EVALUATION OF THE BIOMECHANICAL PROPERTIES OF IN VIVO USED ORTHODONTIC MINI-IMPLANTS

被引:0
|
作者
Pop, Silvia Izabella [1 ]
Dudescu, Mircea [2 ]
Ana, Procopciuc [3 ]
Kerekes-Mathe, Bernadette-Evelyn [4 ]
Mitariu, Mihai [5 ]
Bud, Eugen [6 ]
Mitariu, Loredana [7 ]
Pacurar, Mariana [8 ]
Pop, Radu Vasile [9 ]
机构
[1] George Emil Palade Univ Med Pharm Sci & Technol T, Fac Med Dent, Dept Orthodont, 38 Gh Marinescu Str, Targu Mures 540139, Romania
[2] Tech Univ Cluj Napoca, Dept Mech Engn, 28 Memorandumului St, Cluj Napoca 400114, Romania
[3] George Emil Palade Univ Med Pharm Sci & Technol T, Dept Dent Surg, Fac Med Dent, 38 Gh Marinescu Str, Targu Mures 540139, Romania
[4] George Emil Palade Univ Med Pharm Sci & Technol T, Morphol Teeth & Dent Arches Dept, Fac Med Dent, 38 Gh Marinescu Str, Targu Mures 540139, Romania
[5] Lucian Balga Univ, Fac Med Dent, Dept Dent Med & Nursing, Blvd Victoriei, Sibiu 550024, Romania
[6] George Emil Palade Univ Med Pharm Sci & Technol T, Dept Orthodont, Fac Med Dent, 38 Gh Marinescu Str, Targu Mures 540139, Romania
[7] Lucian Balga Univ, Fac Med Dent, Dept Dent Med & Nursing, Sibiu 550024, Romania
[8] George Emil Palade Univ Med Pharm Sci & Technol T, Dept Orthodont, Fac Med Dent, 38 Gh Marinescu Str, Targu Mures 540139, Romania
[9] Nat Smile Dent Clin, 76 Gh Doja Str, Targu Mures 540232, Romania
来源
ROMANIAN JOURNAL OF ORAL REHABILITATION | 2024年 / 16卷 / 01期
关键词
Orthodontic mini-implants; Biomechanical properties; Insertion torque; STABILITY; ANCHORAGE; INSERTION; SUCCESS; DESIGN;
D O I
10.6261/RJOR.2024.1.16.20
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim of the study Orthodontic mini-implants, are devices that are temporarily attached to the jaws bones in order to improve orthodontic anchorage and to facilitate tooth movements. The aim of this study is to evaluate the biomechanical properties, namely, the insertion torque of 2 types of orthodontic mini-implants used intraorally for different periods of time. The secondary objective is to compare their properties with the new, unused mini-implants of same type and to correlate the intraoral use time with changing parameters. Material and methods The present study included a total of 51 orthodontic mini-implants, from two different manufacturers, Yesanchor OrlusTM (conical) and Link MIS TM (cylindrical), both having the same dimension 1.6 x 8 mm. These two types of mini-implants were divided into GC-control group: GCY (n=5), GCM (n=5) and GI-group used intraorally, the latter was divided according to the period of intraoral use in: GIY1 <= 12 months (n=10), GIY2 >= 12 months (n=15), GIM1 <= 12 months (n=7), GIM2 >= 12 months (n=9). In order to test the insertion torque, all the samples from the GI groups (n=41) were inserted in high-density artificial bone. Results The maximum insertion torque values for Yesanchor mini-implants varied from 35.58 Ncm to 42.34 Ncm and for MIS mini-implants from 19.44 Ncm to 25.36 Ncm. Significant decrease of the intraorally used mini-implants maximum insertion torque was demonstrated both for the Yesanchor group (p<0.05) and MIS group (p<0.05). Regarding the intraorally usage time, the differences were not significant. Conclusions Our study showed that there exist differences in maximum insertion torque between the mini-implants from two different manufacturers. We observed a decrease of the maximum insertion torque for the group of mini-implants used intraorally compared to the Control Group.
引用
收藏
页码:238 / 245
页数:8
相关论文
共 50 条
  • [11] Stress distribution in orthodontic mini-implants
    Dobranszki, A.
    Levy Neto, F.
    Dobranszki, N. C.
    Vuolo, J. H.
    Laboissiere Jr, M. A.
    MATERIA-RIO DE JANEIRO, 2010, 15 (03): : 405 - 412
  • [12] Convenient removal of orthodontic mini-implants
    Kim, Hyewon
    Kim, Tae Kyung
    Lee, Shin-Jae
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 131 (04) : S90 - S91
  • [14] Orthodontic mini-implants: clinical and peri-implant evaluation
    Vicioni-Marques, Fernanda
    Brandariz Pimentel, Diego Jesus
    Nakane Matsumoto, Mirian Aiko
    Sasso Stuani, Maria Bernadete
    Romano, Fabio Lourenco
    JOURNAL OF THE WORLD FEDERATION OF ORTHODONTISTS, 2022, 11 (01) : 22 - 28
  • [15] The validation of Periotest values for the evaluation of orthodontic mini-implants' stability
    Cha, Jung-Yul
    Yu, Hyung-Seog
    Hwang, Chung-Ju
    KOREAN JOURNAL OF ORTHODONTICS, 2010, 40 (03) : 167 - 175
  • [16] In vitro evaluation of insertion and removal torques of orthodontic mini-implants
    Pithon, M. M.
    Nojima, M. G.
    Nojima, L. I.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 40 (01) : 80 - 85
  • [17] Removal torque of osseointegrated mini-implants:: an in vivo evaluation
    Favero, Luca Giuseppe
    Pisoni, Andrea
    Paganelli, Corrado
    EUROPEAN JOURNAL OF ORTHODONTICS, 2007, 29 (05) : 443 - 448
  • [18] Load-related implant reaction of mini-implants used for orthodontic anchorage
    Büchter, A
    Wiechmann, D
    Koerdt, S
    Wiesmann, HP
    Piffko, J
    Meyer, U
    CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) : 473 - 479
  • [19] Fracture resistance of commonly used self-drilling orthodontic mini-implants
    Smith, Angie
    Hosein, Yara K.
    Dunning, Cynthia E.
    Tassi, Ali
    ANGLE ORTHODONTIST, 2015, 85 (01) : 26 - 32
  • [20] Mechanical strength of orthodontic infrazygomatic mini-implants
    Wang, Chau-Hsiang
    Wu, Ju-Hui
    Lee, Kun-Tsung
    Hsu, Kun-Rong
    Wang, Huang Chi
    Chen, Chun-Ming
    ODONTOLOGY, 2011, 99 (01) : 98 - 100