Internal flow analysis of a shaft-less double-suction blood pump

被引:0
|
作者
Zhuang, Bao-Tang [1 ]
Luo, Xian-Wu [2 ]
Wei, Wei [2 ]
Xu, Hong-Yuan [2 ]
Ji, Jing-Jing [2 ]
机构
[1] Beijing Institute of Control Engineering, Beijing,100190, China
[2] State Key Laboratory of Hydroscience and Engineering, Tsinghua University, Beijing,100084, China
关键词
Numerical models - Shear flow - Blood - Shear stress - Biocompatibility;
D O I
暂无
中图分类号
学科分类号
摘要
Based on the concept of double-suction impeller combined with motor rotor, a novel shaft-less blood pump has been designed. In order to investigate the hydraulic performance and the biocompatibility of the pump, the unsteady numerical simulation for three dimensional flows in entire pump passage near the design condition is conducted. In the calculations, the pulsatile flow of a natural artery is used as the boundary condition at pump inlet so as to observe the flow characteristics such as the velocity, pressure and wall shear stress in flow passage, and the effect on the operating stability of the pump. Based on the results of numerical simulation, the internal flow will be further discussed. © 2015, Science Press. All right reserved.
引用
收藏
页码:2608 / 2613
相关论文
共 50 条
  • [1] A novel shaft-less double suction mini pump
    NISHI Michihiro
    Science China(Technological Sciences) , 2010, (01) : 105 - 110
  • [2] A novel shaft-less double suction mini pump
    Luo XianWu
    Zhu Lei
    Zhuang BaoTang
    Xu HongYuan
    Yu WeiPing
    Nishi, Michihiro
    SCIENCE CHINA-TECHNOLOGICAL SCIENCES, 2010, 53 (01) : 105 - 110
  • [3] A novel shaft-less double suction mini pump
    NISHI Michihiro
    Science China(Technological Sciences), 2010, 53 (01) : 105 - 110
  • [4] A novel shaft-less double suction mini pump
    XianWu Luo
    Lei Zhu
    BaoTang Zhuang
    HongYuan Xu
    WeiPing Yu
    Michihiro Nishi
    Science in China Series E: Technological Sciences, 2010, 53 : 105 - 110
  • [5] Design optimization for a shaft-less double suction mini turbo pump
    Zhuang, B.
    Luo, X.
    Zhang, Y.
    Wang, X.
    Xu, H.
    Nishi, M.
    25TH IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND SYSTEMS, 2010, 12
  • [6] Flow Characteristic Analysis of the Impeller Inlet Diameter in a Double-Suction Pump
    Jang, Hyunjun
    Suh, Junho
    ENERGIES, 2024, 17 (09)
  • [7] Research of the Flow Attenuation Mechanism of a Double-Suction Pump
    Meng, Dongrong
    Jiang, Ting
    Liu, Zhengguang
    Zhao, Meng
    Jiang, Wei
    Hou, Gaoyang
    FRONTIERS IN ENERGY RESEARCH, 2022, 10
  • [8] Internal flow numerical simulation of double-suction centrifugal pump using DES model
    Zhou, P. J.
    Wang, F. J.
    Yang, M.
    26TH IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND SYSTEMS, PTS 1-7, 2013, 15
  • [9] Analysis of internal flow field of OTS 125-500A low flow rate double-suction centrifugal pump
    Yuan, Wen
    Wang, Changbin
    Zhang, Kai
    Chen, Xi
    ENERGY SCIENCE AND APPLIED TECHNOLOGY, 2016, : 173 - 177
  • [10] DMD Analysis of Unsteady Flow of Impeller in Double-suction Centrifugal Pump as Turbine
    Miao S.
    Liu L.
    Wang X.
    Yang J.
    Wu J.
    Nongye Jixie Xuebao/Transactions of the Chinese Society for Agricultural Machinery, 2024, 55 (06): : 150 - 158