Ultra-low-field magnetization transfer imaging at 0.055T with low specific absorption rate

被引:1
|
作者
Su, Shi [1 ,2 ]
Zhao, Yujiao [1 ,2 ]
Ding, Ye [1 ,2 ]
Lau, Vick [1 ,2 ]
Xiao, Linfang [1 ,2 ]
Leung, Gilberto K. K. [3 ]
Lau, Gary K. K. [4 ]
Huang, Fan [5 ]
Vardhanabhuti, Vince [5 ]
Leong, Alex T. L. [1 ,2 ]
Wu, Ed X. [1 ,2 ]
机构
[1] Univ Hong Kong, Lab Biomed Imaging & Signal Proc, Hong Kong, Peoples R China
[2] Univ Hong Kong, Dept Elect & Elect Engn, Hong Kong, Peoples R China
[3] Univ Hong Kong, Dept Surg, Hong Kong, Peoples R China
[4] Univ Hong Kong, Dept Med, Hong Kong, Peoples R China
[5] Univ Hong Kong, Dept Diagnost Radiol, Hong Kong, Peoples R China
关键词
brain; magnetization transfer; MRI; specific absorption rate; tissue contrast; ultra-low-field; INHOMOGENEOUSLY BROADENED LINES; SPIN-LOCK; MRI; RELAXATION; TISSUE; T-1;
D O I
10.1002/mrm.30231
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To demonstrate magnetization transfer (MT) effects with low specific absorption rate (SAR) on ultra-low-field (ULF) MRI. Methods: MT imaging was implemented by using sinc-modulated RF pulse train (SPT) modules to provide bilateral off-resonance irradiation. They were incorporated into 3D gradient echo (GRE) and fast spin echo (FSE) protocols on a shielding-free 0.055T head scanner. MT effects were first verified using phantoms. Brain MT imaging was conducted in both healthy subjects and patients. Results: MT effects were clearly observed in phantoms using six SPT modules with total flip angle 3600 degrees at central primary saturation bands of approximate offset +/- 786Hz, even in the presence of large relative B-0 inhomogeneity. For brain, strong MT effects were observed in gray matter, white matter, and muscle in 3D GRE and FSE imaging using six and sixteen SPT modules with total flip angle 3600 degrees and 9600 degrees, respectively. Fat, cerebrospinal fluid, and blood exhibited relatively weak MT effects. MT preparation enhanced tissue contrasts in T2-weighted and FLAIR-like images, and improved brain lesion delineation. The estimated MT SAR was 0.0024 and 0.0008W/kg for two protocols, respectively, which is far below the US Food and Drug Administration (FDA) limit of 3.0W/kg. Conclusion: Robust MT effects can be readily obtained at ULF with extremely low SAR, despite poor relative B-0 homogeneity in ppm. This unique advantage enables flexible MT pulse design and implementation on low-cost ULF MRI platforms to achieve strong MT effects in brain and beyond, potentially augmenting their clinical utility in the future.
引用
收藏
页码:2420 / 2432
页数:13
相关论文
共 50 条
  • [11] Current-density imaging using ultra-low-field MRI with adiabatic pulses
    Nieminen, Jaakko O.
    Zevenhoven, Koos C. J.
    Vesanen, Panu T.
    Hsu, Yi-Cheng
    Ilmoniemi, Risto J.
    MAGNETIC RESONANCE IMAGING, 2014, 32 (01) : 54 - 59
  • [12] Magnetization transfer imaging using non-balanced SSFP at ultra-low field
    Balaji, Sharada
    Wiley, Neale
    Dvorak, Adam
    Padormo, Francesco
    Teixiera, Rui P. A. G.
    Poorman, Megan E.
    Mackay, Alex
    Wood, Tobias
    Cassidy, Adam R.
    Traboulsee, Anthony
    Li, David K. B.
    Vavasour, Irene
    Williams, Steven C. R.
    Deoni, Sean C. L.
    Ljungberg, Emil
    Kolind, Shannon H.
    MAGNETIC RESONANCE IN MEDICINE, 2025,
  • [13] ULTRA-LOW-FIELD MAGNETIC-RESONANCE-IMAGING OF ACUTE CRUCIATE LIGAMENT TEARS
    EKELUND, L
    BJORNEBRINK, J
    ELMQVIST, LG
    MAGNETIC RESONANCE IMAGING, 1991, 9 (02) : 179 - 185
  • [14] Solving the problem of concomitant gradients in ultra-low-field MRI
    Nieminen, Jaakko O.
    Ilmoniemi, Risto J.
    JOURNAL OF MAGNETIC RESONANCE, 2010, 207 (02) : 213 - 219
  • [15] Current-density imaging using ultra-low-field MRI with zero-field encoding
    Vesanen, Panu T.
    Nieminen, Jaakko O.
    Zevenhoven, Koos C. J.
    Hsu, Yi-Cheng
    Ilmoniemi, Risto J.
    MAGNETIC RESONANCE IMAGING, 2014, 32 (06) : 766 - 770
  • [16] Ultra-low-field MRI using superconductive magnetic sensors
    Kanazawa Institute of Technology, Applied Electronics Laboratory, Kanazawa-shi, Japan
    J. Inst. Electron. Inf. Commun. Eng., 1 (40-47):
  • [17] Conductive shield for ultra-low-field magnetic resonance imaging: Theory and measurements of eddy currents
    Zevenhoven, Koos C. J.
    Busch, Sarah
    Hatridge, Michael
    Oisjoen, Fredrik
    Ilmoniemi, Risto J.
    Clarke, John
    JOURNAL OF APPLIED PHYSICS, 2014, 115 (10)
  • [18] Optical Atomic Magnetometry for Ultra-Low-Field NMR Detection
    Bevilacqua, G.
    Biancalana, V.
    Dancheva, Y.
    Moi, L.
    ANNUAL REPORTS ON NMR SPECTROSCOPY, VOL 78, 2013, 78 : 103 - 148
  • [19] Low-Cost Purpose-Built Ultra-Low-Field NMR Spectrometer
    Chowdhury, Md Raduanul H.
    Ahmed, Firoz
    Oladun, Clementinah
    Adelabu, Isaiah
    Abdurraheem, Abubakar
    Nantogma, Shiraz
    Birchall, Jonathan R.
    Gafar, Tobi Abdulbasit
    Chekmenev, Yuri A.
    Nikolaou, Panayiotis
    Barlow, Michael J.
    Goodson, Boyd M.
    Shcherbakov, Anton
    Chekmenev, Eduard Y.
    ANALYTICAL CHEMISTRY, 2024, 96 (42) : 16724 - 16734
  • [20] Ultra-low-field magnetic resonance imaging in upper airways obstruction in sleep apnea syndrome
    Okada, T
    Fukatsu, H
    Ishigaki, T
    Yasuma, F
    Kayukawa, Y
    PSYCHIATRY AND CLINICAL NEUROSCIENCES, 1996, 50 (05) : 285 - 289