Evaluation of biventricular function by cadmium-zinc-telluride SPECT gated tomographic radionuclide angiography: Comparison to conventional SPECT

被引:0
|
作者
Chen, Yue [1 ]
Pang, Zekun [1 ]
Wang, Jiao [1 ]
Yang, Xuewen [1 ]
Li, Jianming [1 ]
机构
[1] TEDA Int Cardiovasc Hosp, Dept Nucl Med, Tianjin, Peoples R China
关键词
biventricular function; cadmium-zinc-telluride; CZT; equilibrium radionuclide ventriculography; ERNV; repeatability; single-photon emission tomography; SPECT; BLOOD-POOL SPECT; RIGHT-VENTRICULAR VOLUMES; EJECTION FRACTION; MYOCARDIAL-PERFUSION; VALUES; CAMERA;
D O I
10.1097/MD.0000000000039821
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We compared and analyzed the consistency and repeatability of left and right ventricular ((LV/RV) functions obtained by gated-equilibrium radionuclide ventriculography (ERNV) with cadmium-zinc-telluride single-photon emission computed tomography (CZT-SPECT) and conventional SPECT (C-SPECT) with sodium iodide crystal detectors. Seventy-seven patients were included in the retrospective study. Both C-SPECT and CZT-SPECT imaging were performed on the same day. Correlations and differences in LV/RV ejection fraction (LVEF and RVEF), peak ejection rate (PER), and peak filling rate (PFR) were compared between the 2 models. Cardiac magnetic resonance (CMR) was partially used as the gold standard, and ultrasound results were included for comparative analysis. Interobserver reproducibility of each parameter obtained by the 2 cameras was compared. Between the 2 cameras, there were no significant difference in LVEF, LVPER, LVPFR, and RVPER (P > .05) and there were in RVEF and RVPFR (P < .05 or .001). The correlations (R value) were 0.831 (LVEF, excellent), 0.619 (RVEF, good), 0.672 (LVPER, good), 0.700 (LVPFR, good), 0.463 (RVPER, normal), and 0.253 (RVPFR, poor). There were no significant difference between CMR and CZT-SPECT in LVEF (P > .05) while there were between CMR and both C-SPECT and ultrasound (P < .05). The correlations were all good (R = 0.660, 0.658, and 0.695). There were no significant difference between CMR and both C-SPECT and CZT-SPET in RVEF (P > .05) and the correlations were good (R = 0.771 and 0.745). For repeatability, the intraclass correlation coefficient of RVPFR by C-SPECT was good (intraclass correlation coefficient = 0.698) and excellent for the rest of the groups (0.823-0.989). The repeatability of LVEF and RVEF was better for CZT-SPECT than for C-SPECT. The repeatability of PER was better for both cameras than PFR. CZT-SPECT tomographic ERNV correlated well with C-SPECT planar ERNV in evaluation of biventricular systolic function and LV diastolic function. Compared with the "gold standard" CMR, both models had good correlation in measuring LV/RVEF. CZT-SPECT had better inter-group reproducibility than C-SPECT. The accuracy of RV diastolic function need further study. CZT-SPECT tomographic ERNV will play an important and unique role in the clinical application of accurate evaluation of biventricular function in the future.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Dosage optimization in cadmium-zinc-telluride SPECT gamma camera gated tomographic radionuclide angiography
    Rydberg, J.
    Nguyen, T.
    Zerahn, B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S404 - S404
  • [2] Evaluation of individualized dosage planning when performing Cadmium-Zinc-Telluride SPECT camera gated tomographic radionuclide angiography
    Huang, C.
    Ismail, A.
    Kristensen, B.
    Zerahn, B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S734 - S735
  • [3] Comparison of biventricular ejection fractions using cadmium-zinc-telluride SPECT and planar equilibrium radionuclide angiography
    Yi-Chieh Chen
    Chi-Lun Ko
    Rouh-Fang Yen
    Mei-Fang Lo
    Yih-Hwen Huang
    Pei-Ying Hsu
    Yen-Wen Wu
    Mei-Fang Cheng
    Journal of Nuclear Cardiology, 2016, 23 : 348 - 361
  • [4] Comparison of biventricular ejection fractions using cadmium-zinc-telluride SPECT and planar equilibrium radionuclide angiography
    Chen, Yi-Chieh
    Ko, Chi-Lun
    Yen, Rouh-Fang
    Lo, Mei-Fang
    Huang, Yih-Hwen
    Hsu, Pei-Ying
    Wu, Yen-Wen
    Cheng, Mei-Fang
    JOURNAL OF NUCLEAR CARDIOLOGY, 2016, 23 (03) : 348 - 361
  • [5] An Algorithm for Individual Dosage in Cadmium-Zinc-Telluride SPECT-Gated Radionuclide Angiography
    Hansen, Maria Normand
    Haarmark, Christian
    Kristensen, Bent
    Zerahn, Bo
    DIAGNOSTICS, 2021, 11 (12)
  • [6] The influence of anthropometric and basic circulatory variables on count rate in cadmium-zinc-telluride SPECT gated radionuclide angiography
    Julie Rydberg
    Jannick Andersen
    Christian Haarmark
    Bo Zerahn
    Journal of Nuclear Cardiology, 2019, 26 : 1974 - 1980
  • [7] The influence of anthropometric and basic circulatory variables on count rate in cadmium-zinc-telluride SPECT gated radionuclide angiography
    Rydberg, Julie
    Andersen, Jannick
    Haarmark, Christian
    Zerahn, Bo
    JOURNAL OF NUCLEAR CARDIOLOGY, 2019, 26 (06) : 1974 - 1980
  • [8] The effect of dose reduction on left ventricular volumes and ejection fraction assessed by cadmium-zinc-telluride SPECT gamma camera gated tomographic radionuclide angiography
    Nguyen, T. Q.
    Rydberg, J.
    Zerahn, B.
    Kristensen, B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S170 - S170
  • [9] Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras
    Jensen, Maria Maj
    Schmidt, Ulla
    Huang, Chenxi
    Zerahn, Bo
    JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (02) : 384 - 396
  • [10] Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras
    Maria Maj Jensen
    Ulla Schmidt
    Chenxi Huang
    Bo Zerahn
    Journal of Nuclear Cardiology, 2014, 21 : 384 - 396