tendinae Rupture After Leadless Pacemaker Implantation
    Kobara, Satoshi
    Okamura, Akihiro
    Kato, Masaru
    Ogura, Kazuyoshi
    Nishimura, Motonobu
    Yamamoto, Kazuhiro
    CIRCULATION JOURNAL, 2022, 86 (05) : 880 - 880
  • [4] Close Proximity of Leadless Pacemaker to Tricuspid Annulus Predicts Worse Tricuspid Regurgitation Following Septal Implantation
    Hai, Jo-Jo
    Mao, Yankai
    Zhen, Zhe
    Fang, Jonathan
    Wong, Chun-Ka
    Siu, Chung-Wah
    Yiu, Kai-Hang
    Lau, Chu-Pak
    Tse, Hung-Fat
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2021, 14 (05): : 539 - 541
  • [5] Prevalence of new-onset tricuspid regurgitation after leadless Micra pacemaker implantation
    La Fazia, V.
    Gianni, C.
    Pierucci, N.
    Della Rocca, D. G.
    Mohanty, S.
    Torlapati, G.
    Al-Ahmad, A.
    Bassiouny, M.
    Gallinghouse, G. J.
    Horton, R.
    Burkhardt, J. D.
    Lakkireddy, D.
    Forleo, G. B.
    Di Biase, L.
    Natale, A.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [6] Conduction Abnormalities and Tricuspid Regurgitation after Cardiac Transplant: Is a Leadless Pacemaker Part of the Solution?
    Araj, Faris G.
    ASAIO JOURNAL, 2020, 66 (03) : E56 - E56
  • [7] Worsening Tricuspid Regurgitation After Leadless Pacemaker Implantation Leads to Higher Hospitalization for Heart Failure
    Lee, Kevin
    Metzl, Mark
    CIRCULATION, 2022, 146
  • [8] Permanent pacemaker implantation associated tricuspid regurgitation
    Niazi, Gul Zaman Khan
    Masood, Arslan
    Ahmed, Nisar
    Qadir, Irfan
    Akhtar, Ammar
    Saleemi, Muhammad Sohail
    Ahmed, Ijaz
    ASIAN CARDIOVASCULAR & THORACIC ANNALS, 2021, 29 (03): : 191 - 194
  • [9] Transfemoral leadless pacemaker implantation after transcatheter tricuspid valve in ring implantation
    Turdubaev, Abai
    Fink, Thomas
    Eitz, Thomas
    Imnadze, Guram
    CLINICAL RESEARCH IN CARDIOLOGY, 2024, 113 (01) : 187 - 189
  • [10] Transfemoral leadless pacemaker implantation after transcatheter tricuspid valve in ring implantation
    Abai Turdubaev
    Thomas Fink
    Thomas Eitz
    Guram Imnadze
    Clinical Research in Cardiology, 2024, 113 : 187 - 189