Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry

被引:0
|
作者
Langenbucher, Achim [1 ]
Wendelstein, Jascha [1 ,2 ]
Cayless, Alan [3 ]
Hoffmann, Peter [4 ]
Szentmary, Nora [5 ,6 ]
机构
[1] Saarland Univ, Dept Expt Ophthalmol, Homburg, Germany
[2] Johannes Kepler Univ Linz, Dept Ophthalmol, Linz, Austria
[3] Open Univ, Sch Phys Sci, Milton Keynes, England
[4] Augen & Laserklin Castrop Rauxel, Castrop Rauxel, Germany
[5] Saarland Univ, Dr Rolf M Schwiete Ctr Limbal Stem Cell & Aniridia, Homburg, Germany
[6] Semmelweis Univ, Dept Ophthalmol, Budapest, Hungary
来源
PLOS ONE | 2024年 / 19卷 / 04期
关键词
MULTIVARIATE; TESTS;
D O I
10.1371/journal.pone.0300576
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose The purpose of this study was to investigate the effect of the corneal back surface by comparing the keratometric astigmatism (K, derived from the corneal front surface) of a modern optical biometer against astigmatism of Total Keratometry (TK, derived from both corneal surfaces) in a large population with cataractous eyes. The results were then used to define linear prediction models to map K to TK.Methods From a large dataset containing bilateral biometric measurements (IOLMaster 700) in 9736 patients prior to cataract surgery, the total corneal astigmatism was decomposed into vectors for K, corneal back surface (BS), and TK. A multivariate prediction model (MV), simplified model with separation of vector components (SM) and a constant model (CM) were defined to map K to TK vector components.Results The K centroid (X/Y) showed some astigmatism with-the-rule (0.1981/-0.0211 dioptre (dpt)) whereas the TK centroid was located around zero (-0.0071/-0.0381 dpt against-the-rule) and the BS centroid showed systematic astigmatism against-the-rule (-0.2367/-0.0145 dpt). The respective TK-K centroid was located at -0.2052/-0.0302 dpt. The MV model showed the same performance (i.e. mean absolute residuum) as the SM did (0.1098 and 0.1099 dpt respectively) while the CM performed only slightly worse (0.1121 dpt mean absolute residuum).Conclusion In cases where tomographic data are unavailable statistical models could be used to consider the overall contribution of the back surface to the total corneal astigmatism. Since the performance of the CM is sufficiently close to that of MV and SM we recommend using the CM which can be directly considered e.g. as surgically induced astigmatism.
引用
收藏
页数:14
相关论文
共 36 条
  • [1] POSTERIOR CORNEAL ASTIGMATISM MEASURED WITH THE IOLMASTER700
    Goggin, Michael
    LaHood, Ben
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2017, 45 : 41 - 41
  • [2] Analysis of biometry and prevalence data for corneal astigmatism in 23 239 eyes
    Hoffmann, Peter Christian
    Huetz, Werner W.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (09): : 1479 - 1485
  • [3] Analysis of differences between keratometric astigmatism and total corneal astigmatism measured by IOLMaster 700C. Yu et al.Analysis of differences between keratometric astigmatism and total corneal astigmatism
    Chunxia Yu
    Xiaoling Fang
    Jinhua Tao
    Yiping Dong
    Shenyu Ben
    Yulan Wang
    International Ophthalmology, 45 (1)
  • [4] Comparison of central corneal thickness measured in myopic eyes by Pentacam, Sirius and IOLMaster 700
    Song, Han
    Liu, Chengyang
    Yang, Wei
    Yang, Chunliu
    Cheng, Xiaodong
    PHOTODIAGNOSIS AND PHOTODYNAMIC THERAPY, 2024, 49
  • [5] Agreement of Corneal Diameter and Anterior Chamber Depth Measurements with the IOLMaster 500 and the IOLMaster 700 Optical Biometers in Myopic Eyes
    Tana-Rivero, Pedro
    Tello-Elordi, Cristina
    Orts-Vila, Paz
    Tana-Sanz, Pedro
    Tana-Sanz, Santiago
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 1245 - 1253
  • [6] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Giacomo Savini
    Leonardo Taroni
    Domenico Schiano-Lomoriello
    Kenneth J. Hoffer
    Eye, 2021, 35 : 307 - 315
  • [7] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Savini, Giacomo
    Taroni, Leonardo
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    EYE, 2021, 35 (01) : 307 - 315
  • [8] Feasibility and repeatability of ocular biometry measured with IOLMaster 700 in a large population-based study
    Michael, Ralph
    Wirkner, Kerstin
    Engel, Christoph
    Loeffler, Markus
    Kirsten, Toralf
    Rauscher, Franziska G.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2023, 43 (04) : 860 - 873
  • [9] Analysis of anterior corneal astigmatism before cataract surgery using power vector analysis in eyes of Korean patients
    Oh, Eun-hye
    Kim, Hyojin
    Lee, Hyun Soo
    Hwang, Kyu-yeon
    Joo, Choun-Ki
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (06): : 1256 - 1263
  • [10] The association between central and peripheral corneal astigmatism based on Fourier analysis in normal eyes
    Maseedupally, Vinod
    Ho, Kerry
    Tran, Kevin
    Swarbrick, Helen A.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)