Elliptical Halbach magnet and gradient modules for low-field portable magnetic resonance imaging

被引:0
|
作者
Galve, Fernando [1 ,2 ]
Pallas, Eduardo [1 ,2 ]
Guallart-Naval, Teresa [1 ,2 ]
Garcia-Cristobal, Pablo [1 ,2 ]
Martinez, Pablo [3 ]
Algarin, Jose M. [1 ,2 ]
Borreguero, Jose [4 ]
Bosch, Ruben [1 ,2 ]
Juan-Lloris, Francisco [3 ]
Benlloch, Jose M. [1 ,2 ]
Alonso, Joseba [1 ,2 ]
机构
[1] CSIC, Inst Mol Imaging & Instrumentat I3M, MRILab, Valencia 46022, Spain
[2] Univ Politecn Valencia UPV, Valencia 46022, Spain
[3] PhysioMRI Tech SL, Valencia, Spain
[4] Tesoro Imaging SL, Valencia, Spain
关键词
elliptical geometry; Halbach magnets; low-field MRI; optimization toolbox;
D O I
10.1002/nbm.5258
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This study aims to develop methods to design the complete magnetic system for a truly portable MRI scanner for neurological and musculoskeletal (MSK) applications, optimized for field homogeneity, field of view (FoV), and gradient performance compared to existing low-weight configurations. We explore optimal elliptic-bore Halbach configurations based on discrete arrays of permanent magnets. In this way, we seek to improve the field homogeneity and remove constraints to the extent of the gradient coils typical of Halbach magnets. Specifically, we have optimized a tightly packed distribution of magnetic Nd2Fe14B cubes with differential evolution algorithms and a second array of shimming magnets with interior point and differential evolution methods. We have also designed and constructed an elliptical set of gradient coils that extend over the whole magnet length, maximizing the distance between the lobe centers. These are optimized with a target field method minimizing a cost function that considers also heat dissipation. We have employed the new toolbox to build the main magnet and gradient modules for a portable MRI scanner designed for point-of-care and residential use. The elliptical Halbach bore has semi-axes of 10 and 14& cm, and the magnet generates a field of 87& mT homogeneous down to 5700& ppm (parts per million) in a 20-cm diameter FoV; it weighs 216& kg and has a width of 65& cm and a height of 72& cm. Gradient efficiencies go up to around 0.8& mT/m/A, for a maximum of 12& mT/m within 0.5& ms with 15& A and 15& V amplifier. The distance between lobes is 28& cm, significantly increased with respect to other Halbach-based scanners. Heat dissipation is around 25& W at maximum power, and gradient deviations from linearity are below 20% in a 20-cm sphere.Elliptic-bore Halbach magnets enhance the ergonomicity and field distribution of low-cost portable MRI scanners, while allowing for full-length gradient support to increase the FoV. This geometry can be potentially adapted for a prospective low-cost whole-body technology. A toolbox to develop optimal elliptic-bore Halbach magnets and gradients for easy industrializability is presented.The toolbox is used to develop an 87-mT magnetic resonance imaging (MRI) magnet for neurological and musculoskeletal (MSK) applications, with 216-kg weight, 44.5-cm length, 65-cm width, and full length gradients with 28-cm lobe-to-lobe distance.image
引用
收藏
页数:19
相关论文
共 50 条
  • [31] LOW-FIELD MAGNETIC RESONANCE
    GARSTENS, MA
    KAPLAN, JI
    PHYSICAL REVIEW, 1955, 99 (02): : 459 - 463
  • [32] Low-Field Portable Magnetic Resonance Imaging for Post-Thrombectomy Assessment of Ongoing Brain Injury
    Sujijantarat, Nanthiya
    Koo, Andrew B.
    Jambor, Ivan
    Malhotra, Ajay
    Mazurek, Mercy H.
    Parasuram, Nethra
    Yadlapalli, Vineetha
    Chavva, Isha R.
    Lalwani, Dheeraj
    Zabinska, Julia
    Roy, Joanna M.
    Antonios, Joseph P.
    Elsamadicy, Aladine A.
    Renedo, Daniela
    Hebert, Ryan M.
    Schindler, Joseph L.
    Gilmore, Emily J.
    Sansing, Lauren H.
    de Havenon, Adam
    Olexa, Madelynne
    Schiff, Steven J.
    Iglesias, Juan Eugenio
    Rosen, Matthew
    Kimberly, William Taylor
    Petersen, Nils H.
    Sheth, Kevin N.
    Matouk, Charles C.
    STROKE-VASCULAR AND INTERVENTIONAL NEUROLOGY, 2023, 3 (05):
  • [33] HEMATOMA VOLUME AND CLINICAL OUTCOME ON LOW-FIELD, PORTABLE, POINT-OF-CARE MAGNETIC RESONANCE IMAGING
    Yuen, M.
    Mazurek, M.
    Cahn, B.
    Shah, J.
    Kundu, P.
    Welch, E. B.
    By, S.
    Ward, A.
    Timario, N.
    Falcone, G.
    Hwang, D.
    Kim, J.
    Jasne, A.
    Gobeske, K.
    Petersen, N.
    Sharma, R.
    Schindler, J.
    Matouk, C.
    Sansing, L.
    Wira, C.
    Gilmore, E.
    Sze, G.
    Rosen, M.
    Kimberly, W. T.
    Sheth, K.
    INTERNATIONAL JOURNAL OF STROKE, 2020, 15 (1_SUPPL) : 96 - 97
  • [34] Imaging of the musculoskeletal system using low-field magnetic resonance imaging
    Pogarell, Tobias
    May, Matthias S.
    Nagel, Armin M.
    Uder, Michael
    Heiss, Rafael
    RADIOLOGIE, 2022, 62 (05): : 410 - 417
  • [35] Imaging of the musculoskeletal system using low-field magnetic resonance imaging
    Pogarell, Tobias
    May, Matthias S.
    Nagel, Armin M.
    Uder, Michael
    Heiss, Rafael
    RADIOLOGE, 2022, 62 (05): : 410 - 417
  • [36] Functional magnetic resonance imaging in a low-field intraoperative scanner
    Schulder, M
    NEUROSURGERY, 2003, 53 (02) : 474 - 474
  • [37] Low-field magnetic resonance imaging using multiplicative regularization
    den Bouter, Merel de Leeuw
    van Gijzen, Martin
    Remis, Rob
    MAGNETIC RESONANCE IMAGING, 2021, 75 : 21 - 33
  • [38] LOW-FIELD MAGNETIC-RESONANCE IMAGING IN MUSCULOSKELETAL TUMORS
    MCMILLAN, NC
    MCCREATH, G
    PATTERSON, J
    BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (728): : 765 - 765
  • [39] Functional magnetic resonance imaging in a low-field intraoperative scanner
    Schulder, M
    Azmi, H
    Biswal, B
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2003, 80 (1-4) : 125 - 131
  • [40] Low-Field Magnetic Resonance Imaging Its History and Renaissance
    Hori, Masaaki
    Hagiwara, Akifumi
    Goto, Masami
    Wada, Akihiko
    Aoki, Shigeki
    INVESTIGATIVE RADIOLOGY, 2021, 56 (11) : 669 - 679