共 50 条
Development of an all-in-one PCPS system
被引:0
|作者:
Funakubo, Akio
[1
]
机构:
[1] Japan Soc for Precision Engineering, Tokyo, Japan
来源:
Seimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering
|
1993年
/
59卷
/
06期
关键词:
Cardiology - Centrifugal pumps - Fluid dynamics - Hemodynamics;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
The present paper deals with development of an all-in-one PCPS (Percutaneous Cardio Pulmonary Support) system. The system consists of a pump-oxygenator unit and pump controller. The pump-oxygenator unit contains centrifugal pump and extracapillary flow type membrane oxygenator (0.8 m2) in one compact body. Dimensions of the pump-oxygenator unit are as follows: length: 250 mm; diameter: 90 mm; weight: 1 kg. The unit has priming volume of 250 ml. The in-vitro tests (gas exchange, fluid dynamic performance, hemolysis) and an ex-vivo test was conducted. The results are as follows : 1. Fluid dynamic: The system products 1.6-3.8 l/min of flow against 100 mmHg of outlet pressure at a rotation speed of 2 500 to 3 500 rpm. 2. Hemolysis: The increase rate of plasma free hemoglobin was 22 mg·dl-1/h with the PCPS system. 3. Gas exchange: O2 transfer rate of the system was 180 ml/min and CO2 transfer rate was 120 ml/min (V/Q=1) at a blood flow rate 3 l/min. 4. Ex-vivo test: The device was operable. In conclusion, the system is considered applicable to cardio pulmonary support.
引用
收藏
页码:981 / 986
相关论文