A method for an accurate T1 relaxation-time measurement compensating B1 field inhomogeneity in magnetic-resonance imaging

被引:0
|
作者
Mihara, Hiroaki [1 ]
Sekino, Masaki [1 ]
Iriguchi, Norio [2 ]
Ueno, Shoogo [1 ]
机构
[1] Department of Biomedical Engineering, Graduate School of Medicine, University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan
[2] Center for Multimedia and Information Technologies, University of Kumamoto, 2-39-1 Kurokami, Kumamoto 860-8555, Japan
来源
Journal of Applied Physics | 2005年 / 97卷 / 10期
关键词
Brain - Data reduction - Image analysis - Magnetic relaxation - Parameter estimation - Vectors;
D O I
暂无
中图分类号
学科分类号
摘要
Two-point gradient echo methods are practical means for estimating T1 value in magnetic-resonance (MR) imaging. Meanwhile, the inhomogeneity of the transmitted B1 field and the imperfect slice profile severely distorts the estimated T1 value. We propose a method for an accurate T1 measurement compensating the inhomogeneity of the B1 field and the effects of the imperfect slice profile. Three MR images of a whole human brain were obtained by multisliced spoiled gradient echo sequences on a 1.5-T scanner with different repetition times and flip angles to calculate the quantitative T1 maps. The histograms of the T1 maps showed two distinguishable peaks of the white matter (672.9±15.5 ms) and the gray matter (921.1±24.7 ms) of the human brain. The results indicate that the T1 value was properly estimated compensating the B1 field inhomogeneity and the slice profile imperfection. The proposed method is applicable to images obtained on any scanners with different parameters, because the method needs neither the information of the flip angle calibrations nor the phantom data to compensate the B1 field inhomogeneity. © 2005 American Institute of Physics.
引用
收藏
相关论文
共 50 条
  • [1] A method for an accurate T1 relaxation-time measurement compensating B1 field inhomogeneity in magnetic-resonance imaging -: art. no. 10E107
    Mihara, H
    Sekino, M
    Iriguchi, N
    Ueno, S
    JOURNAL OF APPLIED PHYSICS, 2005, 97 (10)
  • [2] MAGNETIC-RESONANCE-IMAGING IN IDIOPATHIC RETROPERITONEAL FIBROSIS - MEASUREMENT OF T1 RELAXATION-TIME
    BROOKS, AP
    REZNEK, RH
    WEBB, JAW
    BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (755): : 842 - 844
  • [3] A METHOD FOR T1 RELAXATION-TIME COMPUTED IMAGES
    INOUYE, T
    ARAKI, T
    MAGNETIC RESONANCE IN MEDICINE, 1984, 1 (02) : 179 - 180
  • [4] MAGNETIC-RESONANCE IMAGING OF BRAIN-TUMORS - MEASUREMENT OF T1
    ARAKI, T
    INOUYE, T
    SUZUKI, H
    MACHIDA, T
    IIO, M
    RADIOLOGY, 1984, 150 (01) : 95 - 98
  • [5] ALTERATIONS OF THE PROTON NUCLEAR MAGNETIC-RESONANCE SPIN-LATTICE RELAXATION-TIME (T1) IN REJECTING CARDIAC ALLOGRAFTS
    RATNER, AV
    BARRETT, LV
    GANG, DL
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 3 (02) : 538 - 538
  • [6] REPRODUCIBILITY OF SPIN-LATTICE RELAXATION-TIME (T1) MEASUREMENT USING AN 0.08 TESLA MD-800 MAGNETIC-RESONANCE IMAGER
    RICHARDS, MA
    GREGORY, WM
    WEBB, JAW
    JEWELL, SE
    REZNEK, RH
    BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (711): : 241 - 244
  • [7] T1 relaxation time and magnetic resonance imaging of inflamed gingival tissue
    Schara, R.
    Sersa, I.
    Skaleric, U.
    DENTOMAXILLOFACIAL RADIOLOGY, 2009, 38 (04) : 216 - 223
  • [8] RETARDED GREEN-FUNCTION CALCULATION OF MAGNETIC-RESONANCE RELAXATION-TIME T1 FOR ROTATING MOLECULES OR ATOMIC GROUPS IN SOLIDS
    MARTINLANDROVE, M
    MORENO, JA
    JOURNAL OF MAGNETIC RESONANCE, 1992, 96 (01): : 14 - 23
  • [9] Quantitative T1ρ imaging using phase cycling for B0 and B1 field inhomogeneity compensation
    Chen, Weitian
    Takahashi, Atsushi
    Han, Eric
    MAGNETIC RESONANCE IMAGING, 2011, 29 (05) : 608 - 619
  • [10] NEW METHOD FOR MEASUREMENT OF RELAXATION TIME T1
    BALDACCHINI, G
    NUCLEAR INSTRUMENTS & METHODS, 1971, 92 (03): : 439 - +