Fast Demodulated White-Light Interferometry-Based Measurement and Theoretical Verification of Mechanical Sensitivity of MEMS Biomimetic Microphone With Fiber-Optic Fabry-Perot Interferometer

被引:0
|
作者
Ren, Dipeng [1 ]
Yang, Xiaonan [1 ]
Qi, Zhi-Mei [2 ,3 ,4 ]
机构
[1] Zhengzhou Univ, Sch Elect & Informat Engn, Zhengzhou 450000, Peoples R China
[2] Chinese Acad Sci, State Key Lab Transducer Technol, Aerosp Informat Res Inst, Beijing 100190, Peoples R China
[3] Univ Chinese Acad Sci, Sch Elect Elect & Commun Engn, Beijing 100049, Peoples R China
[4] Univ Chinese Acad Sci, Coll Mat Sci & Optoelect Technol, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
Fast demodulated white-light interferometry (WLI); fiber-optic Fabry-Perot interferometer (FPI); improved theoretical model; mechanical sensitivity; MEMS biomimetic microphone; SOUND SOURCE LOCALIZATION; EARS;
D O I
10.1109/JSEN.2024.3479197
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
MEMS microphones inspired by the coupled ears of the fly Ormia ochracea have been extensively studied to achieve miniaturization of sound source localization (SSL). However, the SSL of all current biomimetic microphones relies on the electrical signals converted from the directional vibration of the biomimetic diaphragms. Consequently, only the electrical performance of the biomimetic microphones under sound pressure excitation (i.e., electrical sensitivity mV/Pa) has been characterized, whereas the mechanical performance (i.e., mechanical sensitivity nm/Pa) has not. Characterizing the mechanical sensitivity can not only directly represent the acoustic response magnitude of the biomimetic microphones but also directly verify the corresponding theoretical models. Therefore, in this article, we designed and fabricated an MEMS biomimetic microphone with a fiber-optic Fabry-Perot interferometer (FPI), based on which we successfully measured the mechanical sensitivity of the biomimetic microphone by the fast demodulated white-light interferometry (WLI) for the first time. In addition, an improved theoretical model with exact expressions was first derived to better describe the acoustic response of the biomimetic microphone, and the theoretical and measurement results can be corroborated by each other in terms of mechanical sensitivity. Therefore, this study verifies a feasible solution to directly measure the acoustic response of the biomimetic microphone and constructs a more effective and detailed theoretical model to describe this acoustic response. The results in this article can not only provide a numerical reference for designing future demodulation systems regarding the dynamic range and the linear demodulation interval but also promote a deeper understanding of the working mechanism of the biomimetic microphones.
引用
收藏
页码:39091 / 39100
页数:10
相关论文
共 28 条
  • [21] High-consistency fiber-optic Fabry-Perot sensor based on MEMS for simultaneous temperature and liquid refractive index measurement
    Wang, Shuang
    Wu, Wen
    Sang, Mei
    Jiang, Junfeng
    Liu, Kun
    Wang, Xue
    Yu, Xun
    Liu, Tiegen
    APPLIED OPTICS, 2020, 59 (30) : 9353 - 9358
  • [22] UV-Curable Polymer Microhemisphere-Based Fiber-Optic Fabry-Perot Interferometer for Simultaneous Measurement of Refractive Index and Temperature
    Tan, X. L.
    Geng, Y. F.
    Li, X. J.
    Deng, Y. L.
    Yin, Z.
    Gao, R.
    IEEE PHOTONICS JOURNAL, 2014, 6 (04):
  • [23] Fiber-Optic Hybrid-Structured Fabry-Perot Interferometer Based On Large Lateral Offset Splicing for Simultaneous Measurement of Strain and Temperature
    Wu, Yongfeng
    Zhang, Yundong
    Wu, Jing
    Yuan, Ping
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2017, 35 (19) : 4311 - 4315
  • [24] A mechanical amplifier based high-finesse fiber-optic Fabry-Perot interferometric sensor for the measurement of static magnetic field
    Feng, Xinxing
    Jiang, Yi
    Zhang, Han
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2021, 32 (12)
  • [25] 3D imaging of mural model in air using sensitivity-improved ultrasonic sensor based on fiber-optic Fabry-Perot interferometer
    Zuo, Chi
    Gang, Tingting
    Hu, Manli
    Bai, Xiaohong
    Tong, Rongxin
    Bian, Ce
    Wang, Jie
    Dong, Leigang
    OPTIK, 2019, 185 : 1205 - 1212
  • [26] High-Order Harmonic-Frequency Cross-Correlation Algorithm for Absolute Cavity Length Interrogation of White-Light Fiber-Optic Fabry-Perot Sensors
    Chen, Haibin
    Liu, Jie
    Zhang, Xiongxing
    Wang, Wei
    Ma, Zhibo
    Lv, Wentao
    Guo, Zilong
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2020, 38 (04) : 953 - 960
  • [27] High-sensitivity and fast-response fiber-optic micro-thermometer based on a piano-concave Fabry-Perot cavity filled with PDMS
    Kong, Ling-Xin
    Zhang, Yan-Xin
    Zhang, Wei-Gang
    Zhang, Yun-Shan
    Yu, Lin
    Wang, Song
    Geng, Peng-Cheng
    Yan, Tie-Yi
    SENSORS AND ACTUATORS A-PHYSICAL, 2018, 281 : 236 - 242
  • [28] Combined Interrogation Algorithm of Phase Function and Minimum Mean Square Error for Fiber-Optic Fabry-Perot Micro-Pressure Sensors Based on White Light Interference
    Liu, Wei
    Yang, Tianyu
    Shi, Yunjie
    Wu, Jianwei
    Dong, Yuming
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2023, 41 (10) : 3241 - 3248