Proton beam spot size and position measurements using a multi-strip ionization chamber

被引:1
|
作者
Lin, Ye [1 ]
Zhang, Haiqun [1 ]
Gu, Shuaizhe [2 ]
Shen, Liren [3 ]
Lv, Ming [1 ]
Zhang, Manzhou [1 ]
Chen, Zhiling [1 ]
机构
[1] Chinese Acad Sci, Shanghai Adv Res Inst, Dept Particle Beam Applicat, Shanghai 201200, Peoples R China
[2] Chinese Acad Sci, Shanghai Inst Appl Phys, Shanghai 201800, Peoples R China
[3] Shanghai Adv Res Inst, Chinese Acad Sci, Dept Gen Technol, Shanghai 201200, Peoples R China
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2024年 / 123卷
关键词
Proton therapy; Multi -strip ionization chamber; Commissioning; Pencil -beam scanning; Quality assurance; QUALITY-ASSURANCE; DOSIMETRY; SYSTEM;
D O I
10.1016/j.ejmp.2024.103411
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop and characterize a large-area multi-strip ionization chamber (MSIC) for efficient measurement of proton beam spot size and position at a synchrotron-based proton therapy facility. Methods and Materials: A 420 mm x 320 mm MSIC was designed with 240 vertical strips and 180 horizontal strips at 1.75 mm pitch. The MSIC was characterized by irradiating a grid of proton spots across 17 energies from 73.5 MeV to 235 MeV and comparing to simultaneous measurements made with a reference Gafchromic EBT3 film. Beam profiles, spot sizes, and positions were analyzed. Short term measurement stability and sensitivity were evaluated. Results: Excellent agreement was demonstrated between the MSIC and EBT3 film for both spot size and position measurements. Spot sizes agreed within +/- 0.18 mm for all energies tested. Measured beam spot positions agreed within +/- 0.17 mm. The detector showed good short term measurement stability and low noise performance. Conclusion: The large-area MSIC enables efficient and accurate proton beam spot characterization across the clinical energy range. The results indicate the MSIC is suitable for pencil beam scanning proton therapy commissioning and quality assurance applications requiring fast spot size and position quantification.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Comprehensive and Efficient Spot Size and Position Quality Assurance for Proton Spot Scanning System
    Robertson, D.
    Shen, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E591 - E592
  • [32] COLLECTION EFFICIENCY OF AN IONIZATION-CHAMBER IN A PULSED SWEPT BEAM - CHAMBER SIZE EFFECTS
    MELI, JA
    WEINHOUS, MS
    PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (10): : 1139 - 1146
  • [33] A Preliminary Study on the Influence of Monitor Chamber Spot Position Threshold Setting on Actual Spot Position Accuracy for a Proton Therapy System
    Li, Z.
    Zheng, Y.
    Li, Z.
    MEDICAL PHYSICS, 2024, 51 (10) : 7876 - 7876
  • [34] Spot Size and Position Analysis of a Pencil Beam Scanning System
    Perles, L.
    Coutinho, L.
    Rana, S.
    Gutierrez, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E426 - E426
  • [35] Experimental determination of the absorbed dose to water in a scanned proton beam using a water calorimeter and an ionization chamber
    Gagnebin, Solange
    Twerenbold, Damian
    Pedroni, Eros
    Meer, David
    Zenklusen, Silvan
    Bula, Christian
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2010, 268 (05): : 524 - 528
  • [36] Operation experience of ionization-proportional chamber for low-current beam measurements at INR RAS proton linac
    Melnikov, A.
    Gavrilov, S.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (09)
  • [37] CALORIMETER AND IONIZATION-CHAMBER MEASUREMENTS OF ABSORBED DOSE IN RON BEAM
    KASE, KR
    DOMEN, SR
    HEALTH PHYSICS, 1973, 25 (03): : 305 - 306
  • [38] Proton and Carbon Ion Beam Spot Size Measurement Using 5 Different Detector Types
    Witt, Matthias
    Weber, Uli
    Adeberg, Sebastian
    Baumann, Kilian-Simon
    Zink, Klemens
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2025, 15
  • [39] Feasibility study of a novel multi-strip silicon detector for use in proton therapy range verification quality assurance
    Merchant, A. H.
    Newall, M.
    Guatelli, S.
    Petasecca, M.
    Lerch, M.
    Perevertaylo, V.
    Milluzzo, G.
    Petringa, G.
    Romano, F.
    Cirrone, G. A. P.
    Cuttone, G.
    Jackson, M.
    Rosenfeld, A. B.
    RADIATION MEASUREMENTS, 2017, 106 : 378 - 384
  • [40] Characterization of multi-strip crystal deflector for high energy proton beams by synchrotron radiation topography with angular scanning
    Kaloyan, A. A.
    Tikhomirov, S. A.
    Podurets, K. M.
    Chesnokov, Yu A.
    Sandomirskiy, Yu E.
    XI INTERNATIONAL SYMPOSIUM ON RADIATION FROM RELATIVISTIC ELECTRONS IN PERIODIC STRUCTURES (RREPS2015), 2016, 732