Quantitative spectral/spatial analysis of phased array coil in magnetic resonance imaging based on method of moment

被引:0
|
作者
Program of Medical Engineering and Medical Physics, Harvard University, Cambridge, MA 02138, United States [1 ]
机构
来源
IEEE Trans. Med. Imaging | / 12卷 / 1129-1137期
关键词
Electric coils - Equivalent circuits - Frequency response - Magnetic fields - Magnetic resonance imaging - Mathematical transformations - Method of moments - Spectrum analysis;
D O I
暂无
中图分类号
学科分类号
摘要
A new approach for analysis of RF coils in magnetic resonance (MR) experiments is reported. Instead of assuming current distribution in conventional quasi-static algorithm, this approach transforms the coil geometry into an equivalent circuit for complex current calculation. Self and mutual inductance are taken into consideration. Frequency responses of RF coils and transverse magnetic field (B1) maps can be simulated. This approach is especially efficient for phased array coil design for its small matrix size when implemented on computers. Experiments on both single surface coil and phased array coils are consistent with simulation results.
引用
收藏
相关论文
共 50 条
  • [21] Study on the Decoupling of Stacked Phased Array Coils for Magnetic Resonance Imaging
    Liang, Dandan
    Hui, Hon Tat
    Yeo, Tat Soon
    Yu, Yantao
    PIERS 2011 SUZHOU: PROGRESS IN ELECTROMAGNETICS RESEARCH SYMPOSIUM, 2011, : 601 - 604
  • [22] A Parasitic Superdirective Electrically Small Coil Array for Magnetic Resonance Imaging
    Tang, Xinyi
    Qing, Xianming
    Chen, Zhi Ning
    2017 11TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2017, : 3702 - 3704
  • [23] Prostate Cancer: Added Value of Subtraction Dynamic Imaging in 3T Magnetic Resonance Imaging with a Phased-array Body Coil
    Yu, Jeong-Sik
    Chung, Jae-Joon
    Hong, Soon Won
    Chung, Byung Ha
    Kim, Joo Hee
    Kim, Ki Whang
    YONSEI MEDICAL JOURNAL, 2008, 49 (05) : 765 - 774
  • [24] Focused, eight-element transceive phased array coil for parallel magnetic resonance imaging of the chest-theoretical considerations
    Li, BK
    Liu, F
    Crozier, S
    MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (06) : 1251 - 1257
  • [25] Moment method analysis of mutual interaction in MRI phased array coils
    Fujita, Hiroyuki
    Missal, John W.
    Morich, Michael A.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2000, 10 (02) : 84 - 92
  • [26] Moment method analysis of mutual interaction in MRI phased array coils
    Hiroyuki Fujita
    John W. Missal
    Michael A. Morich
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2000, 10 (2) : 84 - 92
  • [27] Accuracy of Thin-Section Magnetic Resonance Imaging With a Pelvic Phased-Array Coil in the Local Staging of Rectal Cancer
    Halefoglu, Ahmet Mesrur
    Atasoy, Serpil Topuz
    Sakiz, Damlanur
    Baykan, Adil
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (01) : 58 - 64
  • [28] How the phased array coil elements placement influences the image quality in cardiac magnetic resonance
    Martini N.
    Flori A.
    De Marchi D.
    Giovannetti G.
    Research on Biomedical Engineering, 2021, 37 (03) : 553 - 561
  • [29] Superconducting 300 MHz phased-array for magnetic resonance imaging applications
    Wosik, J.
    Kamel, M. R.
    Xue, L.
    Xie, L-M
    Nesteruk, K.
    SIXTH INT KHARKOV SYMPOSIUM ON PHYSICS AND ENGINEERING OF MICROWAVES, MILLIMETER AND SUBMILLIMETER WAVES/WORKSHOP ON TERAHERTZ TECHNOLOGIES, VOLS 1 AND 2, 2007, : 152 - 152
  • [30] Phased array magnetic resonance imaging for staging clinically localized prostate cancer
    Borre, M
    Lundorf, E
    Marcussen, N
    Langkilde, NC
    Wolf, H
    ACTA ONCOLOGICA, 2005, 44 (06) : 589 - 592