Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength

被引:11
|
作者
Ogawa, Masato [1 ,2 ]
Matsumoto, Tsuyoshi [3 ]
Harada, Risa [3 ]
Yoshikawa, Ryo [3 ]
Ueda, Yuya [4 ]
Takamiya, Daiki [3 ]
Sakai, Yoshitada [3 ]
机构
[1] Kobe Univ Hosp, Div Rehabil Med, Kobe, Japan
[2] Kobe Univ, Grad Sch Hlth Sci, Dept Publ Hlth, Kobe, Japan
[3] Kobe Univ, Grad Sch Med, Div Rehabil Med, Kobe, Japan
[4] Kobe Univ, Grad Sch Hlth Sci, Dept Rehabil Med, Kobe, Japan
关键词
muscle thickness; reliability; ultrasonography; validity; BODY-COMPOSITION; ULTRASOUND;
D O I
10.2490/prm.20230008
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objectives: Measurement of skeletal muscle using ultrasonography (US) has received considerable attention as an alternative method of muscle assessment. However, intra- and inter-rater reliability remains controversial. Furthermore, there is no consensus regarding the relationship between muscle assessment using US and muscle mass or physical assessment. We aimed to verify the validity and reliability of muscle measurements using US and its relationships with muscle strength and physical assessment. Methods: The 22 participants were all healthy men. Quadriceps muscle thickness was measured by US by three different raters. Intraclass correlation coefficient (ICC) was used to assess inter- and intra-rater reliability. The maximum isokinetic strength of the quadriceps and handgrip strength were used as measures of lower and upper muscle strength, respectively. Leg muscle mass was assessed using the leg skeletal muscle index (SMI), measured by body impedance analysis, and calf circumference. Results: The intra-rater reliability was excellent which the ICC(1,1) ranges 0.957-0.993, and ICC(1,3) ranges 0.985-0.998. For inter-rater reliability, the values of 0.904 for ICC(2,1) and 0.966 for ICC(2,3) indicated excellent reliability. Leg SMI was significantly correlated with quadriceps thickness (r=0.36). Maximum isokinetic strength and handgrip strength showed weak but statistically significant correlations with quadriceps thickness (r=0.20, r=0.30, respectively). The correlation between quadriceps thickness and calf circumference was not statistically significant. Conclusions: Quadriceps muscle assessment using US is a valid and reliable technique for healthy individuals. Quadriceps muscle thickness was significantly positively correlated with upper and lower muscle strength and leg SMI. Muscle thickness assessment could replace full body muscle assessment in clinical settings.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Reliability of muscle thickness measurements in ultrasonography
    Barotsis, Nikolaos
    Tsiganos, Panagiotis
    Kokkalis, Zinon
    Panayiotakis, George
    Panagiotopoulos, Elias
    INTERNATIONAL JOURNAL OF REHABILITATION RESEARCH, 2020, 43 (02) : 123 - 128
  • [2] Assessing the Reliability of Handheld Dynamometry and Ultrasonography to Measure Quadriceps Strength and Muscle Thickness in Children, Adolescents, and Young Adults
    Rock, Kelly
    Nelson, Christa
    Addison, Odessa
    Marchese, Victoria
    PHYSICAL & OCCUPATIONAL THERAPY IN PEDIATRICS, 2021, 41 (05) : 540 - 554
  • [3] Reliability of ultrasound measurements of quadriceps muscle thickness in critically ill patients
    Emmanuel Pardo
    Hanen El Behi
    Priscilla Boizeau
    Franck Verdonk
    Corinne Alberti
    Thomas Lescot
    BMC Anesthesiology, 18
  • [4] Reliability of ultrasound measurements of quadriceps muscle thickness in critically ill patients
    Pardo, Emmanuel
    El Behi, Hanen
    Boizeau, Priscilla
    Verdonk, Franck
    Alberti, Corinne
    Lescot, Thomas
    BMC ANESTHESIOLOGY, 2018, 18
  • [5] Reliability of muscle thickness and echo intensity measurements of the quadriceps: A novice examiner
    Abiko, Teppei
    Ohmae, Kento
    Murata, Shin
    Shiraiwa, Kayoko
    Horie, Jun
    JOURNAL OF BODYWORK AND MOVEMENT THERAPIES, 2022, 31 : 164 - 168
  • [6] Reliability and temporal variation of masseter muscle thickness measurements utilizing ultrasonography
    Emshoff, R
    Emshoff, I
    Rudisch, A
    Bertram, S
    JOURNAL OF ORAL REHABILITATION, 2003, 30 (12) : 1168 - 1172
  • [7] Reliability and Validity of an Ultrasound-Based Protocol for Measurement of Quadriceps Muscle Thickness in Children
    Hogelin, Emil Rydell
    Thulin, Kajsa
    von Walden, Ferdinand
    Fornander, Lotta
    Michno, Piotr
    Alkner, Bjorn
    FRONTIERS IN PHYSIOLOGY, 2022, 13
  • [8] Reliability of the Supraspinatus Muscle Thickness Measurement by Ultrasonography
    Yi, Tae Im
    Han, In Soo
    Kim, Joo Sup
    Jin, Ju Ryeon
    Han, Jea Shin
    ANNALS OF REHABILITATION MEDICINE-ARM, 2012, 36 (04): : 488 - 495
  • [9] Reliability of the Deep Abdominal Muscle Thickness Measurements Using Ultrasonography in Normal Subjects
    Otani, Yoshitaka
    Itotani, Keisuke
    Maeda, Noriaki
    Ikada, Michi
    Hatanaka, Megumi
    Kato, Junichi
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2011, 23 (03) : 357 - 359
  • [10] ISOMETRIC STRENGTH AND THICKNESS RELATIONSHIPS IN HUMAN QUADRICEPS MUSCLE
    FREILICH, RJ
    KIRSNER, RLG
    BYRNE, E
    NEUROMUSCULAR DISORDERS, 1995, 5 (05) : 415 - 422