A highly sensitive and selective electrochemical sensor based on polythiourea modified glassy carbon electrode for detecting fenitrothion in vegetables

被引:1
|
作者
Yihun, Tihunie Fentahun [1 ,3 ]
Aragaw, Belete Asefa [1 ]
Tsegaye, Abebaw Adgo [1 ]
Tefera, Molla [2 ]
Ambo, Melaku Metto [4 ]
机构
[1] Bahir Dar Univ, Coll Sci, Dept Chem, POB 79, Bahir Dar, Ethiopia
[2] Univ Gondar, Dept Chem, POB 196, Gondar, Ethiopia
[3] Mekdela Amba Univ, Coll Nat & Computat Sci, Dept Chem, Tulu Awuliya, Ethiopia
[4] Injibara Univ, Coll Nat & Computat Sci, Dept Chem, Injibara, Ethiopia
来源
关键词
Electrochemical sensor; Fenitrothion; Polythiourea; Voltammetry; REDUCED GRAPHENE OXIDE; CHROMATOGRAPHY; NANOCOMPOSITE; PESTICIDES;
D O I
10.1016/j.ijoes.2024.100754
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
A sensitive and selective sensor based on a polythiourea-modified glassy carbon electrode (polythiourea/GCE) was fabricated for electrochemical determination of fenitrothion (FT). Cyclic voltammetry and impedance spectroscopy were used to characterize the electrochemical behavior of polythiourea/GCE. The chemical and structural characteristics of the modified electrode were studied with UV-V is and FTIR spectroscopy techniques. The polythiourea (GCE) demonstrated faster electron transfer kinetics, a small charge transfer resistance, and an enhanced electroactive surface area of 0.173 cm2, 2 , which was two times greater than that of the bare GCE. The electrochemical behavior of FT at polythiourea/GCE was studied using cyclic voltammetry and square wave voltammetry. The reduction of FT on the surface of the modified electrode followed an adsorption-controlled process, and interestingly, the electrode surface coverage determined from chronocoulometric measurements showed 3.3 times higher accumulation of FT than at the bare GCE. Under optimized experimental parameters, the adsorptive stripping square-wave voltammetric current signal of FT at polythiourea/GCE exhibited an excellent sensitivity of 10.05 mu A mu M-1,- 1 , good linearity ranging from 0.05 to 10 mu M and 10-50 mu M, and low limits of detection and quantification of 2.1 and 7.0 nM, respectively. The selectivity of the sensor was also evaluated in the presence of several common interfering substances, such as nitrophenol, malathion, diazinon, ascorbic acid, K+, + , Mg2+,and 2+, and Zn2+ 2+ ions, and the percent error was less than +/- 5 %. The practical application of the suggested electrochemical sensor for FT determination was carried out in tomato, sweet potato, and cabbage samples. The average recoveries of the spiked samples were between 94.3 % and 110 %, indicating that the developed method is accurate for FT determination.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] A Simple Electrochemical Sensor Based on Graphene Nanoplatelets Modified Glassy Carbon Electrode (GrNPs/GCE) for Highly Sensitive Detection of Yohimbine (YOH)
    Rodsud, Siraprapa
    Limbut, Warakorn
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2019, 166 (10) : B771 - B779
  • [22] Sensitive and selective electrochemical sensor using silver nanoparticles modified glassy carbon electrode for determination of cholesterol in bovine serum
    Nantaphol, Siriwan
    Chailapakul, Orawon
    Siangproh, Weena
    SENSORS AND ACTUATORS B-CHEMICAL, 2015, 207 : 193 - 198
  • [23] Highly sensitive hydrogen peroxide sensor based on a glassy carbon electrode modified with platinum nanoparticles on carbon nanofiber heterostructures
    Yang, Yang
    Fu, Renzhong
    Yuan, Jianjun
    Wu, Shiyuan
    Zhang, Jialiang
    Wang, Haiying
    MICROCHIMICA ACTA, 2015, 182 (13-14) : 2241 - 2249
  • [24] Determination of Fenitrothion in Water Using a Voltammetric Sensor Based on a Polymer-Modified Glassy Carbon Electrode
    Amare, Meareg
    Abicho, Samuel
    Admassie, Shimelis
    JOURNAL OF AOAC INTERNATIONAL, 2014, 97 (02) : 580 - 585
  • [26] Highly sensitive hydrogen peroxide sensor based on a glassy carbon electrode modified with platinum nanoparticles on carbon nanofiber heterostructures
    Yang Yang
    Renzhong Fu
    Jianjun Yuan
    Shiyuan Wu
    Jialiang Zhang
    Haiying Wang
    Microchimica Acta, 2015, 182 : 2241 - 2249
  • [27] Electrochemical Sensor for Highly Sensitive Detection of Gallic Acid Based on Graphitized and Carboxylated Multi-Walled Carbon Nanotubes modified Glassy Carbon Electrode
    Zhu, Gan
    Wang, Qian
    Liu, Yunhang
    Guo, Meimei
    Liu, Runqiang
    Zhao, Hongyuan
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2022, 17 (10):
  • [28] A sensitive electrochemical sensor for cryptotanshinone based on electrochemical reduced graphene oxide and carbon nanotube composite film modified glassy carbon electrode
    Liu, Hui
    Han, Huai-yuan
    Liu, Ming-hui
    Zhao, Yong-fu
    INDIAN JOURNAL OF CHEMISTRY SECTION A-INORGANIC BIO-INORGANIC PHYSICAL THEORETICAL & ANALYTICAL CHEMISTRY, 2020, 59 (12): : 1822 - 1827
  • [29] Fabrication of a Highly Sensitive and Selective Electrochemical Imidacloprid Sensor Using a Glassy Carbon Electrode Modified With MWCNTs/SBA-15Si-CDs Nanocomposite
    Kazemifard, Nafiseh
    Ensafi, Ali A.
    Shirani, Marziyeh Poshteh
    IEEE SENSORS JOURNAL, 2021, 21 (08) : 9763 - 9770
  • [30] Electrochemical sensor based on phospholipid modified glassy carbon electrode - determination of paraquat
    Tomkova, H.
    Sokolova, R.
    Opletal, T.
    Kucerova, P.
    Kucera, L.
    Souckova, J.
    Skopalova, J.
    Bartak, P.
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 821 : 33 - 39