Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report

被引:0
|
作者
Urrechaga, Eloisa [1 ,2 ]
Merino, Monica [2 ]
Fernandez, Monica [3 ]
Aguirrezabal, Jaione [4 ]
del Orbe, Rafael [5 ]
机构
[1] Biocruces Bizkaia Hlth Res Inst, Cruces Plaza, Baracaldo 48903, Bizkaia, Spain
[2] Hosp Galdakao Usansolo, Core Lab, Galdakao, Vizcaya, Spain
[3] Hosp Univ Araba, Hematol Dept, Araba, Vizcaya, Spain
[4] Hosp Gernika Luno, Family & Community Med, Gernika, Vizcaya, Spain
[5] Hosp Univ Cruces, Erythropatol & Hematol, Baracaldo, Vizcaya, Spain
关键词
Hemoglobin variant (Hb variant); leukocytes differential (WBC differential); hematology analyzer; case report; VARIANT; LEIDEN;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Hemoglobin (Hb) variants are the world's most frequent monogenic disorders; up to 150 are unstable. Mild hemolytic anemia can be the only clinical evidence that a patient has unstable Hb variants, but the minimal changes in complete blood count (CBC) are difficult to observe and often mean that the condition is undetected in most cases. Case Description: We present a report of a Hb Shelby carrier. An asymptomatic 47-year-old woman born in Bolivia was referred to our medical center for a routine health control. CBC was analyzed with a Mindray BC6800 plus counter and showed values within reference ranges. Only reticulocytosis was detected. The white blood cell differential channel showed an abnormally low fluorescent signal in the leukocyte (WBC) differential scattergram. This pattern has been observed in other counters based on similar fluorescence detection principles, linked to the presence of Hb variants, so reflect tests were performed to confirm this clinical condition. Conclusions: Incidental findings in CBC are not uncommon, an abnormal WBC scatterplot must be investigated and can be used to detect carriers in order to reach the diagnosis efficiently.
引用
收藏
页数:5
相关论文
共 17 条
  • [1] Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report
    Urrechaga, Eloisa
    Merino, Monica
    Fernandez, Monica
    Aguirrezabal, Jaione
    del Orbe, Rafael
    JOURNAL OF LABORATORY AND PRECISION MEDICINE, 2022, 7
  • [2] Reference intervals for clinically reportable platelet parameters on the Mindray BC-6800Plus hematology analyzer
    Kim, Hanah
    Hur, Mina
    Kim, Seung-Wan
    Moon, Hee-Won
    Yun, Yeo-Min
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (10) : E213 - E215
  • [3] Detection of nucleated red blood cells using the Mindray BC-6800Plus hematology analyzer: a clinical performance evaluation
    Wang, Nengyong
    Jiang, Junyao
    Wang, Ji'an
    He, Yao
    Wen, Ning
    Liu, Yulin
    Wang, Yashu
    Li, Haijun
    Shi, Ping
    ANNALS OF PALLIATIVE MEDICINE, 2021, 10 (08) : 8808 - 8817
  • [4] COMPARISON BETWEEN SYSMEX XN-1000 AND MINDRAY BC-6800PLUS HEMATOLOGY ANALYZERS
    Molero-Vilches, Ignacio
    Bermejo-Rodriguez, Alfredo
    Nieto-Borrajo, Maria Elva
    Morales-Garcia, Luis Javier
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 121 - 121
  • [5] Establishing reference intervals for extended red blood cell parameters with the Mindray BC-6800Plus hematology analyzer in a Chinese population
    Luo, Yanping
    Wang, Shuaiyan
    Chen, Zhuocheng
    Lin, Shan
    Guo, Sanping
    Mo, Hongmei
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024,
  • [6] ESTABLISHING REFERENCE INTERVALS FOR EXTENDED RED BLOOD CELL PARAMETERS WITH THE MINDRAY BC-6800PLUS HEMATOLOGY ANALYZER IN A CHINESE POPULATION
    Luo, Yanping
    Wang, Shuaiyan
    Chen, Zhuocheng
    Lin, Shan
    Guo, Sanping
    Mo, Hongmei
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 : 60 - 61
  • [7] Reference intervals for 10 platelet parameters on Mindray BC-6800 plus hematology analyzer
    Kim, H.
    Hur, M.
    Kim, S.
    Moon, H.
    Yun, Y.
    CLINICA CHIMICA ACTA, 2022, 530 : S180 - S180
  • [8] REFERENCE VALUES OF LEUKOCYTE CELL POPULATION DATA (LEUKOCYTE AUTOMATED MORPHOLOGY) IN THE INSTRUMENT BC-6800 PLUS MINDRAY® HEMATOLOGY ANALYZER.
    Dhanani, Hiren
    Simon-Lopez, Ramon
    Tandel, Bharat
    Kothia, Jaysukh
    Dhanani, Nandini
    Tandel, Palak
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 : 53 - 54
  • [9] Detection of a Lymphoproliferative Disorder With Suspected Scattergram Analysis Using the Mindray BC-6800 Plus Automated Hematology Analyzer: A Case Report
    Sacchetti, Sara
    Bellia, Matteo
    Zanotti, Valentina
    Giacomini, Luca
    Rolla, Roberta
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024,
  • [10] AUTOMATED SCREENING FOR MALARIA PARASITES ON MINDRAY BC-6800 HEMATOLOGY ANALYZER: A FOLLOW-UP REPORT
    Bhide, Milind
    Parekh, Vijay
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 80 - 81