Comparison of efficacy and safety of the new generation helixone dialyzers

被引:0
|
作者
Maduell, Francisco [1 ]
Broseta, Jose Jesus [1 ]
Rodriguez-Espinosa, Diana [1 ]
Rodas, Lida Maria [1 ]
Gomez, Miquel [1 ]
Arias-Guillen, Marta [1 ]
Fontsere, Nestor [1 ]
Vera, Manel [1 ]
del Carmen Salgado, Maria [2 ]
Rico, Nayra [2 ]
Ramos, Rosa [1 ]
机构
[1] Hosp Clin Barcelona, Serv Nefrol & Trasplante Renal, Barcelona, Spain
[2] Hosp Clin Barcelona, Serv Bioquim, Barcelona, Spain
来源
NEFROLOGIA | 2024年 / 44卷 / 03期
关键词
Dialyzer performance; Helixone; Hemodiafiltration; Adverse reactions; ONLINE HEMODIAFILTRATION; DIALYSIS MEMBRANES; HEMODIALYSIS; POLYVINYLPYRROLIDONE; HYDROPHILICITY; MORTALITY; SURFACE;
D O I
10.1016/j.nefro.2024.02.005
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction: New generation helixone dialyzers has recently been developed as part of the ongoing effort to improve dialyzer hemocompatibility and avoid adverse reactions to synthetic dialyzers. This study aimed to assess the performance and albumin loss of this new dialyzer series in hemodiafiltration and compare it with the previous generation helixone series. Material and methods: A prospective study was conducted in 19 patients. Each patient underwent eight dialysis sessions with the same routine dialysis parameters; only the dialyzer varied: FX60 CorDiax, FX CorAL 60, FX600 CorDiax, FX CorAL 600, FX80 CorDiax, FX CorAL 80, FX800 CorDiax, and FX CorAL 800. The reduction ratios (RR) of urea, creatinine, (32-microglobulin, 2-microglobulin, myoglobin, kappa-free immunoglobulin light chains (KFLC), prolactin, a1-microglobulin, 1-microglobulin, a1 1-acid glycoprotein, lambda immunoglobulin light chains (XFLC), and albumin were compared intra-individually. Dialysate albumin loss was also measured. Results: All treatments were well tolerated. The mean amount of replacement fluid ranged from 31 to 34 litres. Comparison of dialysis treatments showed no differences between small molecules and even up to those the size of (32-microglobulins. 2-microglobulins. Little differences were found between myoglobin, KFLC, prolactin, a1-microglobulin, 1-microglobulin, and XFLC RRs, and only FX80 CorDiax was slightly superior to the others. Mean dialysate albumin losses were similar, with less than 2.5 grams lost in each dialyzer. The FX80 CorDiax showed slightly higher global removal scores than the other dialyzers evaluated, except for FX CorAL 800. Conclusion: The new generation helixone dialyzers series has been updated to minimise the risk of adverse reactions, while maintaining the effectiveness and albumin loss achieved by the previous most advanced helixone generation. (c) 2024 Published by Elsevier Espana, S.L.U. on behalf of Sociedad Espanola ola de Nefrolog & imath;a. This is an open access article under the CC BY-NC-ND license (http://creativecommons. org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:354 / 361
页数:8
相关论文
共 50 条
  • [31] SAFETY AND EFFICACY OF NEW DRUGS
    PINCKNEY, ER
    SCIENCE, 1973, 181 (4106) : 1205 - 1205
  • [32] HIGH-EFFICIENCY MINIATURE DIALYZERS - COMPARISON
    HOENICH, NA
    LUNO, J
    WHITE, T
    KERR, DNS
    KIDNEY INTERNATIONAL, 1977, 12 (01) : 82 - 82
  • [33] SAFETY OF LONG-TERM MULTIPLE USE OF DIALYZERS
    POLLAK, VE
    KANT, K
    CATHEY, M
    PARNELL, S
    ZASUWA, G
    LEVIN, NW
    KIDNEY INTERNATIONAL, 1985, 27 (01) : 169 - 169
  • [34] Inflammatory response of a new synthetic dialyzer membrane.: A randomised cross-over comparison between polysulfone and helixone
    Stefoni, S
    Colì, L
    Cianciolo, G
    Donati, G
    Ruggeri, G
    Ramazzotti, E
    Pohlmeier, R
    Lang, D
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (01): : 26 - 32
  • [35] THERE IS A SYNDROME ASSOCIATED WITH THE USE OF NEW DIALYZERS
    ROBSON, M
    POLLAK, VE
    KANT, KS
    CHAROENPANICH, R
    CATHEY, M
    KIDNEY INTERNATIONAL, 1985, 27 (01) : 170 - 170
  • [36] NEW HOLLOW-FIBER DIALYZERS
    SIGDELL, JE
    ARTIFICIAL ORGANS, 1986, 10 (02) : 156 - 165
  • [37] Inflammatory response of a new synthetic dialyzer membrane. A randomised cross-over comparison between polysulfone and helixone
    Colì, L. (emodia@orsola-malpighi.med.unibo.it), 1600, Wichtig Editore s.r.l. (26):
  • [38] NEW HOLLOW-FIBER DIALYZERS
    SIGDELL, JE
    ARTIFICIAL ORGANS, 1985, 9 (02) : 208 - 209
  • [39] NEW HOLLOW-FIBER DIALYZERS
    SIGDELL, JE
    ARTIFICIAL ORGANS, 1985, 9 (01) : 69 - 70
  • [40] Comparison of efficacy and safety of treatment with second generation stents releasing paclitaxel or Everolimus. COMPARE Study
    Sosnowski, Cezary
    KARDIOLOGIA POLSKA, 2010, 68 (04) : 493 - 499