Calibration of temperature-sensitivity coefficient of fiber Bragg grating at ultra-low temperature

被引:0
|
作者
Jin K. [1 ]
Ding L. [1 ]
Guo H. [1 ]
Chen G. [1 ]
Hu Y. [1 ]
机构
[1] National Engineering Laboratory of Optical Fiber Sensing Technology, Wuhan University of Technology, Wuhan
关键词
Fiber Bragg grating; Optical fiber sensign; Organic modified ceramic(ORMOCER) coating; Temperature calibration; Temperature-sensitivity coefficient; Ultra-low temperature;
D O I
10.37188/OPE.20223001.0056
中图分类号
学科分类号
摘要
In order to address the limitation of poor reliability of temperature-sensitive coefficient calibration of grating in an ultra-low temperature environment, a reference thermometer probe and fiber Bragg grating sensor were encapsulated in a self-designed non-contact liquid-nitrogen-cooled temperature measuring mold, and calibration experiments were conducted at ultra-low temperatures ranging from 93 K to 293 K. The thermal sensitivity coefficient of the bare grating and thermal expansion coefficient of the coating were used to verify the credibility of the experimental design. The experimental results indicate that the maximum initial temperature change rate of the reference thermometer is 1.8 K/min, which effectively reduces the temperature change rate of the temperature measuring mold and improves the temperature consistency between the reference thermometer and labeled grating. The test results are in good agreement with those of comparable studies. The temperature sensitivity of the bare grating decreases from 9.18 pm/K@293 K to 2.19 pm/K@93 K due to its low temperature nonlinearity. The thermal expansion coefficient of organic modified ceramic (ORMOCER) is 3.7×10-6 K-1 at room temperature. The temperature-sensitivity coefficient of one layer of the ORMOCER coating with a thickness of 50 μm is 4.43 pm/K. At 93 K, the temperature-sensitive coefficient is 7.17 pm/K, the temperature-sensitivity coefficient and linearity of the coating grating are significantly improved. © 2022, Chinese Institute of Electronics. All right reserved.
引用
收藏
页码:56 / 61
页数:5
相关论文
共 17 条
  • [11] ZHANG H J, DENG F P, XIAO J, Et al., Study on temperature, response of fiber Bragg grating at liquid helium temperature, Journal of Optoelectronics·Laser, 5, pp. 581-583, (2008)
  • [12] GUO Z S, FENG J, HUI W., Cryogenic temperature characteristics of the fiber Bragg grating sensors, Cryogenics, 52, 10, pp. 457-460, (2012)
  • [13] HABISREUTHER T, HAILEMICHAEL E, ECKE W, Et al., ORMOCER coated fiber-optic Bragg grating sensors at cryogenic temperatures, Sensors Journal, IEEE, 12, 1, pp. 13-16, (2012)
  • [14] REID M B, OZCAN M., Temperature dependence of fiber optic Bragg gratings at low temperatures, Optical Engineering, 37, 1, pp. 237-240, (1998)
  • [15] DENG F P, SHAO J Y, HUANG G J, Et al., Study on temperature characteristics of optical fiber Bragg grating under the 77 K environment, Journal of Optoelectronics·Laser, 4, pp. 404-406, (2007)
  • [16] GUO H Y, TANG J G, LI X F, Et al., On-line writing identical and weak fiber Bragg grating arrays, Chinese Optics Letters, 11, 3, (2013)
  • [17] SU F., Influence of coating parameters on temperature sensitivity of FBG, Optical Fiber & Electric Cable and Their Applications, 5, pp. 29-31, (2009)