Is day-case surgical procedure safe for MICRA leadless pacemaker implantation?

被引:1
|
作者
Toon, Lin-Thiri [1 ,2 ]
Elrefai, Mohammed [1 ]
Abouelasaad, Mohamed [1 ]
Patil, Roopa [1 ]
Paisey, John [1 ]
Yue, Arthur [1 ]
Roberts, Paul [1 ,2 ]
机构
[1] Univ Hosp Southampton NHS Fdn Trust, Cardiac Rhythm Management Res Dept, Tremona Rd, Southampton SO16 6YD, England
[2] Univ Southampton, Southampton, England
关键词
MICRA; Pacemaker; Patient management; Day-case surgery;
D O I
10.1007/s10840-024-01907-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background MICRA implantation is not commonly done as a day-case procedure. Elective leadless pacemakers are implanted routinely in our centre. Objective To assess whether the day-case MICRA procedure is safe. Methods We retrospectively collected data from all patients undergoing elective MICRA implantation at our centre between May 2014 and Nov 2022 (n = 81). Two patient groups were stratified: those planned to be discharged on the same day (SD, n = 52) and those planned to be observed overnight after the procedure (ON, n = 29). Patient demographics, size of the sheath used, type of MICRA device, and rate of complications were recorded. In patients with successful implants (n = 80), device function at discharge and first routine follow-up were evaluated. Results There were 58% males in the SD group and 45% in the ON group. Median age was 49 years in the SD and 67 years in the ON. Among patients who were planned as a day case, 8 patients had to stay in the hospital but for < 48 h: 2 due to minor groin bleeding, 1 due to patient's request despite fit to discharge, 4 due to the procedure carried out later in the day, and 1 for observation due to procedural complexity. MICRA implantation was successful in 80 patients. The rate of the major complications was 2% in the SD group and 7% in the ON group (p = 0.223), and none of the co-morbidities assessed showed an association with any complications. Device parameters at the follow-up were available in 76 patients. The rate of patients with low and stable PCT at follow-up was also 98% in the SD group and 96% in the ON group. Conclusions Day case MICRA procedure can be performed safely in an appropriately selected patient population.
引用
收藏
页码:117 / 123
页数:7
相关论文
共 50 条
  • [1] Micra leadless pacemaker and transcatheter aortic valve implantation at the same procedure
    Bastante, Teresa
    Cuesta, Javier
    Rivero, Fernando
    Del Val, David
    Martinez-Avial, Maria
    Alfonso, Fernando
    REC-INTERVENTIONAL CARDIOLOGY, 2022, 4 (04): : 338 - 340
  • [2] Mortality After Micra Leadless Pacemaker Implantation
    Kamath, Akshay R.
    Heard, Brittney Z.
    Darby, Andrew E.
    Malhotra, Rohit
    Mangrum, James M.
    Bilchick, Kenneth C.
    Mason, Pamela K.
    Ferguson, John D.
    Mehta, Nishaki K.
    CIRCULATION, 2018, 138
  • [3] Is the new Micra-leadless pacemaker entirely safe?
    Da Costa, Antoine
    Romeyer-Bouchard, Cecile
    Guichard, Jean Baptiste
    Gerbay, Antoine
    Isaaz, Karl
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 212 : 97 - 99
  • [4] OUTCOMES OF MICRA LEADLESS PACEMAKER IMPLANTATION WITH UNINTERRUPTED ANTICOAGULATION
    Kiani, Soroosh
    Black, George
    Rao, Birju
    Thakkar, Nancy
    Massad, Christopher
    Akshar, Patel
    Merchant, Faisal
    Hoskins, Michael
    Delurgio, David
    Patel, Anshul
    Shah, Anand
    Leon, Angel
    Westerman, Stacy
    Lloyd, Michael
    El Chami, Mikhael
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 420 - 420
  • [5] Implantation of Micra AV: A leadless atrioventricular synchronous pacemaker
    Li, Ke
    Liu, Jinfeng
    ASIAN JOURNAL OF SURGERY, 2023, 46 (12) : 5761 - 5762
  • [6] Outcomes of Micra leadless pacemaker implantation with uninterrupted anticoagulation
    Kiani, Soroosh
    Black, George B.
    Rao, Birju
    Thakkar, Nancy
    Massad, Christopher
    Patel, Akshar V.
    Merchant, Faisal M.
    Hoskins, Michael H.
    De Lurgio, David B.
    Patel, Anshul M.
    Shah, Anand D.
    Leon, Angel R.
    Westerman, Stacy B.
    Lloyd, Michael S.
    El-Chami, Mikhael F.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (08) : 1313 - 1318
  • [7] Leadless Pacemaker Implantation in Hemodialysis Patients Experience With the Micra Transcatheter Pacemaker
    El-Chami, Mikhael F.
    Clementy, Nicolas
    Garweg, Christophe
    Omar, Razali
    Duray, Gabor Z.
    Gornick, Charles C.
    Leyva, Francisco
    Sagi, Venkata
    Piccini, Jonathan P.
    Soejima, Kyoko
    Stromberg, Kurt
    Roberts, Paul R.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (02) : 162 - 170
  • [8] Tricuspid leaflet flail after Micra™ leadless pacemaker implantation: a case report
    Pingitore, Annachiara
    Calcagno, Simone
    Salvador, Loris
    Mennuni, Silvia
    Cavarretta, Elena
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2022, 6 (06)
  • [9] Micra leadless pacemaker after transcatheter aortic valve implantation
    Cuesta, Javier
    Bastante, Teresa
    Alfonso, Fernando
    MEDICINA CLINICA, 2020, 154 (06): : 239 - 240
  • [10] Intraoperative Implantation of Micra Leadless Pacemaker During Valve Surgery
    Marai, Ibrahim
    Diab, Sammer
    Ben-Avi, Ronny
    Kachel, Erez
    ANNALS OF THORACIC SURGERY, 2018, 105 (05): : E211 - E212