Patients dose and risk assessment in fluoroscopically guided interventional procedures in Ghana

被引:0
|
作者
Issahaku, Shiraz [1 ,2 ]
Hasford, Francis [1 ,2 ]
Azantillow, Norbert [3 ]
Sackey, Theophilus Akumea [1 ,2 ]
Nunoo, George [4 ]
Dzefi-Tettey, Klenam [5 ]
Mensah, Yaw Boateng [5 ]
Sarkodie, Benjamin [5 ]
Amponsah, Becky Appiah [3 ]
Tagoe, Samuel [6 ]
Esposito, Marco [7 ]
Padovani, Renato [7 ]
机构
[1] Ghana Atom Energy Commiss, Accra, Ghana
[2] Univ Ghana, Dept Med Phys, Accra, Ghana
[3] Int Maritine Hosp, Accra, Ghana
[4] Univ Ghana, Dept Radiol, Med Ctr, Accra, Ghana
[5] Korle Bu Teaching Hosp, Dept Radiol, Accra, Ghana
[6] Natl Ctr Radiotherapy & Nucl Med, Accra, Ghana
[7] Abdus Salaam Int Ctr Theoret Phys, Trieste, Italy
关键词
KAP; Fluoroscopy; Interventional procedure; Kerma; Dose optimisation;
D O I
10.1007/s12553-024-00919-7
中图分类号
R-058 [];
学科分类号
摘要
IntroductionThe significant increases in the use of imaging and interventional procedures in recent times have brought along the issue of safety and protection of patients and users of diagnostic and interventional radiology equipment for complex procedures. Additionally, due to potential risk associated with these procedures, patients' radiation protection issues have become the focus of most fluoroscopy guided interventional procedures globally. The objective of this study is to evaluate optimized angiographic exposure parameters, radiation dose, estimated risk levels and develop a radiation dose reference chart for the most commonly routine protocol in interventional procedures in Ghana.MethodologyPatients dose parameters were obtained from the structured dose reports of the PAC system of five diagnostic interventional equipment facilities, together with number of images per procedure and image quality assessment for dose optimisation protocols. The data retrieved from the PAC System include patients' biodata, input and output indices including Kerma-Area Product (KAP) Fluoro and total Kerma-Area Product values. Additionally, fluoroscopy time and the number of radiographic images of at least 50 patients for each FGI procedure based on ICRP Publication 135 recommendations were also collected. The PC based Monte Carlo simulation software (PCXMC) was used to estimate ED from the total KAP and the KAP Fluoro values. Image quality (IQ) was assessed using human observer assessment and evaluation based on the assessment of the radiologists, who ranked the images from most acceptable images to least acceptable images and whether accepted on not accepted for clinical use. Patient radiation exposure was evaluated using estimated KAP Fluoro and Total KAP to estimate ED from each of the measured parameters. That is radiation risk levels were estimated based on Biological Effects of Ionising radiation (BEIR) report VII predictive models. These were done by assessing the dose and risk to patients and practioners during fluoroscopy guided interventional procedures for each of the nine most common routine procedure and the possible corrective measures needed for effective management of these procedure.ResultsThe study shows that the total KAP median dose value was 109 with a minimum and maximum values in the range of 2 to 835 Gy.cm2 were recorded during Abdominal Aorta Endoprosthesis procedure. While KAP Fluoro minimum and maximum dose value were ranged from 1.59 to 734 Gy.cm2 with a median value of 86 Gy.cm2. The trend of the study shows significant comparative estimated dose values in terms of effective dose from the KAP median values with other studies. Additionally, abdominal aorta endoprosthesis procedures took much longer time to perform relative to the other studies with resultant high ED values. It was observed that, time taken, and images acquired for abdominal aorta endoprosthesis procedure could be as much as three-fold with significant dose value compared to other procedures. Furthermore, higher radiation dose quantity, in terms of KAP Fluoro and total KAP, in the case of Abdominal Aorta Endoprosthesis had a direct correlation to the number of images and the time taken to complete the procedure. Comparative analysis between these procedures shows significant variation among them. For instance Percutaneous Liver Drainage had the least effective dose and risk factors of 0.94 mSv and 0.34 LAR respectively. Whilst Abdominal Aorta Endoprosthesis had the highest effective dose and risk factor of 32 mSv and 29 LAR respectively. Finally, the number of images and the corresponding image quality assessment shows significant variation from a minimum of 2.7 to a maximum of 167 Gy.cm2. This resulted in highest ED of 59.4 mSv with significant risk levels to both patients and practitioners (1 in 1000 to 1 in 500). Most of the procedures had low radiation risk levels (1 in 10,000 to 1 in 1000). Additionally, all the images were good enough for clinical use and ranged from 3 to 5 with less than 2% of the procedure having relative poor image quality with significant positive correlation to the dose and risk factor metrics.ConclusionIn conclusion it was observed that, the study results were comparable to other regional and international studies. In addition, it was also observed that, the time taken for each procedure to be completed had direct positive correlation with estimated patient dose metrics. That is, the higher the time taken for a procedure to be completed, the higher the dose values in terms of KAP Fluoro (Gy<middle dot>cm2) and Total KAP (Gy<middle dot>cm2) with significant correlation with the quality of images used during the procedure. Finally, the study found that the angiographic exposure input parameters were optimized and within accepted and recommened dose estimates (effective dose, estimated risk levels) and radiation dose reference chart has been developed for the most commonly routine protocol in interventional procedures in Ghana.
