Optimization and experimental validation of deuterium-tritium neutron monoenergetics for diamond detector testing

被引:0
|
作者
Yuan, Guoliang [1 ]
Wen, Zuowei [1 ]
Zhao, Wei [1 ]
Luo, Xiaobing [2 ]
Feng, Lei [1 ]
Wei, Lingfeng [1 ]
Zhang, Jinwen [1 ]
Yang, Qingwei [1 ]
机构
[1] Southwestern Inst Phys, Chengdu 610225, Sichuan, Peoples R China
[2] Sichuan Univ, Inst Nucl Sci & Technol, Chengdu 610065, Sichuan, Peoples R China
基金
中国国家自然科学基金;
关键词
Neutron source; Plasma diagnostics; Dimond detector; Neutron energy spectrum; Neutron energy spectroscopy; SPECTRA;
D O I
10.1016/j.nima.2024.169708
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Diamond detectors are excellent tools for measuring deuterium-tritium (D-T) neutron energy spectra. It has great potential for application in the field of magnetic confinement fusion. However, the current experimental values of the energy resolution of diamond detectors are much lower than the theoretical estimates. The purpose of this study is to find a method for experimental optimization. For this purpose, the D-T neutron energy distribution of the accelerator is theoretically analyzed and simulated. A method to increase the deuteron energy to obtain a better monoenergetic neutron source is proposed. This strategy was validated experimentally. The energy resolution of diamond for D-T neutrons achieves 1.21%, a Full Width at Half Maximum of 93 keV at 7.65 MeV. It is the highest resolution recorded to date. These results demonstrate the feasibility of the optimization method.
引用
收藏
页数:5
相关论文
共 50 条
  • [11] Experimental measurement of nuclear heating in a graphite-cantered assembly in deuterium-tritium neutron environment for the validation of data and calculation
    Kumar, A
    Ikeda, Y
    Youssef, MZ
    Abdou, MA
    Maekawa, F
    Kasugai, Y
    FUSION ENGINEERING AND DESIGN, 1998, 42 : 307 - 318
  • [12] CROSS CALIBRATION OF NEUTRON DETECTORS FOR DEUTERIUM-TRITIUM OPERATION IN TFTR
    JOHNSON, LC
    BARNES, CW
    DUONG, HH
    HEIDBRINK, WW
    JASSBY, DL
    LOUGHLIN, MJ
    ROQUEMORE, AL
    RUSKOV, E
    STRACHAN, JD
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 894 - 896
  • [13] Neutron-Resonance Transmission Analysis with a Compact Deuterium-Tritium Neutron Generator
    Klein, Ethan A.
    Naqvi, Farheen
    Bickus, Jacob E.
    Lee, Hin Y.
    Danagoulian, Areg
    Goldston, Robert J.
    PHYSICAL REVIEW APPLIED, 2021, 15 (05)
  • [14] Heat transfer analysis of rotating tritium target of deuterium-tritium fusion neutron generator
    Wang Gang
    Yu Qian-Feng
    Wang Wen
    Song Gang
    Wu Yi-Can
    ACTA PHYSICA SINICA, 2015, 64 (10)
  • [15] Neutron emission Doppler-shift measurements in deuterium-tritium plasmas
    Frenje, J
    Ballabio, L
    Conroy, S
    Ericsson, G
    Gorini, G
    Källne, J
    Prandoni, P
    Tardocchi, M
    Traneus, E
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01): : 1176 - 1180
  • [16] Neutron spectrometry of radio-frequency heated deuterium-tritium plasmas
    Tardocchi, M
    Ballabio, L
    Conroy, S
    Ericsson, G
    Frenje, J
    Gorini, G
    Guadagno, C
    Källne, J
    Prandoni, P
    Traneus, E
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01): : 1171 - 1175
  • [17] ANALYSIS OF THE AMPLIFICATION OF DEUTERIUM-TRITIUM NEUTRON SOURCES .2. APPLICATION
    SEGEV, M
    TACZANOWSKI, S
    NUCLEAR SCIENCE AND ENGINEERING, 1985, 91 (02) : 153 - 161
  • [19] Modification of neutron emission spectrum by Alfven eigenmodes in a deuterium-tritium plasma
    Sugiyama, Shota
    Matsuura, Hideaki
    FUSION ENGINEERING AND DESIGN, 2019, 146 : 320 - 324
  • [20] Development and testing of a fission-like neutron field for dosimetry and instrument calibrations based on deuterium-tritium generator
    Mozhayev, Andrey V.
    Piper, Roman K.
    Christ, Josef F.
    Turner, Jarrod E.
    Guerrero, Rodrigo
    Maine, Andrew L.
    Harris Jr, William S.
    Lockwood, Gabriel E.
    RADIATION PHYSICS AND CHEMISTRY, 2024, 216