Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging

被引:1
|
作者
Li, Da [1 ]
Ren, Yong [2 ]
Liu, Feng [3 ]
机构
[1] Univ Sci & Technol China, Hefei 230031, Peoples R China
[2] Chinese Acad Sci, Hefei Inst Phys Sci, Inst Plasma Phys, Hefei 200031, Peoples R China
[3] Univ Queensland, Sch Elect Engn & Comp Sci, Brisbane, Qld 4072, Australia
关键词
Coils; Magnetic resonance imaging; Superconducting magnets; Magnetic fields; Windings; Stress; Electromagnetic forces; Head scanner; magnetic resonance imaging (MRI); REBCO coil; superconducting magnet;
D O I
10.1109/TASC.2024.3426543
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A 0.5 T REBCO MRI coil with a warm bore of 560 mm was designed for head imaging using linear programming (LP) and nonlinear programming (NLP). The LP method was used to determine the initial distribution of the coil cluster, followed by the NLP method, which optimized the radii and heights of the coil blocks using this initial layout as a reference. Both layer-winding and double-pancake (DP) winding coils were explored for developing MRI magnets. The layer-winding coil enabled the MRI magnet to achieve a uniformity of 10 parts per million (ppm) over a 200-mm diameter spherical volume. In contrast, the DP winding coils attained a field uniformity of 17.6 ppm without employing either iron or superconducting shim coils. The stray magnetic field of the MRI magnet was designed to be within 0.5 mT at radial and axial positions of 3.2 and 4.2 m, respectively. Because the REBCO magnet has a substantial temperature margin, it can be effectively cooled using solid nitrogen at 35 K or through conduction cooling. This article presents the electromagnetic design and mechanical analysis results of the coils.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 50 条
  • [31] THE COIL DESIGN OF THE SUPERCONDUCTING MRI MAGNET
    FUJITA, M
    IEEE TRANSACTIONS ON MAGNETICS, 1988, 24 (06) : 2907 - 2909
  • [32] Genetic algorithms for MRI magnet design
    Shaw, NR
    Ansorge, RE
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2002, 12 (01) : 733 - 736
  • [33] Development of a 0.5-T/240-mm MgB2 MRI Magnet: Assembly Design and Module Coils
    Ling, Jiayin
    Voccio, John
    Hahn, Seungyong
    Kim, Youngjae
    Song, Jungbin
    Bascunan, Juan
    Iwasa, Yukikazu
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2014, 24 (03)
  • [34] Temperature Dependence of Optimal Shape and DC Current Transport Characteristics of 3 T Whole Body REBCO MRI Magnet
    Nakamura, Taketsune
    Yamada, Shoko
    Kido, Yushi
    Wei, Liangliang
    Matsuda, Tetsuya
    Miura, Hideaki
    Hattori, Taisuke
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2022, 32 (06)
  • [35] Design and Magnetic Field Simulation of 3.0T MRI Superconducting Magnet
    Yue, Zengkai
    Sun, Zhizhou
    Liu, Zhizhen
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (02)
  • [36] Structural Design of a 9.4 T Whole-Body MRI Superconducting Magnet
    Dai, Yinming
    Wang, Qiuliang
    Wang, Chunzhong
    Li, Lankai
    Wang, Housheng
    Ni, Zhipeng
    Song, Shousen
    Chen, Shunzhong
    Zhao, Baozhi
    Wang, Hui
    Li, Yi
    Hu, Xinning
    Cui, Chunyan
    Cheng, Junsheng
    Lei, Yuanzhong
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)
  • [37] Reduction of TMS Strength Near MRI Scanner Could be Explained by Electromagnetic Coupling to MRI Magnet
    Peterchev, Angel V.
    BRAIN STIMULATION, 2014, 7 (06) : 916 - 917
  • [38] Design of an MRI magnet with an ellipsoidal uniform field
    Konovalov, AB
    Volegov, PL
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1995, 38 (05) : 678 - 682
  • [39] Compact MRI magnet design by stochastic optimization
    Crozier, S
    Doddrell, DM
    JOURNAL OF MAGNETIC RESONANCE, 1997, 127 (02) : 233 - 237
  • [40] Gradient coil design in permanent magnet MRI
    Hu Guo-jun
    Peng Yu-hua
    Zhao Wei
    Tang Xiao-ying
    JOURNAL OF IRON AND STEEL RESEARCH INTERNATIONAL, 2006, 13 : 419 - 422