We describe the properties of a detector array made up of 64 NaI(Tl) 406 multiplied by 63 multiplied by 63 mm**3 modules, used as an intermediate energy photon spectrometer. We obtain an energy resolution of 6% FWHM at 129 Mev, a time resolution of 1 ns FWHM and a resolution of 48 mm FWHM for the location of the impact point on the front face of the detector. The modularity allows to some extent a discrimination between photons and neutrons. We also present the response of the detector to 69 Mev neutrons.
引用
收藏
页码:497 / 506
相关论文
共 50 条
  • [1] A MODULAR NAI(TI) DETECTOR FOR INTERMEDIATE ENERGY PHOTONS
    BAY, A
    JOSEPH, C
    LOUDE, JF
    PERROUD, JP
    RUEGGER, D
    SCHORI, O
    STEINER, D
    TRAN, MT
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 271 (03): : 497 - 506
  • [2] A MODULAR NAI(TL) DETECTOR FOR 20-1000 MEV PHOTONS
    BLUM, P
    GUIGAS, R
    KOCH, H
    MEYER, M
    POTH, H
    RAICH, U
    RICHTER, B
    BACKENSTOSS, G
    HASINOFF, M
    PAVLOPOULOS, P
    REPOND, J
    TAUSCHER, L
    TROSTER, D
    ADIELS, L
    BERGSTROM, I
    FRANSSON, K
    KEREK, A
    SUFFERT, M
    ZIOUTAS, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 213 (2-3): : 251 - 259
  • [3] Compensation for collimator detector response in tumor SPECT imaging using 111In with dual energy photons.
    Sayeram, S
    Tsui, BM
    Zhao, XD
    Frey, EC
    Falen, SW
    McCartney, WH
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (05) : 197P - 197P
  • [4] Reducing the unusable areas of modular NaI(Tl) detector
    Li, Yulan
    Jing, Yongjie
    Hedianzixue Yu Tance Jishu/Nuclear Electronics & Detection Technology, 2000, 20 (02): : 112 - 114
  • [5] SPECTRAL TRANSFORMATION OBTAINED WITH SCINTILLATION DETECTOR NAI(TL) RELATIVE TO ENERGY-DISTRIBUTION OF PHOTONS DETECTED
    CHARRIER, J
    NUCLEAR INSTRUMENTS & METHODS, 1975, 128 (01): : 101 - 108
  • [6] An improved method for measurement of gamma camera sensitivity for high energy photons.
    Wagenaar, DJ
    Zimmerman, RF
    Hill, TC
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 876 - 876
  • [7] K alpha L1 MULTIPLET STRUCTURE OF Na EXCITED BY PHOTONS.
    Murti, M.V.R.
    Rao, K.S.
    Gopalakrishna, V.
    Raju, M.L.N.
    Rao, Thirumala
    Parthasaradhi, K.
    Murti, V.Radhakrishna
    1600, (144):
  • [8] MULTIFILAMENT COUNTER WITH A 5% ENERGY RESOLUTION FOR 5. 9 KeV PHOTONS.
    Peskov, V.D.
    Podolyak, E.R.
    Instruments and experimental techniques New York, 1982, 25 (6 pt 2): : 1498 - 1501
  • [9] ENERGY RESOLUTION OF SCINTILLATION DETECTOR WITH NAI(T1) SINGLE-CRYSTAL
    PERKOV, AI
    FEDOTOV, SN
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1988, 31 (03) : 590 - 592
  • [10] Comparative study for intermediate crystal size of NaI(Tl) scintillation detector
    Singh, Inderjeet
    Singh, Bhajan
    Sandhu, B. S.
    Sabharwal, Arvind D.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (07):