Comparison of Pulse Sequences of Magnetic Resonance Imaging for Optimization of Timing and Image Quality

被引:0
|
作者
Amin N. [1 ]
Yousaf M. [2 ]
Javid M.A. [3 ]
Farooq A. [1 ]
机构
[1] Department of Physics, COMSATS University Islamabad, Lahore Campus
[2] Mathematics Department, COMSATS University Islamabad, Lahore Campus
[3] Department of Basic Sciences, UET, Taxila
关键词
Diagnosis; Imaging; Magnetic Resonance; Phantoms; Signal to noise ratio;
D O I
10.22038/ijmp.2019.38606.1501
中图分类号
学科分类号
摘要
Introduction: The present study aimed to three frequently used pulse sequences of magnetic resonance imaging (MRI) to assess the image quality of theses pulse sequences at short acquisition time. Material and Methods: For the purpose of study two tissue equivalent gels were prepared. One gel was made from Polysaccharide and Agarose, whereas second gel was obtained from Ferrous Benzoic Xylenol Orange (FBX) which is tissue equivalent material. 6MV photons were used to irradiate FBX gel from linear accelerator with 25 Gray dose. Imaging parameters are performed in repetition time (TR) for experimental variations. The quantitative analysis included contrast-to-noise ratio (CNR) and signal to noise ratio (SNR). Results: As evidenced by obtained results at 1.5 Tesla, Fast Spin Echo (FSE) and Fast Fluid Attenuated Inversion Recovery (FLAIR) were most comparable in SNR although, acquisition time of FSE is 62%, 9 %, and 15% less than FLAIR at different values of 4000ms, 4200ms and 4600ms of TR. CNR of Conventional Spin Echo (CSE) was 143% and 93% better than FSE and FLAIR respectively. The time difference between CSE and FSE was 6 min and 34 sec while this difference was 6 min and 43 sec between CSE and FLAIR. Conclusion: FSE and FLAIR produced optimal image quality for many tissues. Their reduced acquisition time could make them perfect option for patients who cannot tolerate longer imaging time. Nonetheless long acquisition time cannot undervalue importance of CSE since it has yielded significantly higher contrast and SNR in T2-weighted images among other pulse sequences of MRI. © 2020. All Rights Reserved.
引用
收藏
页码:350 / 358
页数:8
相关论文
共 50 条
  • [21] ASSESSMENT OF IMAGE QUALITY IN MAGNETIC-RESONANCE-IMAGING
    TODDPOKROPEK, A
    GORDON, I
    JASZCZAK, R
    BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (704): : 822 - 822
  • [22] Optimization of pulse sequences in ultrafast magnetic resonance imaging for the diagnosis of acute abdominal pain caused by gastrointestinal disease
    Inoue, Akitoshi
    Furukawa, Akira
    Nitta, Norihisa
    Takaki, Kai
    Ohta, Shinichi
    Murata, Kiyoshi
    ACTA RADIOLOGICA OPEN, 2020, 9 (08)
  • [23] Magnetic resonance imaging with fluid attenuated pulse sequences in evaluation of craniospinal lesions
    Palmieri, A
    Muto, M
    Izzo, R
    Girardi, C
    RIVISTA DI NEURORADIOLOGIA, 1997, 10 (02): : 223 - 234
  • [24] Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: Technical considerations
    Tyler, Damian J.
    Robson, Matthew D.
    Henkelman, R. Mark
    Young, Ian R.
    Bydder, Graeme M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (02) : 279 - 289
  • [25] Magnetic resonance imaging of entheses using ultrashort TE (UTE) pulse sequences
    Benjamin, Michael
    Bydder, Graeme M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (02) : 381 - 389
  • [26] Multi-parametric optimization of magnetic resonance imaging sequences for magnetic resonance-guided radiotherapy
    Fahad, Hafiz Muhammad
    Dorsch, Stefan
    Zaiss, Moritz
    Karger, Christian P.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 28
  • [27] Optimization of Pulse Sequences in Magnetic Resonance Lymphography of Axillary Lymph Nodes Using Magnetic Nanoparticles
    Gharehaghaji, Nahideh
    Oghabian, Mohammad Ali
    Sarkar, Saeed
    Amirmohseni, Saeedeh
    Ghanaati, Hossein
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2009, 9 (07) : 4448 - 4452
  • [28] Image quality at synthetic brain magnetic resonance imaging in children
    So Mi Lee
    Young Hun Choi
    Jung-Eun Cheon
    In-One Kim
    Seung Hyun Cho
    Won Hwa Kim
    Hye Jung Kim
    Hyun-Hae Cho
    Sun-Kyoung You
    Sook-Hyun Park
    Moon Jung Hwang
    Pediatric Radiology, 2017, 47 : 1638 - 1647
  • [29] Image quality at synthetic brain magnetic resonance imaging in children
    Lee, So Mi
    Choi, Young Hun
    Cheon, Jung-Eun
    Kim, In-One
    Cho, Seung Hyun
    Kim, Won Hwa
    Kim, Hye Jung
    Cho, Hyun-Hae
    You, Sun-Kyoung
    Park, Sook-Hyun
    Hwang, Moon Jung
    PEDIATRIC RADIOLOGY, 2017, 47 (12) : 1638 - 1647
  • [30] Magnetic resonance imaging - a new quality in measurement of pulse wave velocity
    Katulska, Katarzyna
    Kociemba, Anna
    Pyda, Malgorzata
    ARTERIAL HYPERTENSION, 2014, 18 (01): : 47 - 50