Multi-Magnet Cochlear Implant Technology and Magnetic Resonance Imaging: The Safety Issue

被引:1
|
作者
Canzi, Pietro [1 ,2 ]
Carlotto, Elena [3 ]
Zanoletti, Elisabetta [4 ]
Frijns, Johan H. M. [5 ]
Borsetto, Daniele [6 ]
Caruso, Antonio [7 ]
Chiapparini, Luisa [8 ]
Ciorba, Andrea [9 ]
Conte, Giorgio [10 ]
Creber, Nathan [6 ]
Criscuolo, Stefania [8 ]
Di Lella, Filippo [11 ]
Franchella, Sebastiano [4 ]
Hensen, Erik F. [5 ]
Lauda, Lorenzo [7 ]
Malpede, Stefano [1 ]
Mandala, Marco [12 ]
Rotteveel, Liselotte J. C. [5 ]
Simoncelli, Anna [8 ]
Stellato, Anna Chiara [1 ]
Zanetti, Diego [13 ,14 ]
Benazzo, Marco [1 ,2 ]
机构
[1] Univ Pavia, Dept Clin Surg Diagnost & Pediat Sci, I-27100 Pavia, Italy
[2] Univ Pavia, Fdn IRCCS Policlin San Matteo, Dept Otorhinolaryngol, I-27100 Pavia, Italy
[3] S Croce Hosp, Dept Otolaryngol, I-12100 Cuneo, Italy
[4] Univ Padua, Dept Neurosci, Sect Otorhinolaryngol Head & Neck Surg, I-35127 Padua, Italy
[5] Leiden Univ, Med Ctr, Dept Otorhinolaryngol, NL-2300 RC Leiden, Netherlands
[6] Cambridge Univ Hosp NHS Fdn Trust, Addenbrookes Hosp, Dept ENT Surg, Cambridge CB2 0QQ, England
[7] Grp Otol, Dept Otol & Skull Base Surg, I-29121 Piacenza, Italy
[8] Univ Pavia, Fdn IRCCS Policlin San Matteo, Dept Diagnost Radiol & Intervent Radiol & Neurorad, I-27100 Pavia, Italy
[9] Univ Hosp Ferrara, ENT & Audiol Dept, I-44122 Ferrara, Italy
[10] t Fdn IRCCS CaGranda Osped Maggiore Policlin Milan, Neuroradiol Dept, I-20122 Milan, Italy
[11] Univ Parma, Otolaryngol & Otoneurosurg Unit, I-43126 Parma, Italy
[12] Azienda Osped Univ Senese, Dept Otol & Skull Base Surg, I-53100 Siena, Italy
[13] Fdn IRCCS CaGranda Osped Maggiore Policlin, Dept Specialist Surg Sci, Audiol Unit, I-20122 Milan, Italy
[14] Univ Milan, Dept Clin Sci & Community Hlth, I-20122 Milan, Italy
关键词
cochlear implants; magnetic resonance imaging; safety; Ultra; 3D; magnet; MRI;
D O I
10.3390/audiolres14030034
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Despite the spread of novel-generation cochlear-implant (CI) magnetic systems, access to magnetic resonance imaging (MRI) for CI recipients is still limited due to safety concerns. The aim of this study is to assess and record the experiences of Hires Ultra 3D (Advanced Bionics) recipients who underwent an MRI examination. A multicentric European survey about this topic was conducted focusing on safety issues, and the results were compared with the current literature. We collected a total of 65 MRI scans performed in 9 otologic referral centers for a total of 47 Hires Ultra 3D recipients, including, for the first time, 2 children and 3 teenagers. Preventive measures were represented by scanning time and sedation for children. Head wrapping was used in eight cases, and six of the eight cases received local anesthesia, even if both measures were not needed. Only three patients complained of pain (3/65 examinations, 4.6%) due to the tight head bandage, and one of the three cases required MRI scan interruption. No other adverse events were reported. We believe that these results should encourage MRI execution in accordance with manufacturer recommendations for Ultra 3D recipients.
引用
收藏
页码:401 / 411
页数:11
相关论文
共 50 条
  • [21] Clinical application of magnetic resonance imaging in 30 cochlear implant patients
    Baumgartner, WD
    Youssefzadeh, S
    Hamzavi, J
    Czerny, C
    Gstoettner, W
    OTOLOGY & NEUROTOLOGY, 2001, 22 (06) : 818 - 822
  • [22] Magnetic Resonance Imaging Compatibility of the Polymer-based Cochlear Implant
    Kim, Jin Ho
    Min, Kyou Sik
    An, Soon Kwan
    Jeong, Joon Soo
    Jun, Sang Beom
    Cho, Min Hyoung
    Son, Young-Don
    Cho, Zang-Hee
    Kim, Sung June
    CLINICAL AND EXPERIMENTAL OTORHINOLARYNGOLOGY, 2012, 5 : S19 - S23
  • [23] Noninvasive management of cochlear implant's inner magnet displacement after magnetic resonance
    Di Nardo, Walter
    Giannantonio, Sara
    Schinaia, Lorenzo
    De Corso, Eugenio
    Paludetti, Gaetano
    LARYNGOSCOPE, 2013, 123 (03): : 783 - 786
  • [24] Safety study of the Cochlear Nucleus® 24 Device with internal magnet in the 1.5 Tesla magnetic resonance imaging scanner
    Gubbels, SP
    McMenomey, SO
    LARYNGOSCOPE, 2006, 116 (06): : 865 - 871
  • [25] Magnetic resonance imaging and cochlear implants: Compatibility and safety aspects
    Teissl, C
    Kremser, C
    Hochmair, ES
    Hochmair-Desoyer, IJ
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 1999, 9 (01) : 26 - 38
  • [26] A Rare Complication of Cochlear Implantation After Magnetic Resonance Imaging: Reversion of the Magnet
    Ozturk, Erkan
    Doruk, Can
    Orhan, Kadir Serkan
    Celik, Mehmet
    Polat, Beldan
    Guldiken, Yahya
    JOURNAL OF CRANIOFACIAL SURGERY, 2017, 28 (04) : E372 - E374
  • [27] Adverse Events and Discomfort During Magnetic Resonance Imaging in Cochlear Implant Recipients
    Kim, Bo Gyung
    Kim, Jin Won
    Park, Jeong Jin
    Kim, Sung Huhn
    Kim, Hee Nam
    Choi, Jae Young
    JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2015, 141 (01) : 45 - 52
  • [28] Value of Routine Magnetic Resonance Imaging for the Preoperative Assessment of Cochlear Implant Candidates
    Alzhrani, Farid
    Babkour, Alaa
    Al-Muhawas, Fida
    Alsanosi, Abdulrahman
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (12)
  • [29] Magnetic Resonance Imaging Artifacts and Cochlear Implant Positioning at 1.5 T In Vivo
    Schroeer, Dirk
    Grupe, Gloria
    Rademacher, Grit
    Mutze, Sven
    Ernst, Arneborg
    Seidl, Rainer
    Mittmann, Philipp
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [30] Adverse events in pediatric cochlear implant patients undergoing magnetic resonance imaging
    Gallant, Sara C.
    Danehy, Amy R.
    Licameli, Greg R.
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2021, 140