Performance of a BeO-based dosimetry system for proton and electron beam dose measurements

被引:0
|
作者
Bossin, Lily [1 ]
Dal Bello, Riccardo [2 ,3 ]
Christensen, Jeppe Brage [1 ]
Schischke, Stefan [4 ]
Motta, Silvia [1 ]
Togno, Michele [5 ]
Yukihara, Eduardo Gardenali [1 ]
机构
[1] Paul Scherrer Inst, Dept Radiat Safety & Secur, Villigen, Switzerland
[2] Univ Hosp Zurich, Dept Radiat Oncol, Zurich, Switzerland
[3] Univ Zurich, Zurich, Switzerland
[4] Radpro, Remscheid, Germany
[5] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
关键词
OSL dosimetry; Proton dosimetry; Electron dosimetry; Ultra-high dose rate; Dosimetry system; BeO; OPTICALLY STIMULATED LUMINESCENCE; BERYLLIUM-OXIDE; RATE DEPENDENCE; OSL; THERMOLUMINESCENCE;
D O I
10.1016/j.radmeas.2024.107310
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This study aims to evaluate the performance of the BeO-based myOSLchip system (RadPro International GmbH, Remscheid, Germany) for dosimetry of proton and electron radiotherapy beams. Although beryllium oxide (BeO) has been recognised as a promising material for luminescence dosimetry in radiotherapy, this research extends beyond material properties and examines the entire BeO-based dosimetry system, including the detector, holder, and readout components. Packages of myOSLchip dosimeters were irradiated either in a ( 70-230 ) MeV proton beam or in a 16 MeV electron beam. The readouts were carried out using the portable myOSLchip reader. In the electron beam, tests on the precision, dose response up to 100 Gy and dose-rate effects of the system were carried out. In the proton beam, the system was tested for its dose response (up to 10 Gy), fading, and linear energy transfer (LET) response. For proton irradiations, the myOSLchip BeO OSLDs exhibited stability within 2% over 135 days, as well as a linear dose response in the tested range, (0.1 - 10) Gy. The efficiency showed a reduction for proton beams with LET values (for water) above 0.6 keV/mu m, with up to 40% loss inefficiency at 4 keV/mu m. For the electron irradiations, they showed a linear dose-response up to 20 Gy and dose-rate independence, with a constant response at least up to 2.99 x 105 Gy s-1. Using individual dosimeter sensitivity correction, the precision fora single dosimeter was around 3.5% (standard deviation of the data of all dosimeters) and fora package comprising four dosimeters was 1.7% (standard deviation of the mean of the four dosimeters). These findings suggest the myOSLchip system's potential as a reliable alternative to existing dosimetry systems in clinical applications.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Ferrous sulphate gel dosimetry and MRI for proton beam dose measurements
    Bäck, SÅJ
    Medin, J
    Magnusson, P
    Olsson, P
    Grusell, E
    Olsson, LE
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (08): : 1983 - 1996
  • [2] PROTON BEAM DOSIMETRY AND DOSE DISTRIBUTIONS
    PRESTON, WM
    KOEHLER, AM
    PHYSICS IN MEDICINE AND BIOLOGY, 1966, 11 (01): : 151 - &
  • [3] Performance of a Portable MOSFET Dosimetry System for InVivo Dose Measurements in Orthovoltage Treatments and Kilovoltage Cone Beam CT
    Amin, M.
    Kohli, K.
    Norrlinger, B.
    Lindsay, P.
    Heaton, R.
    Ramaseshan, R.
    Hallil, A.
    Islam, M.
    MEDICAL PHYSICS, 2009, 36 (06) : 2611 - +
  • [4] Investigation of the HyperscintTM plastic scintillation dosimetry system performance in a 15 MeV electron beam at FLASH dose rates
    Duzenli, Cheryl
    Mendez, Claudia
    Petric, Peter
    Karan, Tania
    MEDICAL PHYSICS, 2021, 48 (08) : 4666 - 4666
  • [5] Three-Dimensional Measurements of Dose and LET From a Proton Beam Via Polymer Gel Dosimetry
    Vredevoogd, K.
    Ibbott, G.
    Gillin, M.
    Sahoo, N.
    Kry, S.
    Gifford, K.
    Maryanski, M.
    MEDICAL PHYSICS, 2012, 39 (06) : 3726 - 3726
  • [6] Advanced Proton Beam Dosimetry Part I: review and performance evaluation of dose calculation algorithms
    Saini, Jatinder
    Traneus, Erik
    Maes, Dominic
    Regmi, Rajesh
    Bowen, Stephen R.
    Bloch, Charles
    Wong, Tony
    TRANSLATIONAL LUNG CANCER RESEARCH, 2018, 7 (02) : 171 - 179
  • [7] MOSFET dose measurements for proton SOBP beam
    Hsing, Chun-Hui
    Oanh, Luu Dang Hoang
    Chao, Tsi-Chian
    Lee, Chung-Chi
    Hong, Ji-Hong
    Cheng, Chun-Chi
    Tseng, Chien-Kai
    Tung, Chuan-Jong
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 185 - 190
  • [8] Performance of a scintillation imaging system for relative dosimetry in pencil beam scanning proton therapy
    Liu, Qi
    Gong, Liangde
    Li, Xiufang
    Grossmann, Martin
    Wang, Jie
    Guo, Mengya
    Gu, Shuaizhe
    Lin, Ye
    Zhang, Manzhou
    Pu, Yuehu
    Chen, Zhiling
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1040
  • [9] Validation of the National Research Council alanine dosimetry system for linac electron beam measurements
    Surensoy, Rodi
    Muir, Bryan
    MEDICAL PHYSICS, 2021, 48 (08) : 4689 - 4689
  • [10] Investigation of Proton Beam Dose Distributions with 3D Dosimetry
    Grant, R.
    Ibbott, G.
    Zhu, X.
    Adamovics, J.
    Oldham, M.
    Followill, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S885 - S886