BAYESIAN EXPERIMENTAL DESIGN FOR HEAD IMAGING BY ELECTRICAL IMPEDANCE TOMOGRAPHY

被引:0
|
作者
Hyvonen, N. [1 ]
Jaaskelainen, A. [1 ]
Maity, R. [2 ]
Vavilov, A. [1 ]
机构
[1] Aalto Univ, Dept Math & Syst Anal, POB 11100, FI-00076 Aalto, Finland
[2] Univ Innsbruck, Engn Math, Technikerstr 13, A-6020 Innsbruck, Austria
基金
芬兰科学院;
关键词
electrical impedance tomography; head imaging; Bayesian experimental design; A-optimality; adaptivity; edge-promoting prior; lagged diffusivity; OPTIMIZING ELECTRODE POSITIONS; OPTIMAL CURRENT PATTERNS; A-OPTIMAL DESIGN; IN-VIVO; CONDUCTIVITY; MODELS; CONVERGENCE; TISSUES; SKULL;
D O I
10.1137/23M1624634
中图分类号
O29 [应用数学];
学科分类号
070104 ;
摘要
This work considers the optimization of electrode positions in head imaging by electrical impedance tomography. The study is motivated by maximizing the sensitivity of electrode measurements to conductivity changes when monitoring the condition of a stroke patient, which justifies adopting a linearized version of the complete electrode model as the forward model. The algorithm is based on finding a (locally) A-optimal measurement configuration via gradient descent with respect to the electrode positions. The efficient computation of the needed derivatives of the complete electrode model is one of the focal points. Two algorithms are introduced and numerically tested on a three-layer head model. The first one assumes a region of interest and a Gaussian prior for the conductivity in the brain, and it can be run offline, i.e., prior to taking any measurements. The second algorithm first computes a reconstruction of the conductivity anomaly caused by the stroke with an initial electrode configuration by combining lagged diffusivity iteration with sequential linearizations, which can be interpreted to produce an approximate Gaussian probability density for the conductivity perturbation. It then resorts to the first algorithm to find new, more informative positions for the available electrodes with the constructed density as the prior.
引用
收藏
页码:1718 / 1741
页数:24
相关论文
共 50 条
  • [21] Current source design for electrical impedance tomography
    Ross, AS
    Saulnier, GJ
    Newell, JC
    Isaacson, D
    PHYSIOLOGICAL MEASUREMENT, 2003, 24 (02) : 509 - 516
  • [22] Imaging of the head using induced current impedance tomography
    Towers, CM
    Freeston, IL
    Tozer, RC
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 774 - 775
  • [23] Study on human brain impedance imaging using electrical impedance tomography
    Xu, GZ
    Dong, GY
    Yan, WL
    Yang, QX
    PROCEEDINGS OF THE FOURTH INTERNATIONAL CONFERENCE ON ELECTROMAGNETIC FIELD PROBLEMS AND APPLICATIONS, 2000, : 378 - 380
  • [24] Experimental design to reduce inductive coupling in spectral electrical impedance tomography (sEIT) measurements
    Wang, Haoran
    Huisman, Johan Alexander
    Zimmermann, Egon
    Vereecken, Harry
    GEOPHYSICAL JOURNAL INTERNATIONAL, 2021, 225 (01) : 222 - 235
  • [25] A review on electrical impedance tomography for pulmonary perfusion imaging
    Nguyen, D. T.
    Jin, C.
    Thiagalingam, A.
    McEwan, A. L.
    PHYSIOLOGICAL MEASUREMENT, 2012, 33 (05) : 695 - 706
  • [26] Single Cell Imaging Using Electrical Impedance Tomography
    Sun, Tao
    Tsuda, Soichiro
    Zauner, Klaus-Peter
    Morgan, Hywel
    2009 4TH IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS, VOLS 1 AND 2, 2009, : 858 - 863
  • [27] Electrical impedance tomography imaging of epileptic activity in humans
    Eadie, L
    Bagshaw, A
    Bayford, R
    Binnie, C
    Holder, D
    NEUROIMAGE, 2001, 13 (06) : S1314 - S1314
  • [28] Capacitively Coupled Electrical Impedance Tomography in Lung Imaging
    Guo, Yuxi
    Zhu, Liying
    Wang, Minmin
    Jiang, Yandan
    Soleimani, Manuchehr
    Zhang, Maomao
    IEEE SENSORS JOURNAL, 2024, 24 (20) : 33072 - 33082
  • [29] Electrical impedance tomography for high speed chest imaging
    Lionheart, WRB
    Lidgey, FJ
    McLeod, CN
    Paulson, KS
    Pidcock, MK
    Shi, Y
    PHYSICA MEDICA, 1997, 13 : 247 - 249
  • [30] Fast imaging for magnetic resonance electrical impedance tomography
    Hamamura, Mark J.
    Muftuler, L. Tugan
    MAGNETIC RESONANCE IMAGING, 2008, 26 (06) : 739 - 745