A Forked Microvascular Phantom for Ultrasound Localization Microscopy Investigations

被引:0
|
作者
Shangguan, Hanyue [1 ,2 ]
Yiu, Billy Y. S. [1 ,2 ,3 ]
Chee, Adrian J. Y. [1 ,2 ]
Yu, Alfred C. H. [1 ,2 ]
机构
[1] Univ Waterloo, Schlegel Res Inst Aging, Waterloo, ON N2L 3G1, Canada
[2] Univ Waterloo, Dept Elect & Comp Engn, Waterloo, ON N2L 3G1, Canada
[3] Tech Univ Denmark, Dept Hlth Technol, DK-2800 Lyngby, Denmark
基金
美国国家卫生研究院;
关键词
Lumen; Phantoms; Microscopy; Acoustics; Location awareness; Containers; Casting; Additive manufacturing; forked lumen; microvascular phantom; ultrasound localization microscopy (ULM); 3-D ULTRASOUND; SUPERRESOLUTION; RESOLUTION; IMAGE; ANGIOGENESIS; LIMIT; FLOW;
D O I
10.1109/TUFFC.2024.3409518
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In the development of ultrasound localization microscopy (ULM) methods, appropriate test beds are needed to facilitate algorithmic performance calibration. Here, we present the design of a new ULM-compatible microvascular phantom with a forked, V-shaped wall-less flow channel pair ( 250 mu m channel width) that is bifurcated at a separation rate of 50 mu m/mm. The lumen core was fabricated using additive manufacturing, and it was molded within a polyvinyl alcohol (PVA) tissue-mimicking slab using the lost-core casting method. Measured using optical microscopy, the lumen core's flow channel width was found to be 252 +/- 15 mu m with a regression-derived flow channel separation gradient of 50.89 mu m/mm. The new phantom's applicability in ULM performance analysis was demonstrated by feeding microbubble (MB) contrast flow (1.67 to 167 mu L/s flow rates) through the phantom's inlet and generating ULM images with a previously reported method. Results showed that, with longer acquisition times (10 s or longer), ULM image quality was expectedly improved, and the variance of ULM-derived flow channel measurements was reduced. Also, at axial depths near the lumen's bifurcation point, the current ULM algorithm showed difficulty in properly discerning between the two flow channels because of the narrow channel-to-channel separation distance. Overall, the new phantom serves well as a calibration tool to test the performance of ULM methods in resolving small vasculature.
引用
收藏
页码:887 / 896
页数:10
相关论文
共 50 条
  • [1] Ultrasound Localization Microscopy for the Assessment of Microvascular Circulation in Ischemic Perinatal Stroke
    Regensburger, Adrian P.
    Wachter, Felix
    Denis, Louise
    Mandelbaum, Henriette
    Schey, Frauke
    Buehler, Adrian
    Siebenlist, Gregor
    Schwarz, Simone
    Schulz-Heise, Susanne
    Rompel, Oliver
    Trollmann, Regina
    Woelfle, Joachim
    Juengert, Joerg
    Couture, Olivier
    Reutter, Heiko
    Hanslik, Gregor
    Knieling, Ferdinand
    STROKE, 2024, 55 (12) : e323 - e325
  • [2] Ultrasound localization microscopy
    Dencks, Stefanie
    Schmitz, Georg
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (03): : 292 - 308
  • [3] A Novel Microflow Phantom Dedicated to Ultrasound Microvascular Measurements
    Grand-Perret, Virginie V.
    Jacquet, Jean-Rene
    Leguerney, Ingrid I.
    Benatsou, Baya
    Gregoire, Jean-Marc
    Willoquet, Georges
    Bouakaz, Ayache
    Lassau, Nathalie
    Pitre-Champagnat, Stephanie
    ULTRASONIC IMAGING, 2018, 40 (05) : 325 - 338
  • [4] Microvascular flow dictates the compromise between spatial resolution and acquisition time in Ultrasound Localization Microscopy
    Hingot, Vincent
    Errico, Claudia
    Heiles, Baptiste
    Rahal, Line
    Tanter, Mickael
    Couture, Olivier
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [5] Optimizing in vivo data acquisition for robust clinical microvascular imaging using ultrasound localization microscopy
    Huang, Chengwu
    Lok, U-Wai
    Zhang, Jingke
    Zhu, Xiang Yang
    Krier, James D.
    Stern, Amy
    Knoll, Kate M.
    Petersen, Kendra E.
    Robinson, Kathryn A.
    Hesley, Gina K.
    Bentall, Andrew J.
    Atwell, Thomas D.
    Rule, Andrew D.
    Lerman, Lilach O.
    Chen, Shigao
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (07):
  • [6] Microvascular flow dictates the compromise between spatial resolution and acquisition time in Ultrasound Localization Microscopy
    Vincent Hingot
    Claudia Errico
    Baptiste Heiles
    Line Rahal
    Mickael Tanter
    Olivier Couture
    Scientific Reports, 9
  • [7] Assessment of microvascular fl ow in human atherosclerotic carotid plaques using ultrasound localization microscopy
    Leroy, Henri
    Wang, Louise Z.
    Jimenez, Anatole
    Mohamedi, Nassim
    Papadacci, Clement
    Julia, Pierre
    El Batti, Salma
    Alsac, Jean-Marc
    Sitruk, Jonas
    Arnoux, Armelle
    Bruneval, Patrick
    Messas, Emmanuel
    Mirault, Tristan
    Goudot, Guillaume
    Pernot, Mathieu
    EBIOMEDICINE, 2025, 111
  • [8] Geometric Ultrasound Localization Microscopy
    Hahne, Christopher
    Sznitman, Raphael
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION, MICCAI 2023, PT X, 2023, 14229 : 217 - 227
  • [9] Volumetric Ultrasound Localization Microscopy
    Denis, Louise
    Chabouh, Georges
    Heiles, Baptiste
    Couture, Olivier
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2024, 71 (12) : 1643 - 1656
  • [10] Aging-related cerebral microvascular changes visualized using ultrasound localization microscopy in the living mouse
    Lowerison, Matthew R.
    Sekaran, Nathiya Chandra
    Zhang, Wei
    Dong, Zhijie
    Chen, Xi
    Llano, Daniel A.
    Song, Pengfei
    SCIENTIFIC REPORTS, 2022, 12 (01)