引用
收藏
页码:119 / 130
页数:12
相关论文
共 50 条
  • [41] Collar Badge Lens Dose Equivalent Values among United States Physicians Performing Fluoroscopically Guided Interventional Procedures
    Borrego, David
    Yoder, Craig
    Balter, Stephen
    Kitahara, Cari M.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 33 (03) : 219 - +
  • [42] Occupational Radiation Dose Trends in US Radiologic Technologists Assisting with Fluoroscopically Guided Interventional Procedures, 1980-2020
    Milder, Cato M.
    Borrego, David
    Preston, Dale L.
    Villoing, Daphnee
    Kwon, Tae-Eun
    Miller, Donald L.
    Alexander, Bruce H.
    Linet, Martha S.
    Lee, Choonsik
    Kitahara, Cari M.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2024, 35 (07) : 1057 - 1065.e4
  • [43] PRELIMINARY ASSESSMENT OF THE DOSE TO THE INTERVENTIONAL RADIOLOGIST IN FLUORO-CT-GUIDED PROCEDURES
    Pereira, M. F.
    Alves, J. G.
    Sarmento, S.
    Santos, J. A. M.
    Sousa, M. J.
    Gouvea, M.
    Oliveira, A. D.
    Cardoso, J. V.
    Santos, L. M.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 448 - 452
  • [44] Patient skin dose assessment during CT-guided interventional procedures
    Tsapaki, V.
    Triantopoulou, C.
    Maniatis, P.
    Kottou, S.
    Tsalafoutas, J.
    Papailiou, J.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 29 - 31
  • [45] Dose reduction for fluoroscopically guided injections: phantom simulation and patient procedures
    C. Y. Chang
    F. J. Simeone
    M. C. DeLorenzo
    W. E. Palmer
    M. A. Bredella
    A. J. Huang
    Skeletal Radiology, 2018, 47 : 223 - 231
  • [46] X-ray dose reduction in fluoroscopically guided electrophysiology procedures
    Davies, AG
    Cowen, AR
    Kengyelics, SM
    Moore, J
    Pepper, C
    Cowan, C
    Sivanathan, MU
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (03): : 262 - 271
  • [47] Dose indices of radiation to skin in fluoroscopically guided invasive cardiology procedures
    Chu, Robert Y. L.
    Schechter, Eliot
    Chu, Nancy
    HEALTH PHYSICS, 2007, 93 (02): : S124 - S127
  • [48] Dose reduction for fluoroscopically guided injections: phantom simulation and patient procedures
    Chang, C. Y.
    Simeone, F. J.
    DeLorenzo, M. C.
    Palmer, W. E.
    Bredella, M. A.
    Huang, A. J.
    SKELETAL RADIOLOGY, 2018, 47 (02) : 223 - 231
  • [49] Real-Time Plastic Scintillation Dosimeter for Fluoroscopically-Guided Interventional Procedures
    Boivin, J.
    Guillemette, M.
    Beaulieu, L.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [50] The Use of Hybrid Computational Phantoms to Estimate the Patient's Peak Skin Dose During Fluoroscopically Guided Interventional Procedures (FGIPs)
    Makkia, R.
    Fadhel, M.
    Hoerner, M.
    Vergara, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E217 - E217