Precise alignment of separated Sagnac interferometer

被引:0
|
作者
Zhang, Xuemin [1 ,2 ]
Wei, Ruyi [1 ,2 ]
Yan, Su [1 ]
Duan, Jiayou [1 ]
Li, Hua [1 ]
Yang, Jianfeng [1 ]
机构
[1] Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an, Shaanxi 710119, China
[2] University of Chinese Academy of Sciences, Beijing 100049, China
来源
关键词
Reflection - Alignment - Fiber optic sensors - Interferometers;
D O I
10.3788/CJL201340.0416001
中图分类号
学科分类号
摘要
A separated Sagnac interferometer with optical diameter of 60 mm and spectral channel of 65 is introduced based on its working principle. Ideal interference fringes and interference fringes are simulated when reflector has an angle deviation, and the effect of alignment error to the interference fringes is analyzed. The position accuracy and angle accuracy of alignment are induced based on the optical design parameters. The method of precision alignment is introduced in detail, which includes the installation of primary alignment plane, the precise location of splitting prism, the precise adjustment of long-arm and short-arm reflectors. Through the installation of alignment refrence and the masterly refrence transformation, the precise alignment of separated Sagnac interferometer is achieved based on the principle of auto-collimation, whose position accuracy is better than 0.01 mm, angle accuracy is better than 1″, super possition accuracy of primary working plane is better than 1″.
引用
收藏
相关论文
共 50 条
  • [31] Nonlinear spectrally selective sagnac interferometer
    Starodumov, AN
    Zenteno, LA
    Monzon, D
    Boyain, AB
    SECOND IBEROAMERICAN MEETING ON OPTICS, 1996, 2730 : 509 - 513
  • [32] Continuous phase amplification with a Sagnac interferometer
    Starling, David J.
    Ben Dixon, P.
    Williams, Nathan S.
    Jordan, Andrew N.
    Howell, John C.
    PHYSICAL REVIEW A, 2010, 82 (01):
  • [33] AN ALIGNMENT INTERFEROMETER
    SAUNDERS, JB
    APPLIED OPTICS, 1963, 2 (05): : 541 - 542
  • [34] ALIGNMENT INTERFEROMETER
    SAUNDERS, JB
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1963, 53 (04) : 526 - &
  • [35] ALIGNMENT INTERFEROMETER
    BALDWIN, RR
    WHITTEN, LG
    MECHANICAL ENGINEERING, 1965, 87 (03) : 83 - &
  • [36] Study on the precise alignment of hybrid refractive-diffractive elements based on Fizeau interferometer
    Zhao, LP
    Wu, MX
    Jin, GF
    Zhao, ZY
    Yan, YB
    CURRENT DEVELOPMENTS IN OPTICAL ELEMENTS AND MANUFACTURING, 1998, 3557 : 138 - 142
  • [37] Double-Diffracted Spatially Separated Heterodyne Grating Interferometer and Analysis on its Alignment Tolerance
    Chang, Di
    Xing, Xu
    Hu, Pengcheng
    Wang, Jianing
    Tan, Jiubin
    APPLIED SCIENCES-BASEL, 2019, 9 (02):
  • [38] Sagnac -like Interferometer with a Double Ring Resonator
    Wang, Jinfang
    Zhang, Yundong
    Zhang, Xuenan
    Wang, Nan
    Yuan, Ping
    2009 LASERS & ELECTRO-OPTICS & THE PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1 AND 2, 2009, : 718 - +
  • [39] Sagnac interferometer for gravitational-wave detection
    Sun, KX
    Fejer, MM
    Gustafson, E
    Byer, RL
    PHYSICAL REVIEW LETTERS, 1996, 76 (17) : 3053 - 3056
  • [40] An optical microfiber Sagnac interferometer with adjustable transmission
    Lim, Sun Do
    Park, Kyoung Jun
    Kim, Byoung Yoon
    Lee, Kwanil
    Lee, Sang Bae
    2010 CONFERENCE ON OPTICAL FIBER COMMUNICATION OFC COLLOCATED NATIONAL FIBER OPTIC ENGINEERS CONFERENCE OFC-NFOEC, 2010,