Agreement of intraocular pressure measurement by scleral Schiotz and Goldmann applanation tonometer

被引:0
|
作者
Shaikh, Zeba Muzammil [1 ]
Rajalakshmi, A. R. [1 ]
Nagarajan, Swathi [1 ]
Lokeshmaran, A. [2 ]
机构
[1] Deemed Be Univ, Sri Balaji Vidyapeeth SBV, Mahatma Gandhi Med Coll & Res Inst, Dept Ophthalmol, Pondicherry, India
[2] Deemed Be Univ, Sri Balaji Vidyapeeth SBV, Mahatma Gandhi Med Coll & Res Inst, Dept Community Med, Pondicherry, India
来源
SCIENTIFIC REPORTS | 2024年 / 14卷 / 01期
关键词
Corneal IOP; Intraocular pressure estimation; Ocular rigidity; Predictive formula for IOP; Schiotz tonometer; Scleral biomechanics; Scleral IOP; CORNEAL; PNEUMATONOMETRY; THICKNESS;
D O I
10.1038/s41598-024-79403-1
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Aim & background : Intraocular pressure (IOP) measured through the sclera has some advantages over IOP measured through cornea. This study aimed to determine the agreement between IOP as estimated by scleral Schiotz (SS) tonometer and Goldmann applanation tonometer (GAT). Methods: In this cross-sectional observational study, IOP was measured by GAT on the cornea and Schiotz tonometer on the temporal sclera. Axial length and central corneal thickness (CCT) were noted. IOP was then estimated from SS IOP as eGAT using a predictive formula. Agreement of IOP corrected for CCT (cGAT) with eGAT and SS IOP were assessed. The effects of age, axial length, refraction, CCT, gender, and ocular and systemic comorbidity on measured IOP were also assessed. Results: The study included 155 patients with a mean age of 54.27 +/- 10.02 years. Mean IOP measured by GAT and scleral Schiotz were 16.58 +/- 2.76 mm Hg and 16.10 +/- 2.96 mm Hg, respectively. The mean IOP by cGAT and eGAT were 16.70 +/- 2.80 mm Hg and 17.07 +/- 3.10 mm Hg, respectively. Both eGAT & SS IOP had good agreement with cGAT on the Bland-Altmann plot (p = 0.96 & 0.51, respectively). Further, SS IOP showed good agreement even with GAT IOP (p = 0.78). Age, axial length, refraction, central corneal thickness, gender, and ocular and systemic comorbidity did not show significant correlation with IOP measurement. Conclusion: Estimated GAT IOP showed good agreement with measured GAT values. SS IOP also agreed well with GAT IOP. Clinical significance IOP measurement on the sclera using Schiotz can be used as an alternative to GAT.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses
    Senthil, Sirisha
    Chary, Raghava
    Ali, Mohammed Hasnat
    Choudhari, Nikhil S.
    Garudadri, Chandra Sekhar
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2022, 106 (04) : 480 - 484
  • [2] ASSESSMENT OF THE FACTORS INFLUENCING AND COMPARING THE INTRAOCULAR PRESSURE WITH THE HELP OF SCHIOTZ INDENTATION TONOMETER AND GOLDMANN'S APPLANATION TONOMETER: A CLINICAL STUDY
    Amarnath, Mridula V.
    Samuel, Christina
    Sundararajan, D.
    INTERNATIONAL JOURNAL OF MEDICAL RESEARCH & HEALTH SCIENCES, 2015, 4 (01): : 58 - 64
  • [3] RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES
    DIELEMANS, I
    VINGERLING, JR
    HOFMAN, A
    GROBBEE, DE
    DEJONG, PTVM
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) : 141 - 144
  • [4] Schiotz Scleral Intraocular Pressure Readings Predict Goldmann Applanation Readings Better Than Rebound Tonometry
    Senthil, Sirisha
    Chary, Raghava
    Ali, Mohammed Hasnat
    Choudhari, Nikhil
    Badakere, Swathi
    Krishnamurthy, Rashmi
    Dikshit, Siddharth
    Garudadri, Chandrasekhar
    CORNEA, 2019, 38 (09) : 1117 - 1123
  • [5] Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea
    Badakere, Swathi Vallabh
    Chary, Raghava
    Choudhari, Nikhil S.
    Rao, Harsha L.
    Garudadri, Chandrasekhar
    Senthil, Sirisha
    OPHTHALMOLOGY GLAUCOMA, 2021, 4 (03): : 238 - 243
  • [6] Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea
    Badakere, Swathi Vallabh
    Chary, Raghava
    Choudhari, Nikhil S.
    Rao, Harsha L.
    Garudadri, Chandrasekhar
    Senthil, Sirisha
    OPHTHALMOLOGY GLAUCOMA, 2021, 4 (01): : 89 - 94
  • [7] Agreement of intraocular pressure measurement with Corvis ST, non-contact tonometer, and Goldmann applanation tonometer in children with ocular hypertension and related factors
    Li, Hou-Gang
    Chen, Yan-Hui
    Lin, Fang
    Li, Si-Yu
    Liu, Qing-Hua
    Yin, Chun-Ge
    Chen, Xi-Yue
    Zhang, Xin-Jie
    Qu, Yue
    Hui, Yan-Nian
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2023, 16 (10) : 1601 - 1607
  • [8] Comparison of intraocular pressure measured by iCare tonometers and Goldmann applanation tonometer
    Sachdeva, Robin
    Iordanous, Yiannis
    Lin, Tony
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2023, 58 (05): : 426 - 432
  • [9] Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure
    Sakaue, Yuta
    Ueda, Jun
    Seki, Masaaki
    Tanaka, Takayuki
    Togano, Tetsuya
    Yoshino, Takaiko
    Fukuchi, Takeo
    JOURNAL OF OPHTHALMOLOGY, 2014, 2014
  • [10] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770