Display stabilising circuit for systolic-diastolic blood-pressure monitors

被引:0
|
作者
Klevenhagen, S.C. [1 ]
Storey, P. [1 ]
机构
[1] Dept. Med. Phys., London Hosp., Whitechapel, London, United Kingdom
来源
关键词
BIOMEDICAL ENGINEERING - Patient Monitoring - DISPLAY DEVICES - Stability - PRESSURE TRANSDUCERS - Digital Readout;
D O I
暂无
中图分类号
学科分类号
摘要
When using a digital panel meter to display blood pressure, it is difficult to achieve a stable, clinically usable display. A circuit is described, which allows the measured parameter to be displayed in a stable manner ignoring small beat-to-beat variations. The described circuit and a systolic-diastolic separator provide the complete circuitry for a low-cost instrument for the numerical display of arterial blood pressure.